Studying of characteristics of the HPGe detector for radioactivity measurements

被引:7
|
作者
Demir, D. [1 ]
Eroglu, M. [1 ]
Tursucu, A. [1 ]
机构
[1] Ataturk Univ, Fac Sci, Dept Phys, TR-25240 Erzurum, Turkey
来源
关键词
Gamma detectors (scintillators; CZT; HPG; HgI etc); X-ray fluorescence (XRF) systems; EFFICIENCY; LIMITS;
D O I
10.1088/1748-0221/8/10/P10027
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The photopeak efficiency, the peak to valley ratio, the energy resolution and the minimum detectable activity of a HPGe detector were determined experimentally for 13.81-1212.95 keV photon energy by using Am-241, Ba-133, Cs-137 and Eu-152. The photopeak efficiency was studied at the geometry with the source to detector distance of 5 cm, 10 cm, 15 cm and 20 cm. Also, it was determined theoretically. It is shown that full energy peak efficiency decreases with increases in gamma ray energy. Also, the HPGe detector offers very good resolution. This makes it a suitable detector for x- and gamma ray spectroscopy and health physics research.
引用
收藏
页数:12
相关论文
共 50 条
  • [1] Determination of the HPGe detector efficiency in measurements of radioactivity in extended environmental samples
    Montalvan Olivares, Diango M.
    Manso Guevara, M. V.
    Velasco, Fermin G.
    [J]. APPLIED RADIATION AND ISOTOPES, 2017, 130 : 34 - 42
  • [2] Evaluation of HPGe Detector Long Term Performance for Environmental Radioactivity Measurements
    Abdel Aziz, Dalal
    Shabasy, A. M.
    Hassan, H. A.
    El-Meligy, Wafaa
    Diab, H. M.
    [J]. EGYPTIAN JOURNAL OF CHEMISTRY, 2022, 65 (13): : 699 - 705
  • [3] Calibration and Performance of HPGe Detector for Environmental Radioactivity Measurements using LabSOCS
    Shyti, Manjola
    [J]. 10TH JUBILEE CONFERENCE OF THE BALKAN PHYSICAL UNION, 2019, 2075
  • [4] A low-energy HPGE detector dedicated to radioactivity measurements far below environmental levels
    Mouchel, D
    Wordel, R
    [J]. APPLIED RADIATION AND ISOTOPES, 1996, 47 (9-10) : 1033 - 1041
  • [5] Measurement of environmental radioactivity for RAON using HPGe detector and LSC
    Gil, Danhye
    Baek, Beomyeol
    Kim, Jongwook
    Jo, Kyungha
    Yang, Keunsoo
    Hong, Inseok
    [J]. JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2024, 85 (03) : 230 - 234
  • [6] Variation of the Detection Efficiency of a HPGe Detector with the Density of the Sample in the Radioactivity Analysis
    Seo, Bum-Kyoung
    Lee, Kil-Yong
    Yoon, Yoon-Yeol
    Jung, Ki-Jung
    Oh, Won-Zin
    Lee, Kune-Woo
    [J]. ANALYTICAL SCIENCE AND TECHNOLOGY, 2005, 18 (01): : 59 - 65
  • [7] Analytical approach for radioactivity correlation of disc sources with HPGe detector efficiency
    Challan, Mohsen B.
    El-Taher, A.
    [J]. APPLIED RADIATION AND ISOTOPES, 2014, 85 : 23 - 27
  • [8] Performance of CdTe, HPGe and NaI(Tl) detectors for radioactivity measurements
    Perez-Andujar, A
    Pibida, L
    [J]. APPLIED RADIATION AND ISOTOPES, 2004, 60 (01) : 41 - 47
  • [9] The BiPo detector for ultralow radioactivity measurements
    Bongrand, Mathieu
    [J]. TOPICAL WORKSHOP ON LOW RADIOACTIVITY TECHNIQUES - LRT 2010, 2011, 1338 : 49 - 58
  • [10] The BiPo detector for ultralow radioactivity measurements
    Bongrand, M.
    [J]. Topical Worshop on Low Radioactivity Techniques, 2007, 897 : 14 - 19