The accuracy of a glucose-oxidase-based point-of-care glucometer in premature infants

被引:3
|
作者
Tsao, Lon-Yen [1 ,4 ]
Chang, Ming-Yeh [2 ]
Hsiao, Chien-Chou [1 ,3 ,4 ]
机构
[1] Changhua Christian Hosp, Dept Neonatol, Changhua 500, Taiwan
[2] Changhua Christian Hosp, Dept Pediat Neurol, Changhua 500, Taiwan
[3] Chung Shan Med Univ, Sch Med, Taichung, Taiwan
[4] Changhua Christian Hosp, Clin Nutr Support Serv Team, Changhua 500, Taiwan
关键词
Neonatology; Paediatric Practice; Intensive Care; Measurement; BLOOD-GLUCOSE;
D O I
10.1136/archdischild-2012-303613
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Objective The purpose of this study was to evaluate the accuracy of a novel commercial glucose-oxidase (GO)-based glucometer in measuring glucose concentrations in arterial blood samples collected from premature infants. Patients and methods A commercial glucometer, which is a plasma separation GO-based glucometer, was employed to compare the accuracy of the commercial blood glucometer with that of a standard glucose analyser for monitoring blood glucose levels in premature infants. Arterial blood samples were collected and analysed, including samples adjusted for specific glucose (3.6mmol/L, 8.3mmol/L and 16.7mmol/L) and PaO2 levels (range, 40-400mmHg). Results In total, 159 samples were collected. Bland-Altman analysis showed good correlation between the commercial glucometer and standard glucose analyser. Error-grid analysis revealed that all of the results were within zone A (ie, the clinically accurate estimate zone). The biases between the two systems were low at different PaO2 levels and haematocrits. Finally, the influence of different PaO2 levels was within acceptable ranges. Conclusions The GO-based glucometer evaluated in this study provides accurate results even when measured at high PaO2 and different haematocrits and is suitable for measuring arterial glucose levels in premature infants.
引用
收藏
页码:F545 / F548
页数:4
相关论文
共 50 条
  • [21] A veterinary-calibrated point-of-care glucometer accurately measures blood glucose concentration in dogs and cats
    Cammarano, Kellie
    Carney, Patrick C.
    Rose, Hailey E.
    Merkhassine, Michael
    Ippolito, Elizabeth
    Arsenault, Anne
    Plotsker, Noah M.
    Loftus, John P.
    Kowalczyk, Marcin
    Prieto, Jennifer M. B.
    AMERICAN JOURNAL OF VETERINARY RESEARCH, 2024, 85 (09)
  • [22] Electrochemical immunosensor for point-of-care quantitative detection of tumor markers based on personal glucometer
    Xu, Yang
    Wang, Huiqiang
    JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 2022, 915
  • [23] Accuracy of point-of-care measurements
    Jorens, Philippe G.
    CANADIAN MEDICAL ASSOCIATION JOURNAL, 2007, 177 (09) : 1070 - 1070
  • [24] Impact of Cyanokit Administration on Point-of-Care Glucose and Laboratory Test Accuracy
    Sauer, Madeline A.
    Rizzo, Natalie M.
    Tolan, Nicole, V
    CLINICAL CHEMISTRY, 2022, 68 (04) : 504 - 508
  • [25] Recent advances in glucose-oxidase-based nanocomposites for diabetes diagnosis and treatment
    Yang, Dejun
    Cai, Chunyan
    Liu, Kai
    Peng, Zhaolei
    Yan, Chunmei
    Xi, Jingjing
    Xie, Fan
    Li, Xiaofang
    JOURNAL OF MATERIALS CHEMISTRY B, 2023, 11 (32) : 7582 - 7608
  • [26] Precision and accuracy of a point of care glucometer for detection of hypoglycaemia in horses
    Hughes, Kristopher
    Moore, Claire
    Woods, Sophie
    Wilkes, Edwina
    VETERINARY JOURNAL, 2024, 305
  • [27] Accuracy and precision of point-of-care testing for glucose and prothrombin time at the critical care units
    Yuoh, C
    Elghetany, MT
    Petersen, JR
    Mohammad, A
    Okorodudu, AO
    CLINICA CHIMICA ACTA, 2001, 307 (1-2) : 119 - 123
  • [28] A Comparison of Venous versus Capillary Blood Samples when Measuring Blood Glucose Using a Point-of-Care, Capillary-Based Glucometer
    Topping, Jessica
    Reardon, Matthew
    Coleman, Jake
    Hunter, Brian
    Shojima-Perera, Haruka
    Thyer, Liz
    Simpson, Paul
    PREHOSPITAL AND DISASTER MEDICINE, 2019, 34 (05) : 506 - 509
  • [29] Point of care glucometer vs serum glucose in neonatal hypoglycemia
    Fujii, AM
    Camara, P
    Fleck, P
    Maloney, J
    PEDIATRIC RESEARCH, 2003, 53 (04) : 369A - 369A
  • [30] ACCURACY OF POINT OF CARE BLOOD GLUCOSE MEASUREMENTS IN HEALTHY NEWBORN INFANTS
    Zayek, A. J.
    Bhat, R.
    Zayek, M.
    JOURNAL OF INVESTIGATIVE MEDICINE, 2019, 67 (02) : 538 - 538