Control of centrifugal blood pump based on the motor current

被引:29
|
作者
Iijima, T [1 ]
Inamoto, T [1 ]
Nogawa, M [1 ]
Takatani, S [1 ]
机构
[1] YAMAGATA UNIV,YONEZAWA,YAMAGATA 992,JAPAN
关键词
implantable centrifugal pump; motor current; fast Fourier transform; suction effect; automatic motor speed control;
D O I
10.1111/j.1525-1594.1997.tb03717.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this study, centrifugal pump performance was examined in a mock circulatory loop to derive an automatic pump rotational speed (rpm) control method. The pivot bearing supported sealless centrifugal pump was placed in the left ventricular apex to aorta bypass mode. The pneumatic pulsatile ventricle was used to simulate the natural ventricle. To simulate the suction effect in the ventricle, a collapsible rubber tube was placed in the inflow port of the centrifugal pump in series with the apex of the simulated ventricle. Experimentally, the centrifugal pump speed (rpm) was gradually increased to simulate the suction effect. The pump flow through the centrifugal pump measured by an electromagnetic flowmeter, the aortic pressure, and the motor current were continuously digitized at 100 Hz and stored in a personal computer. The analysis of the cross-spectral density between the pump flow and motor current waveforms revealed that 2 waveforms were highly correlated at the frequency range between 0 and 4 Hz, with the coherence and phase angles being close to 1.0 and 0 degrees, respectively. The fast Fourier transform analysis of the motor current indicated that the second harmonic component of the motor current power density increased with the occurrence of the suction effect in the circuit. The ratio of the fundamental to the second harmonic component decreased less than 1.3 as the suction effect developed in the circuit. It is possible to detect and prevent the suction effect of the centrifugal blood pump in the natural ventricle through analysis of the motor current waveform.
引用
收藏
页码:655 / 660
页数:6
相关论文
共 50 条
  • [41] Modeling Study of an Implantable Centrifugal Blood Pump Actuator with Redundant Sensorless Control
    Leao, Tarcisio F.
    Bock, Eduardo
    Fonseca, Jeison
    Andrade, Aron
    Cavalheiro, Andre
    Uebelhart, Beatriz
    Cardoso, Jose R.
    Chabu, Ivan
    Campo, Alexandre
    2012 44TH SOUTHEASTERN SYMPOSIUM ON SYSTEM THEORY (SSST), 2012, : 174 - 178
  • [42] In vitro evaluation of multi-objective physiological control of the centrifugal blood pump
    Leao, Tarcisio
    Utiyama, Bruno
    Fonseca, Jeison
    Bock, Eduardo
    Andrade, Aron
    ARTIFICIAL ORGANS, 2020, 44 (08) : 785 - 796
  • [43] Modeling and control of a magnetic bearing system for the magnetically suspended centrifugal blood pump
    Kim, H
    Kim, HC
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2000, 23 (10): : 689 - 696
  • [44] Optimization Design of Centrifugal Pump Flow Control System Based on Adaptive Control
    Wang, Yuqin
    Zhang, Haodong
    Han, Zhibo
    Ni, Xiaoqiang
    PROCESSES, 2021, 9 (09)
  • [45] Numerical study of blood pump weaning effects on hemocompatibility of centrifugal blood pump
    Xie N.
    Tang Y.
    Zhang Y.
    Liu Y.
    Beijing Hangkong Hangtian Daxue Xuebao/Journal of Beijing University of Aeronautics and Astronautics, 2023, 49 (07): : 1680 - 1688
  • [46] Vibration control for an implantable blood pump on a bearingless slice motor
    Huettner, C
    JSME INTERNATIONAL JOURNAL SERIES C-MECHANICAL SYSTEMS MACHINE ELEMENTS AND MANUFACTURING, 2003, 46 (03) : 908 - 915
  • [47] A new technique to control brushless motor for blood pump application
    Fonseca, Jeison
    Andrade, Aron
    Nicolosi, Denys E. C.
    Biscegli, Jose F.
    Legendre, Daniel
    Bock, Eduardo
    Lucchi, Julio CeSar
    ARTIFICIAL ORGANS, 2008, 32 (04) : 355 - 359
  • [48] Feedback Control of Superconducting Current Based on Linear-Motor Type Magnetic Flux Pump
    Zhou, Qun
    Huang, Shikang
    Wang, Wei
    Leng, Minrui
    Wu, Chenghuai
    Tang, Fuling
    Liu, Peng
    Zhou, Li
    Li, Yuxiao
    Dian, Yulin
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2024, 34 (05) : 1 - 5
  • [49] A new blood pump for cardiopulmonary bypass: The HiFlow centrifugal pump
    Gobel, C
    Eilers, R
    Reul, H
    Schwindke, P
    Jorger, M
    Rau, G
    ARTIFICIAL ORGANS, 1997, 21 (07) : 841 - 845
  • [50] A new blood pump for cardiopulmonary bypass: The HiFlow centrifugal pump
    Helmholtz-Inst. for Biomed. Eng., Aachen University of Technology, Aachen, Germany
    不详
    ARTIF. ORGANS, 7 (841-845):