MICROBIOLOGICAL CONTAMINATION OF A HEMODIALYSIS CENTER WATER DISTRIBUTION SYSTEM

被引:30
|
作者
Montanari, Lilian Bueno [1 ]
Sartori, Flavio Garcia [1 ]
de Oliveira Cardoso, Miguel Jorge [1 ]
Varo, Samuel Dutra [1 ]
Pires, Regina Helena [1 ]
Fujimura Leite, Clarice Queico [2 ]
Prince, Karina [2 ]
Gomes Martins, Carlos Henrique [1 ]
机构
[1] Univ Franca, Lab Res Appl Microbiol, Franca, SP, Brazil
[2] State Univ Sao Paulo, Dept Biol Sci, Araraquara, SP, Brazil
关键词
Dialysis; Water; Microbiological contamination; Water treatment; Dialysis fluid quality; Bacterial count; PYROGENIC REACTIONS; DIALYSIS WATER; MYCOBACTERIA; IDENTIFICATION; QUALITY; INFECTIONS; BACTERIA; FLUID;
D O I
10.1590/S0036-46652009000100007
中图分类号
R51 [传染病];
学科分类号
100401 ;
摘要
The microbiological monitoring of the water used for hemodialysis is extremely important, especially because of the debilitated immune system of patients suffering from chronic renal insufficiency. To investigate the occurrence and species diversity of bacteria in waters, water samples were collected monthly from a hemodialysis center in upstate Sao Paulo and tap water samples at the terminal sites of the distribution system was sampled repeatedly (22 times) at each of five points in the distribution system; a further 36 samples were taken from cannulae in 19 hemodialysis machines that were ready for the next patient, four samples from the reuse system and 13 from the water storage system. To identify bacteria, samples were filtered through 0.22 mu m-pore membranes; for mycobacteria, 0.45 mu m pores were used. Conventional microbiological and molecular methods were used in the analysis. Bacteria were isolated from the distribution system (128 isolates), kidney machine water (43) and reuse system (3). Among these isolates, 32 were Gram-positive rods, 120 Gram-negative rods, 20 Gram-positive cocci and 11 mycobacteria. We propose the continual monitoring of the water supplies in hemodialysis centers and the adoption of effective prophylactic measures that minimize the exposure of these immunodeficient patients to contaminated sources of water.
引用
收藏
页码:37 / 43
页数:7
相关论文
共 50 条
  • [1] Design requirements for a water distribution system in a hemodialysis center
    Andrysiak, P
    Varughese, PM
    DIALYSIS & TRANSPLANTATION, 2002, 31 (10) : 683 - +
  • [2] Microbiological contamination of drinking water in a commercial household water filter system
    Daschner, FD
    Ruden, H
    Simon, R
    Clotten, J
    EUROPEAN JOURNAL OF CLINICAL MICROBIOLOGY & INFECTIOUS DISEASES, 1996, 15 (03) : 233 - 237
  • [3] Microbiological Contamination of Water and Health
    Wyer M.
    Stapleton C.
    Kay D.
    Kay, David (dave@crehkay.demon.co.uk), 1600, Royal Society of Chemistry (2021-January): : 70 - 94
  • [4] BACTERIAL-CONTAMINATION OF HEMODIALYSIS CENTER WATER AND DIALYSATE - ARE CURRENT ASSAYS ADEQUATE
    HARDING, GB
    PASS, T
    MILLION, C
    WRIGHT, R
    DEJARNETTE, J
    KLEIN, E
    ARTIFICIAL ORGANS, 1989, 13 (02) : 155 - 159
  • [5] Evaluation of exposure scenarios on intentional microbiological contamination in a drinking water distribution network
    Schijven, Jack
    Foret, Jean Marie
    Chardon, Jurgen
    Teunis, Peter
    Bouwknegt, Martijn
    Tangena, Ben
    WATER RESEARCH, 2016, 96 : 148 - 154
  • [6] Effect of reservoirs on microbiological water qualities in a drinking water distribution system
    Lee, Dong-Geun
    Kim, Sang-Jong
    Park, Seong Joo
    JOURNAL OF MICROBIOLOGY AND BIOTECHNOLOGY, 2006, 16 (07) : 1060 - 1067
  • [7] Optical System for Detection of microbiological Contamination
    Steinmueller, Rolf
    DEUTSCHE LEBENSMITTEL-RUNDSCHAU, 2010, 106 (10) : 573 - 579
  • [8] AN INVESTIGATION OF A WATERBORNE OUTBREAK CAUSED BY MICROBIOLOGICAL CONTAMINATION OF THE DRINKING WATER SUPPLY SYSTEM
    Ucakar, Veronika
    Grilc, Eva
    Jeraj, Irena
    ZDRAVSTVENO VARSTVO, 2012, 51 (02): : 112 - 119
  • [9] Tricholroethylene water contamination in hemodialysis centre
    Pavone, L
    Poli, D
    Tansinda, P
    Tagliavini, D
    Goldoni, M
    Franchini, I
    Mutti, A
    Cambi, V
    David, S
    NEPHROLOGY DIALYSIS TRANSPLANTATION, 2005, 20 : V331 - V331
  • [10] MICROBIAL CONTAMINATION OF WATER USED FOR HEMODIALYSIS
    BLAGG, CR