Digital flat-panel X-ray technology based on cesium iodide and amorphous silicon:: experimental studies and preliminary clinical results.

被引:0
|
作者
Strotzer, M [1 ]
Gmeinwieser, J [1 ]
Völk, M [1 ]
Fründ, R [1 ]
Feuerbach, S [1 ]
机构
[1] Klinikum Univ Regensburg, Inst Rontgendiagnostik, D-93042 Regensburg, Germany
来源
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN | 1999年 / 170卷 / 01期
关键词
radiography; digital; technology; amorphous silicon;
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Experimental and clinical evaluation of a digital flat-panel X-ray system based on cesium iodide (CsI) and amorphous silicon (a-Si). Methods: Performance of a prototype detector was compared with conventional screen-film radiography (SFR) using several phantom studies. Foreign bodies, fractures, osteolyses, and pulmonary lesions were analyzed. Additionally, 120 patients were studied prospectively, resulting in 400 comparative X-ray studies. The fiat-panel detector was exposed with standard dose and with a dose reduction of up to 75%. Detector size was 15 x 15 cm, pixel matrix was 1 x Ik with a pixel size of 143 mu m. Modulation-transfer function was determined to be 18% at the maximum spatial resolution of 3.5 Ip/mm. Results: The diagnostic results achieved with the digital detector were similar to those of conventional SFR; even at reduced radiation exposure. A potential for dose reduction was observed: 50% with respect to osteoarthrosis and fractures, end 75% for determining bony alignment. Discussion: This new technology can be used in thoracic and skeletal radiography. A significant dose reduction is possible, depending on the suspected disease.
引用
收藏
页码:66 / 72
页数:7
相关论文
共 50 条
  • [41] Development and evaluation of a portable amorphous silicon flat panel X-ray detector
    Watanabe, N
    Mochizuki, C
    Kameshima, T
    Yamazaki, T
    Court, L
    Hayashida, S
    Morishita, M
    Ohta, S
    MEDICAL IMAGING 2001: PHYSICS OF MEDICAL IMAGING, 2001, 4320 : 103 - 114
  • [42] Erratum to: Clinical evaluation of digital radiography based on a large-area cesium iodide-amorphous silicon flat-panel detector compared with screen-film radiography for skeletal system and abdomen
    Terue Okamura
    Saori Tanaka
    Koichi Koyama
    Norifumi Nishida
    Hideo Daikokuya
    Toshiyuki Matsuoka
    Kenji Kishimoto
    Masakatsu Hatagawa
    Hiroaki Kudoh
    Ryusaku Yamada
    European Radiology, 2002, 12 (4) : 951 - 951
  • [43] Simulated bone erosions in a hand phantom:: Detection with conventional screen-film technology versus cesium iodide-amorphous silicon flat-panel detector
    Strotzer, M
    Völk, M
    Wild, T
    von Landenberg, P
    Feuerbach, S
    RADIOLOGY, 2000, 215 (02) : 512 - 515
  • [44] Monte Carlo studies on the amorphous silicon based digital x-ray imagers
    Lee, HK
    Shinn, KS
    Suh, TS
    Choe, BY
    Chung, SK
    Cho, G
    1997 IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, VOLS 1 & 2, 1998, : 1288 - 1292
  • [45] Development and Evaluation of a selenium-based flat-panel digital x-ray detector system based on quality factor
    Park, JK
    Choi, JY
    Kang, SS
    Mun, CW
    Lee, HW
    Nam, SH
    MEDICAL IMAGING 2003: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2003, 5030 : 799 - 809
  • [46] Radiation exposure in full-field digital mammography with a flat-panel x-ray detector based on amorphous silicon in comparison with conventional screen-film mammography.
    Hermann, KP
    Obenauer, S
    Grabbe, E
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2000, 172 (11): : 940 - 945
  • [47] X-ray photoconductors and stabilized a-Se for direct conversion digital flat-panel X-ray image-detectors
    Kasap, SO
    Rowlands, JA
    JOURNAL OF MATERIALS SCIENCE-MATERIALS IN ELECTRONICS, 2000, 11 (03) : 179 - 198
  • [48] Comparison of an amorphous silicon/cesium iodide flat-panel digital chest radiography system with screen/film and computed radiography systems - A contrast-detail phantom study
    Rong, XJJ
    Shaw, CC
    Liu, XM
    Lemacks, MR
    Thompson, SK
    MEDICAL PHYSICS, 2001, 28 (11) : 2328 - 2335
  • [49] Performance of an amorphous silicon flat panel X-ray imaging detector for general radiography
    Granfors, PR
    Aufrichtig, R
    RADIOLOGY, 1999, 213P : 234 - 234
  • [50] Review X-ray photoconductors and stabilized a-Se for direct conversion digital flat-panel X-ray image-detectors
    S. O. Kasap
    J. A. Rowlands
    Journal of Materials Science: Materials in Electronics, 2000, 11 : 179 - 198