3-D Vector Flow Estimation With Row-Column-Addressed Arrays

被引:41
|
作者
Holbek, Simon [1 ]
Christiansen, Thomas Lehrmann [2 ]
Stuart, Matthias Bo [1 ]
Beers, Christopher [3 ]
Thomsen, Erik Vilain [2 ]
Jensen, Jorgen Arendt [1 ]
机构
[1] Tech Univ Denmark, Dept Elect Engn, Ctr Fast Ultrasound Imaging, DK-2800 Lyngby, Denmark
[2] Tech Univ Denmark, Dept Micro & Nanotechnol, DK-2800 Lyngby, Denmark
[3] Sound Technol Inc, State Coll, PA 16803 USA
关键词
3-D vector flow imaging; blood flow; medical ultrasound; row-column addressed arrays; transverse oscillation; TRANSVERSE OSCILLATION APPROACH; 3-DIMENSIONAL VELOCITY VECTORS; ARTERY STENOSIS; ULTRASOUND; DOPPLER; TRANSDUCER; DIRECTION;
D O I
10.1109/TUFFC.2016.2582536
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Simulation and experimental results from 3-D vector flow estimations for a 62 + 62 2-D row-column (RC) array with integrated apodization are presented. A method for implementing a 3-D transverse oscillation (TO) velocity estimator on a 3-MHz RC array is developed and validated. First, a parametric simulation study is conducted, where flow direction, ensemble length, number of pulse cycles, steering angles, transmit/ receive apodization, and TO apodization profiles and spacing are varied, to find the optimal parameter configuration. The performance of the estimator is evaluated with respect to relative mean bias (B) over tilde and mean standard deviation (sigma) over tilde. Second, the optimal parameter configuration is implemented on the prototype RC probe connected to the experimental ultrasound scanner SARUS. Results from measurements conducted in a flow-rig system containing a constant laminar flow and a straight-vessel phantom with a pulsating flow are presented. Both an M-mode and a steered transmit sequence are applied. The 3-D vector flow is estimated in the flow rig for four representative flow directions. In the setup with 90 degrees beam-to-flow angle, the relative mean bias across the entire velocity profile is (-4.7, -0.9, 0.4)% with a relative standard deviation of (8.7, 5.1, 0.8)% for (v(x), v(y), v(z)). The estimated peak velocity is 48.5 +/- 3 cm/s giving a -3% bias. The out-of-plane velocity component perpendicular to the cross section is used to estimate volumetric flow rates in the flow rig at a 90 degrees beam-to-flow angle. The estimated mean flow rate in this setup is 91.2 +/- 3.1 L/h corresponding to a bias of -11.1%. In a pulsating flow setup, flow rate measured during five cycles is 2.3 +/- 0.1 mL/stroke giving a negative 9.7% bias. It is concluded that accurate 3-D vector flow estimation can be obtained using a 2-D RC-addressed array.
引用
收藏
页码:1799 / 1814
页数:16
相关论文
共 50 条
  • [1] 3-D Vector Velocity Estimation with Row-Column Addressed Arrays
    Holbek, Simon
    Christiansen, Thomas Lehrmann
    Rasmussen, Morten Fischer
    Stuart, Matthias Bo
    Thomasen, Erik Vilain
    Jensen, Jorgen Arendt
    [J]. 2015 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2015,
  • [2] Experimental 3-D Vector Velocity Estimation with Row-Column Addressed Arrays
    Holbek, Simon
    Stuart, Matthias Bo
    Jensen, Jorgen Arendt
    [J]. 2016 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2016,
  • [3] 3-D Imaging using Row-Column-Addressed 2-D Arrays with a Diverging Lens
    Bouzari, Hamed
    Engholm, Mathias
    Stuart, Matthias Bo
    Nikolov, Svetoslav Ivanov
    Thomsen, Erik Vilain
    Jensen, Jorgen Arendt
    [J]. 2016 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2016,
  • [4] Improved Focusing Method for 3-D Imaging using Row-Column-Addressed 2-D Arrays
    Bouzari, Hamed
    Engholm, Mathias
    Stuart, Matthias Bo
    Thomsen, Erik Vilain
    Jensen, Jorgen Arendt
    [J]. 2017 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2017,
  • [5] 3-D Imaging using Row-Column-Addressed 2-D Arrays with a Diverging Lens: Phantom Study
    Bouzari, Hamed
    Engholm, Mathias
    Beers, Christopher
    Stuart, Matthias Bo
    Nikolov, Svetoslav Ivanov
    Thomsen, Erik Vilain
    Jensen, Jorgen Arendt
    [J]. 2017 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2017,
  • [6] Fourier-Based Fast 3-D Ultrasound Imaging Using Row-Column-Addressed 2-D Arrays
    Lin, Shang-Ching
    Li, Pai-Chi
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2024, 71 (01) : 85 - 101
  • [7] Curvilinear 3-D Imaging Using Row-Column-Addressed 2-D Arrays With a Diverging Lens: Feasibility Study
    Bouzari, Hamed
    Engholm, Mathias
    Beers, Christopher
    Stuart, Matthias Bo
    Nikolov, Svetoslav Ivanov
    Thomsen, Erik Vilain
    Jensen, Jorgen Arendt
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2017, 64 (06) : 978 - 988
  • [8] Curvilinear 3-D Imaging Using Row-Column-Addressed 2-D Arrays With a Diverging Lens: Phantom Study
    Bouzari, Hamed
    Engholm, Mathias
    Beers, Christopher
    Stuart, Matthias Bo
    Nikolov, Svetoslav Ivanov
    Thomsen, Erik Vilain
    Jensen, Jorgen Arendt
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2018, 65 (07) : 1182 - 1192
  • [9] 3-D Vector Flow Using a Row-Column Addressed CMUT Array
    Holbek, Simon
    Christiansen, Thomas Lehrmann
    Engholm, Mathias
    Lei, Anders
    Stuart, Matthias Bo
    Beers, Christopher
    Moesner, Lars Nordahl
    Bagge, Jan Peter
    Thomsen, Erik Vilain
    Jensen, Jurgen Arendt
    [J]. MEDICAL IMAGING 2016: ULTRASONIC IMAGING AND TOMOGRAPHY, 2016, 9790
  • [10] Capacitive Substrate Coupling of Row-Column-Addressed 2-D CMUT Arrays
    Engholm, Mathias
    Bouzari, Hamed
    Jensen, Jorgen Arendt
    Thomsen, Erik Vilain
    [J]. 2016 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2016,