New weighting factor of weighted CTDI equation for PMMA phantom diameter from 8 to 40 cm: A Monte Carlo study

被引:6
|
作者
Haba, Tomonobu [1 ]
Koyama, Shuji [2 ]
Kinomura, Yutaka [1 ]
Ida, Yoshihiro [1 ]
Kobayashi, Masanao [3 ]
机构
[1] Fujita Hlth Univ Hosp, Dept Radiol, 1-98 Dengakugakubo,Kutsukake Cho, Toyoake, Aichi 4701192, Japan
[2] Nagoya Univ, Brain & Mind Res Ctr, Higashi Ku, 1-1-20 Daiko Minami, Nagoya, Aichi 4618673, Japan
[3] Fujita Hlth Univ, Fac Radiol Technol, 1-98 Dengakugakubo,Kutsukake Cho, Toyoake, Aichi 4701192, Japan
关键词
dosimetry; Monte Carlo simulation; weighted CTDI; x-ray CT; COMPUTED-TOMOGRAPHY; DOSIMETRY; SIMULATIONS; MODELS;
D O I
10.1002/mp.12601
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The weighted computed tomography dose index (CTDIw) uses measured CTDI values at the center and periphery of a cylindrical phantom. The CTDIw value is calculated using conventional, Bakalyar's, and Choi's weighting factors. However, these weighting factors were produced from only 16- and 32-cm-diameter cylindrical phantoms. This study aims to devise new weighting factors to provide more accurate average dose in the central cross-sectional plane of cylindrical phantoms over a wide range of object diameters, by using Monte Carlo simulations. Methods: Simulations were performed by modeling a Toshiba Aquilion ONE CT scanner, in order to compute the cross-sectional dose profiles of polymethyl methacrylate (PMMA) cylindrical phantoms of each diameter (8-40 cm at 4-cm steps), for various tube voltages and longitudinal beam widths. Two phantom models were simulated, corresponding to the CTDI100 method and the method recommended by American Association of Physicists in Medicine (AAPM) task group 111. The dose-computation PMMA cylinders of 1 mm diameter were located between the phantom surfaces and the centers at intervals of 1 mm, from which cross-sectional dose profiles were calculated. By using linear least-squares fits to the obtained cross-sectional dose profiles data, we determined new weighting factors to estimate more accurate average doses in the PMMA cylindrical phantoms by using the CTDIw equation: CTDIw = W-center . CTDIcenter + W-periphery . CTDIperiphery. In order to demonstrate the validity of the devised new weighting factors, the percentage difference between average dose and CTDIw value was evaluated for the weighting factors (conventional, Bakalyar's, Choi's, and devised new weighting factors) in each calculated cross-sectional dose profile. Results: With the use of linear least-squares techniques, new weighting factors (W-center = 3/8 and W-periphery = 5/8 where W-center and W-periphery are weighting factors for CTDIcenter and CTDIperiphery) were determined. The maximum percentage differences between average dose and CTDIw value were 16, -12, -8, and -6% for the conventional, Bakalyar's, Choi's, and devised new weighting factors, respectively. Conclusions: We devised new weighting factors (W-center = 3/8 and W-periphery = 5/8) to provide more accurate average dose estimation in PMMA cylindrical phantoms over a wide range of diameter. The CTDIw equation with devised new weighting factors could estimate average dose in PMMA cylindrical phantoms with a maximum difference of -6%. The results of this study can estimate the average dose in PMMA cylindrical phantoms more accurately than the conventional weighting factors (W-center = 1/3 and W-periphery = 2/3). (C) 2017 American Association of Physicists in Medicine.
引用
收藏
页码:6603 / 6609
页数:7
相关论文
共 2 条
  • [1] Weighted CTDI Equation for 3D Rotational Angiography: A Monte Carlo Study
    Azzi, A.
    Hidayat, R.
    Rosa, A.
    Lubis, L. E.
    ATOM INDONESIA, 2024, 50 (01) : 43 - 48
  • [2] Emission of Secondary Particles From a PMMA Phantom During Proton Irradiation: A Simulation Study with the Geant4 Monte Carlo Toolkit
    Lau, A.
    Chen, Y.
    Ahmad, S.
    MEDICAL PHYSICS, 2014, 41 (06) : 346 - 346