Embedded Control in Wearable Medical Devices: Application to the Artificial Pancreas

被引:25
|
作者
Zavitsanou, Stamatina [1 ]
Chakrabarty, Ankush [1 ]
Dassau, Eyal [1 ]
Doyle, Francis J., III [1 ]
机构
[1] Harvard Univ, Harvard John A Paulson Sch Engn & Appl Sci, Cambridge, MA 02138 USA
来源
PROCESSES | 2016年 / 4卷 / 04期
基金
美国国家卫生研究院;
关键词
embedded control systems; artificial pancreas; software architecture; model predictive control (MPC); safety-critical applications; MODEL-PREDICTIVE CONTROL; CLOSED-LOOP CONTROL; INSULIN DELIVERY; GLUCOSE CONTROL; NONLINEAR-SYSTEMS; CLINICAL-EVALUATION; CONTROL ALGORITHM; BIONIC PANCREAS; YOUNG-PEOPLE; CONTROL; 1ST;
D O I
10.3390/pr4040035
中图分类号
TQ [化学工业];
学科分类号
0817 ;
摘要
Significant increases in processing power, coupled with the miniaturization of processing units operating at low power levels, has motivated the embedding of modern control systems into medical devices. The design of such embedded decision-making strategies for medical applications is driven by multiple crucial factors, such as: (i) guaranteed safety in the presence of exogenous disturbances and unexpected system failures; (ii) constraints on computing resources; (iii) portability and longevity in terms of size and power consumption; and (iv) constraints on manufacturing and maintenance costs. Embedded control systems are especially compelling in the context of modern artificial pancreas systems (AP) used in glucose regulation for patients with type 1 diabetes mellitus (T1DM). Herein, a review of potential embedded control strategies that can be leveraged in a fully-automated and portable AP is presented. Amongst competing controllers, emphasis is provided on model predictive control (MPC), since it has been established as a very promising control strategy for glucose regulation using the AP. Challenges involved in the design, implementation and validation of safety-critical embedded model predictive controllers for the AP application are discussed in detail. Additionally, the computational expenditure inherent to MPC strategies is investigated, and a comparative study of runtime performances and storage requirements among modern quadratic programming solvers is reported for a desktop environment and a prototype hardware platform.
引用
收藏
页数:29
相关论文
共 50 条
  • [21] Embedded Artificial Intelligence: Intelligence on Devices
    Lin, Hsiao-Ying
    [J]. COMPUTER, 2023, 56 (09) : 90 - 93
  • [22] SURGICAL APPLICATION OF AN ARTIFICIAL PANCREAS
    LAMBOTTE, L
    REYNAERT, M
    OTTE, JB
    MOULIN, D
    GRIBOMONT, B
    RAHIER, J
    LAMBERT, AE
    BUYSSCHAERT, M
    KESTENS, PJ
    [J]. EUROPEAN SURGICAL RESEARCH, 1979, 11 : 5 - 5
  • [23] A Wire-embedded Converter Used for Wearable Devices
    Tian, Mofan
    Wang, Naizeng
    Wang, Kangping
    Jia, Haiyang
    Li, Zhenwei
    Yang, Xu
    Wang, Laili
    [J]. 2017 THIRTY SECOND ANNUAL IEEE APPLIED POWER ELECTRONICS CONFERENCE AND EXPOSITION (APEC), 2017, : 121 - 125
  • [24] Perioperative Considerations for Evolving Artificial Pancreas Devices
    Long, Micah T.
    Coursin, Douglas B.
    Rice, Mark J.
    [J]. ANESTHESIA AND ANALGESIA, 2019, 128 (05): : 902 - 906
  • [25] Control of an artificial human pancreas
    Saleem, Muhammad Umer
    Farman, Muhammad
    Ahmad, M. O.
    Rizwan, Muhammad
    [J]. CHINESE JOURNAL OF PHYSICS, 2017, 55 (06) : 2273 - 2282
  • [26] CONTROL ASPECTS OF THE ARTIFICIAL PANCREAS
    SALZSIEDER, E
    FISCHER, U
    JUTZI, E
    ALBRECHT, A
    [J]. ARTIFICIAL ORGANS, 1980, 4 (02) : 148 - 148
  • [27] Enhanced, simplified glucose sensors: Long-term clinical application of wearable artificial endocrine pancreas
    Shichiri, M
    Sakakida, M
    Nishida, K
    Shimoda, S
    [J]. ARTIFICIAL ORGANS, 1998, 22 (01) : 32 - 42
  • [28] FEASIBILITY OF THE WEARABLE ARTIFICIAL ENDOCRINE PANCREAS WITH A NEEDLE-TYPE GLUCOSE SENSOR IN CLINICAL-APPLICATION
    SHICHIRI, M
    KAWAMORI, R
    YAMASAKI, Y
    HAKUI, N
    ABE, H
    [J]. DIABETES, 1983, 32 : A36 - A36
  • [29] Wearable Medical Devices: Regulatory Affairs in Argentina
    Rodriguez Campos, Theo
    Monsalve, Leandro N.
    [J]. ADVANCES IN BIOENGINEERING AND CLINICAL ENGINEERING, VOL 2, SABI 2023, 2024, 114 : 297 - 305
  • [30] Optimal FFT for Wearable Wireless Medical Devices
    Varaleshmi, K.
    Venkatalakshmi, B.
    [J]. 2015 GLOBAL CONFERENCE ON COMMUNICATION TECHNOLOGIES (GCCT), 2015, : 799 - 803