Trans-pulmonary echocardiography as a guide for device closure of patent ductus arteriosus

被引:4
|
作者
Kudo, Yoshiyuki [1 ]
Suda, Kenji [1 ]
Yoshimoto, Hironaga [1 ]
Teramachi, Yozo [2 ]
Kishimoto, Shintaro [1 ]
Iemura, Motofumi [2 ]
Matsuishi, Toyojiro [1 ]
机构
[1] Kurume Univ, Sch Med, Dept Pediat & Child Hlth, Kurume, Fukuoka 8300011, Japan
[2] St Marys Hosp, Dept Pediat Cardiol, Kurume, Fukuoka, Japan
关键词
intracardiac echocardiography; Amplatzer duct occluder; aortography; monitoring tool; residual shunt; 3-DIMENSIONAL TRANSESOPHAGEAL ECHOCARDIOGRAPHY; ATRIAL SEPTAL-DEFECTS; INTRACARDIAC ECHOCARDIOGRAPHY; TRANSCATHETER CLOSURE; VALVE-REPLACEMENT; GUIDANCE;
D O I
10.1002/ccd.25879
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
ObjectivesThe aim of this study was to develop trans-pulmonary echocardiography (TPE) to guide device closure of patent ductus arteriosus (DC-PDA). BackgroundAortography requires a large amount of contrast yet may give us an inadequate image to evaluate anatomy or residual shunt in patients with large PDA or dilated vessels and is precluded in patients with renal dysfunction. Practically, there is no imaging modality to monitor the entire procedure except for trans-esophageal echocardiography that requires general anesthesia. MethodsSubjects were seven patients with ages ranged from 6- to 77-years old and body weight>15 kg. The size of the PDA ranged from 1.8 to 6.3 mm with pulmonary to systemic flow ratios from 1.2 to 2.2. During DC-PDA using Ampaltzer Duct Occluder or coil, an intra-cardiac echocardiographic (ICE) catheter was advanced into pulmonary arteries and standard views were developed to guide DC-PDA. ResultsWe have developed two standard views; the main pulmonary artery view (MPA view) and the left pulmonary artery view (LPA view). The MPA view provided aortic short axis view equivalent to that seen by trans-thoracic echocardiography in children. The LPA view, obtained by the echo probe in the LPA and turned it up upside down, provided long axis view of the PDA allowing more precise anatomical evaluation. TPE allowed us to monitor the entire procedure and determine residual shunts. ConclusionsTPE in the MPA and LPA view can be an effective guide for DC-PDA. This report leads to new application of this imaging device. (c) 2015 Wiley Periodicals, Inc.
引用
收藏
页码:264 / 270
页数:7
相关论文
共 50 条
  • [1] Intracardiac Echocardiography as a Guide for Transcatheter Closure of Patent Ductus Arteriosus
    Yoshimoto, Hironaga
    Yasuto, Maeda
    Inoue, Tadashi
    Kagiyama, Yoshiyuki
    Teramachi, Yozo
    Takase, Ryuta
    Koteda, Yusuke
    Fukumoto, Yoshihiro
    Iemura, Motofumi
    Suda, Kenji
    [J]. JOURNAL OF INTERVENTIONAL CARDIOLOGY, 2020, 2020
  • [2] Device closure of patent ductus arteriosus with severe pulmonary arterial hypertension
    Roy, A
    Juneja, R
    Kothari, SS
    Naik, N
    Saxena, A
    [J]. EUROPEAN HEART JOURNAL, 2004, 25 : 228 - 228
  • [3] Device Closure of Patent Ductus Arteriosus in Adults
    O'Byrne, Michael L.
    Smith, Christopher L.
    Gillespie, Matthew J.
    [J]. CANADIAN JOURNAL OF CARDIOLOGY, 2020, 36 (06) : 795 - 796
  • [4] Transcatheter Device Closure of Patent Ductus Arteriosus
    Sultan, Mehboob
    Ullah, Maad
    Sadiq, Nadeem
    Akhtar, Khurram
    Akbar, Hajira
    [J]. JCPSP-JOURNAL OF THE COLLEGE OF PHYSICIANS AND SURGEONS PAKISTAN, 2014, 24 (10): : 710 - 713
  • [5] Transcatheter closure of very large patent ductus arteriosus with the modified patent ductus arteriosus closure device.
    Zhang, F.
    Ge, J.
    Zhou, D.
    Guan, L.
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 2006, 97 (8B): : 50D - 50D
  • [6] Echocardiography for Patent Ductus Arteriosus Including Closure in Adults
    Chugh, Reema
    Salem, Morris M.
    [J]. ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES, 2015, 32 : 17 - 31
  • [7] Intracardiac echocardiography: use during transcatheter device closure of a patent ductus arteriosus in a dog
    Chetboul, V.
    Damoiseaux, C.
    Behr, L.
    Morlet, A.
    Moise, N. S.
    Gouni, V.
    Lavennes, M.
    Pouchelon, J. -L.
    Laborde, F.
    Borenstein, N.
    [J]. JOURNAL OF VETERINARY CARDIOLOGY, 2017, 19 (03) : 293 - 298
  • [8] Trans-Catheter Closure of Patent Ductus Arteriosus-What Is the Best Device?
    Ghasemi, Abdolrahim
    Pandya, Swati
    Reddy, Surendranath Veeram
    Turner, Daniel R.
    Du, Wei
    Navabi, Mohamad Ali
    Mirzaaghayan, Mohammad Reza
    Kiani, Abdolrazagh
    Sloan, Katherine
    Forbes, Thomas J.
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2010, 76 (05) : 687 - 695
  • [9] CLOSURE OF PATENT DUCTUS ARTERIOSUS
    不详
    [J]. POSTGRADUATE MEDICINE, 1971, 50 (04) : 24 - +
  • [10] Evaluation of pulmonary flow by Doppler echocardiography and scintigraphy after percutaneous closure of patent ductus arteriosus
    Soares, AM
    Andrade, JL
    Leal, SM
    Soares, J
    Kajita, L
    Mathias, W
    Lopes, AA
    Aielo, V
    [J]. EUROPEAN HEART JOURNAL, 2003, 24 : 213 - 213