Development and implementation of a remote audit tool for high dose rate (HDR) Ir-192 brachytherapy using optically stimulated luminescence dosimetry

被引:19
|
作者
Casey, Kevin E. [1 ,2 ]
Alvarez, Paola [1 ]
Kry, Stephen F. [1 ,2 ]
Howell, Rebecca M. [1 ,2 ]
Lawyer, Ann [1 ]
Followill, David [1 ,2 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
[2] Univ Texas Grad Sch Biomed Sci Houston, Houston, TX 77030 USA
关键词
HDR; brachytherapy; OSL; audit; RPC; DEPENDENCE; PHOTON; OSL; IRRADIATION; ELECTRON; BEAMS; KERMA; AIR;
D O I
10.1118/1.4824915
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The aim of this work was to create a mailable phantom with measurement accuracy suitable for Radiological Physics Center (RPC) audits of high dose-rate (HDR) brachytherapy sources at institutions participating in National Cancer Institute-funded cooperative clinical trials. Optically stimulated luminescence dosimeters (OSLDs) were chosen as the dosimeter to be used with the phantom. Methods: The authors designed and built an 8 x 8 x 10 cm(3) prototype phantom that had two slots capable of holding Al2O3:C OSLDs (nanoDots; Landauer, Glenwood, IL) and a single channel capable of accepting all Ir-192 HDR brachytherapy sources in current clinical use in the United States. The authors irradiated the phantom with Nucletron and Varian Ir-192 HDR sources in order to determine correction factors for linearity with dose and the combined effects of irradiation energy and phantom characteristics. The phantom was then sent to eight institutions which volunteered to perform trial remote audits. Results: The linearity correction factor was k(L) = (-9.43 x 10(-5) x dose) + 1.009, where dose is in cGy, which differed from that determined by the RPC for the same batch of dosimeters using Co-60 irradiation. Separate block correction factors were determined for current versions of both Nucletron and Varian Ir-192 HDR sources and these vendor-specific correction factors differed by almost 2.6%. For the Nucletron source, the correction factor was 1.026 [95% confidence interval (CI) = 1.023-1.028], and for the Varian source, it was 1.000 (95% CI = 0.995-1.005). Variations in lateral source positioning up to 0.8 mm and distal/proximal source positioning up to 10 mm had minimal effect on dose measurement accuracy. The overall dose measurement uncertainty of the system was estimated to be 2.4% and 2.5% for the Nucletron and Varian sources, respectively (95% CI). This uncertainty was sufficient to establish a +/- 5% acceptance criterion for source strength audits under a formal RPC audit program. Trial audits of four Nucletron sources and four Varian sources revealed an average RPC-to-institution dose ratio of 1.000 (standard deviation = 0.011). Conclusions: The authors have created an OSLD-based Ir-192 HDR brachytherapy source remote audit tool which offers sufficient dose measurement accuracy to allow the RPC to establish a remote audit program with a 5% acceptance criterion. The feasibility of the system has been demonstrated with eight trial audits to date. (C) 2013 American Association of Physicists in Medicine.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Development and Implementation of a Remote Audit Tool for High Dose Rate (HDR) Ir-192 Brachytherapy Using Optically Stimulated Luminescence Dosimetry
    Casey, K.
    Alvarez, P.
    Lawyer, A.
    Kry, S.
    Howell, R.
    Davidson, S.
    Followill, D.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3721 - 3721
  • [2] HIGH-DOSE RATE (HDR) IR-192 BRACHYTHERAPY APPLICATION OF BASAL CELL CARCINOMA
    Basaric, Borko
    Petrovic, Borislava
    Teodorovic, Milan
    Baucal, Milutin
    Rutonjski, Laza
    Cudic, Ozren
    Djuran, Branislav
    Mitric-Askovic, Milana
    [J]. RAD 2015: THE THIRD INTERNATIONAL CONFERENCE ON RADIATION AND APPLICATIONS IN VARIOUS FIELDS OF RESEARCH, 2015, : 65 - 68
  • [3] The Safety Aspect Design for remote Afterloading High Dose Rate Brachytherapy Ir-192
    Isnaini, Ismet
    Triyanto, Joko
    Sukandar
    [J]. 3RD INTERNATIONAL CONFERENCE ON NUCLEAR ENERGY TECHNOLOGIES AND SCIENCES (ICONETS) 2019, 2019, 2180
  • [4] Development of An Absorbed Dose to Water Primary Standard for HDR Ir-192 Brachytherapy Based On the Fricke Dosimetry System
    El Gamal, I.
    Cojocaru, C.
    Mainegra-Hing, E.
    Ross, C.
    McEwen, M.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [5] Comparative dosimetry of GammaMed Plus high-dose rate Ir-192 brachytherapy source
    Patel, N. P.
    Majumdar, B.
    Vijayan, V.
    [J]. JOURNAL OF MEDICAL PHYSICS, 2010, 35 (03) : 137 - 143
  • [6] Design and implementation of a film dosimetry audit tool for comparison of planned and delivered dose distributions in high dose rate (HDR) brachytherapy
    Palmer, Antony L.
    Lee, Chris
    Ratcliffe, Ailsa J.
    Bradley, David
    Nisbet, Andrew
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (19): : 6623 - 6640
  • [7] Transit dose of an Ir-192 high dose rate brachytherapy stepping source
    Wong, TPY
    Fernando, W
    Johnston, PN
    Bubb, IF
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (02): : 323 - 331
  • [8] TLD, diode and monte carlo dosimetry of an Ir-192 source for high dose-rate brachytherapy
    Kirov, AS
    Williamson, JF
    Meigooni, AS
    Zhu, Y
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (12): : 2015 - 2036
  • [9] Optically stimulated luminescent dosimetry for high dose rate brachytherapy
    Tien, Christopher Jason
    Ebeling, Robert, III
    Hiatt, Jessica R.
    Curran, Bruce
    Sternick, Edward
    [J]. FRONTIERS IN ONCOLOGY, 2012, 2 : 1 - 7
  • [10] High dose dosimetry using optically stimulated luminescence
    Miller, SD
    [J]. RADIATION PROTECTION DOSIMETRY, 1996, 66 (1-4) : 201 - 204