摘要
Objectives Multiple studies show orthopedic health problems for medical staff due to wearing radiation protection aprons. The aim of this study was to evaluate the weight pressure on the shoulder as a marker of physical strain caused by different radiation-protection devices. Methods For the weight pressure measurement, a pressure sensor (OMD-30-SE-100N, OptoForce, Budapest, Hungary) placed on the left and right shoulder was used. Wearing different radiation protection systems the force measurement system was used to quantify the weight pressure. Measurements were acquired in still standing position and during various movements. Results A mean significant decreasing weight pressure on the shoulder between 74% and 84% (p<0.001) was measured, when the free-hanging radiation protection system was used in comparison to one-piece and two-piece radiation protection aprons and coats. Using two-piece radiation protection aprons, the weight pressure was significantly lower than that of one-piece radiation protection coats. If a belt was used for the one-piece radiation protection coat, the weight pressure on the shoulder was reduced by 32.5% (p = 0.003). For a two-piece radiation protection apron and a one-piece radiation protection coat (with and without belt) a significant different weight pressure distribution between the right and left shoulder could be measured. Conclusions The free-hanging radiation protection system showed a significant lower weight pressure in comparison to the other radiation protection devices. Apart from this, use of a two-piece radiation protection apron or addition of a belt to a radiation protection coat proved to be further effective options to reduce weight pressure.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] Radiation protection in interventional radiology
    Johnson, DR
    Kyriou, J
    Morton, EJ
    Clifton, A
    Fitzgerald, M
    MacSweeney, E
    [J]. CLINICAL RADIOLOGY, 2001, 56 (02) : 99 - 106
  • [2] Radiation protection in interventional radiology
    Faulkner, K
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1997, 70 (832): : 325 - 326
  • [3] Radiation Protection in Interventional Radiology
    Garg, Tushar
    Shrigiriwar, Apurva
    [J]. INDIAN JOURNAL OF RADIOLOGY AND IMAGING, 2021, 31 (04): : 939 - 945
  • [4] Personal Radiation Protection and Corresponding Dosimetry in Interventional Radiology: An Overview and Future Developments
    Koenig, Alexander Marc
    Etzel, Robin
    Thomas, Rohit Philip
    Mahnken, Andreas H.
    [J]. ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2019, 191 (06): : 512 - 521
  • [5] Occupational orthopedic problems and its relation to personal radiation protection in interventional radiology
    Koenig, A. M.
    Froehlich, L.
    Viniol, S.
    Thomas, R. P.
    Mahnken, A. H.
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2024, 175
  • [6] Radiation protection in interventional radiology.
    Adamus, R.
    Loose, R.
    Wucherer, M.
    Uder, M.
    Galster, M.
    [J]. RADIOLOGE, 2016, 56 (03): : 275 - 281
  • [7] Radiation Protection Tools in Interventional Radiology
    Schueler, Beth A.
    Baiter, Stephen
    Miller, Donald L.
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF RADIOLOGY, 2012, 9 (11) : 844 - 845
  • [8] Radiation Protection and Safety in Interventional Radiology
    Wagner, Jessyca B.
    [J]. RADIOLOGIC TECHNOLOGY, 2020, 91 (05) : 431 - 444
  • [9] Radiation protection of patients in interventional radiology
    Roche, A.
    [J]. JOURNAL DE RADIOLOGIE, 2010, 91 (11): : 1231 - 1235
  • [10] Exoskeleton System for Radiation Protection in Interventional Radiology
    Hosse, Clarissa
    Kolck, Johannes
    Can, Elif
    Fehrenbach, Uli
    Auer, Timo A.
    Pivetta, Fabio
    Collettini, Federico
    Gebauer, Bernhard
    de Bucourt, Maximilian
    [J]. JOURNAL OF RADIATION PROTECTION AND RESEARCH, 2024, 49 (02): : 78 - 84