Accuracy and Reliability of Length Measurements on Three-Dimensional Computed Tomography Using Open-Source OsiriX Software

被引:88
|
作者
Kim, Gihyeon [2 ]
Jung, Ho-Joong [1 ]
Lee, Han-Jun [1 ]
Lee, Jae-Sung [1 ]
Koo, Seungbum [3 ]
Chang, Seung-Hwan [3 ]
机构
[1] Chung Ang Univ, Coll Med, Dept Orthoped Surg, Seoul 140757, South Korea
[2] Chung Ang Univ, Coll Med, Dept Radiol, Seoul 140757, South Korea
[3] Chung Ang Univ, Sch Mech Engn, Seoul 140757, South Korea
基金
新加坡国家研究基金会;
关键词
Computed tomography; Accuracy; Reliability; Length; OsiriX; RECONSTRUCTION; TUNNELS;
D O I
10.1007/s10278-012-9458-6
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
There is a growing interest in three-dimensional computed tomography (3D-CT) as a research tool for the study of bone, joint anatomy, and kinematics. However, when CT data are processed and handled manually using image processing programs to yield 3D image and coordinate value, systematic and random errors should be validated. We evaluated the accuracy and reliability of length measurement on CT with OsiriX software. 3D-CT scans were made of 14 frozen pig knees with five transosseous holes in the metaphyseal portion of femur. The lengths between tunnel orifices were measured using Mitutoyo Digimatic digital calipers to establish the gold standard, and with the OsiriX program in 3D multi-planar reformatting mode for comparison. All measurements were recorded by a principal (replicate 1, trial 1) and a secondary observer (replicate 2, trial 1) and were repeated once by each observer (trial 2). The mean differences between OsiriX and real measurements were less than 0.1 mm in both replicates, and maximum differences were less than 0.3 mm. There were no significant differences between the replicates and real measurements (p = 0.544 and 0.622 for replicates 1 and 2, respectively). The intraclass correlation coefficients (ICC) were very high between trials and between replicates (ICC = 0.998 and 0.999, respectively). For kinematic analysis of the knees, length measurements on 3D-CT using OsiriX program can be used as alternatives to real measurements with less than 0.3-mm accuracy and very high reliability.
引用
收藏
页码:486 / 491
页数:6
相关论文
共 50 条
  • [31] Quantitative diffusion measurements using the open-source software PyFRAP
    Alexander Bläßle
    Gary Soh
    Theresa Braun
    David Mörsdorf
    Hannes Preiß
    Ben M. Jordan
    Patrick Müller
    Nature Communications, 9
  • [32] Bone mineral density measurements in postmortem computed tomography: An open-source approach
    Hansen, Kasper
    Vinther, Dennis
    Boel, Lene Warner Thorup
    Agacan, Guelislam
    Thygesen, Jesper
    Uhrenholt, Lars
    Lauridsen, Henrik
    FORENSIC IMAGING, 2024, 37
  • [33] Open-Source Three-Dimensional Printable Infant Clubfoot Brace
    Savonen, Benjamin
    Gershenson, John
    Bow, Jennifer K.
    Pearce, Joshua M.
    JOURNAL OF PROSTHETICS AND ORTHOTICS, 2020, 32 (02) : 149 - 158
  • [34] SOWOS: an open-source program for the three-dimensional Wulff construction
    Scopece, Daniele
    JOURNAL OF APPLIED CRYSTALLOGRAPHY, 2013, 46 : 811 - 816
  • [35] SOWOS: An open-source program for the three-dimensional Wulff construction
    Department of Materials Science, University of Milano-Bicocca, via R. Cozzi 53, I-20125 Milano, Italy
    不详
    J. Appl. Crystallog., 2013, 3 (811-816):
  • [36] Reliability of Three-Dimensional Micro-Computed Tomography Root Canal Analysis Using Drishti Software: A Pilot Study
    Iwani, Samsudin Nurulaqmar
    Kamaruzaman, Marlena
    Jawami, Afiq Azizi
    JOURNAL OF DENTISTRY INDONESIA, 2024, 31 (01)
  • [37] Landmark precision and reliability and accuracy of linear distances estimated by using 3D computed micro-tomography and the open-source TINA Manual Landmarking Tool software
    Simon, Monique Nouailhetas
    Marroig, Gabriel
    FRONTIERS IN ZOOLOGY, 2015, 12
  • [38] A Semi-Automatic Method to Create an Affordable Three-Dimensional Printed Splint Using Open-Source and Free Software
    Wang, Zhujiang
    Dubrowski, Adam
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2021, 13 (03)
  • [39] Landmark precision and reliability and accuracy of linear distances estimated by using 3D computed micro-tomography and the open-source TINA Manual Landmarking Tool software
    Monique Nouailhetas Simon
    Gabriel Marroig
    Frontiers in Zoology, 12
  • [40] Improving Access to Intracranial Volume Measurements for Clinical Use: Open-Source Method using Computed Tomography
    Surapaneni, Akhil
    Frommer, Sarah A.
    Harshbarger, Raymond
    FACE, 2024, 5 (04): : 579 - 585