Comparison of the Accuracy of PET/CT and PET/MRI Spatial Registration of Multiple Metastatic Lesions

被引:36
|
作者
Rakheja, Rajan [1 ]
DeMello, Linda [1 ]
Chandarana, Hersh [2 ]
Glielmi, Christopher [3 ]
Geppert, Christian [3 ]
Faul, David [3 ]
Friedman, Kent P. [1 ]
机构
[1] NYU, Langone Med Ctr, Dept Radiol, Div Nucl Med, New York, NY 10016 USA
[2] NYU, Langone Med Ctr, Dept Radiol, Abdominal Imaging Sect, New York, NY 10016 USA
[3] Siemens Healthcare, MR R&D Collaborat, New York, NY USA
关键词
MRI; PET/CT; PET/MRI; registration; respiratory motion; MOTION ARTIFACTS; ATTENUATION CORRECTION; FDG PET; MRI; CT; REDUCTION; SCANNER;
D O I
10.2214/AJR.13.11305
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of this study was to compare the accuracy of the spatial registration of conventional PET/CT with that of hybrid PET/MRI of patients with FDG-avid metastatic lesions. SUBJECTS AND METHODS. Thirteen patients with known metastatic lesions underwent FDG PET/CT followed by PET/MRI with a hybrid whole-body system. The inclusion criterion for tumor analysis was spherical or oval FDG-avid tumor clearly identified with both CT and MRI. The spatial coordinates (x, y, z) of the visually estimated centers of the lesions were determined for PET/CT (PET and CT independently) and PET/MRI (PET, T1-weighted gradient-echo sequence with radial stack-of-stars trajectory, T2-weighted sequence), and the b0 images of an echo-planar imaging (EPI) diffusion-weighted imaging (DWI) acquisition. All MRI sequences were performed in the axial plane with free breathing. The spatial coordinates of the estimated centers of the lesions were determined for PET and CT and PET and MRI sequences. Distance between the isocenter of the lesion on PET images and on the images obtained with the anatomic modalities was measured, and misregistration (in millimeters) was calculated. The degree of misregistration was compared between PET/CT and PET/MRI with a paired Student t test. RESULTS. Nineteen lesions were evaluated. On PET/CT images, the average of the total misregistration in all planes of CT compared with PET was 4.13 +/- 4.24 mm. On PET/MR images, lesion misregistration between PET and T1-weighted gradient-echo images had a shift of 2.41 +/- 1.38 mm and between PET and b0 DW images was 5.97 +/- 2.83 mm. Similar results were calculated for 11 lesions on T2-weighted images. The shift on T2-weighted images compared with PET images was 2.24 +/- 1.12 mm. Paired Student t test calculations for PET/CT compared with PET/MRI T1-weighted gradient-echo images with a radial stack-of-stars trajectory, b0 DW images, and T2-weighted images showed significant differences (p < 0.05). Similar results were seen in the analysis of six lung lesions. CONCLUSION. PET/MRI T1-weighted gradient-echo images with a radial stack-of-stars trajectory and T2-weighted images had more accurate spatial registration than PET/CT images. This may be because that the whole-body PET/MRI system used can perform simultaneous acquisition, whereas the PET/CT system acquires data sequentially. However, the EPI-based b0 DWI datasets were significantly misregistered compared with the PET/CT datasets, especially in the thorax. Radiologists reading PET/MR images should be aware of the potential for misregistration on images obtained with EPI-based DWI sequences because of inherent spatial distortion associated with this type of MRI acquisition.
引用
收藏
页码:1120 / 1123
页数:4
相关论文
共 50 条
  • [1] Accuracy of PET and CT spatial registration of lung lesions with PET-CT.
    Cohade, C
    Osman, M
    Wahl, RL
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2002, 43 (05): : 14P - 15P
  • [2] PET-CT: accuracy of PET and CT spatial registration of lung lesions
    Christian Cohade
    Medhat Osman
    Laura T. Marshall
    Richard L. Wahl
    [J]. European Journal of Nuclear Medicine and Molecular Imaging, 2003, 30 : 721 - 726
  • [3] PET-CT: accuracy of PET and CT spatial registration of lung lesions
    Cohade, C
    Osman, M
    Marshall, LT
    Wahl, RL
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2003, 30 (05) : 721 - 726
  • [4] PET/MRI and PET/CT in Lung Lesions and Thoracic Malignancies
    Flechsig, Paul
    Mehndiratta, Amit
    Haberkorn, Uwe
    Kratochwil, Clemens
    Giesel, Frederik L.
    [J]. SEMINARS IN NUCLEAR MEDICINE, 2015, 45 (04) : 268 - 281
  • [5] Comparison of the Accuracy of a Deep Learning Method for Lesion Detection in PET/CT and PET/MRI Images
    Pang, Lifang
    Zhang, Zheng
    Liu, Guobing
    Hu, Pengcheng
    Chen, Shuguang
    Gu, Yushen
    Huang, Yukun
    Zhang, Jia
    Shi, Yuhang
    Cao, Tuoyu
    Zhang, Yiqiu
    Shi, Hongcheng
    [J]. MOLECULAR IMAGING AND BIOLOGY, 2024, 26 (05) : 802 - 811
  • [6] Quantitative PET/CT-PET/MRI comparison of lesions and normal tissues SUV values
    Lord, M.
    Garibotto, V.
    Figueiral, A.
    Heinzer, S.
    Zaidi, H.
    Ratib, O.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2011, 38 : S389 - S389
  • [7] Accuracy of PET/MRI coregistration of cervical lesions
    Vogel, Wouter V.
    Bloemers, Monique C. W. M.
    van der Heide, Uulke A.
    Beets-Tan, Regina G.
    [J]. NUCLEAR MEDICINE COMMUNICATIONS, 2016, 37 (09) : 997 - 998
  • [8] Diagnostic accuracy in gynecological malignant tumors: A comparison between PET/CT and PET/MRI fusion images
    Nakajo, Kazuya
    Tatsumi, Mituaki
    Inoue, Atsuo
    Isohashi, Kayako
    Higuchi, Ichiro
    Kato, Hiroki
    Shimosegawa, Eku
    Hatazawa, Jun
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2009, 50
  • [9] Comparison of 18F-FDG PET/CT and PET/MRI for the assessment of multiple myeloma
    Sachpekidis, C.
    Hillengass, J.
    Goldschmidt, H.
    Mosebach, J.
    Schlemmer, H.
    Haberkorn, U.
    Dimitrakopoulou-Strauss, A.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2015, 42 : S324 - S324
  • [10] Comparison and evaluation of pet/ct image registration
    Gong, Sylvia J.
    O'Keefe, Graerne J.
    Scott, Andrew M.
    [J]. 2005 27TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-7, 2005, : 1599 - 1603