Filtered Back Projection, Adaptive Statistical Iterative Reconstruction, and a Model-based Iterative Reconstruction in Abdominal CT: An Experimental Clinical Study

被引:207
|
作者
Deak, Zsuzsanna [1 ]
Grimm, Jochen M. [1 ]
Treitl, Marcus [1 ]
Geyer, Lucas L. [1 ]
Linsenmaier, Ulrich [1 ]
Koerner, Markus [1 ]
Reiser, Maximilian F. [1 ]
Wirth, Stefan [1 ]
机构
[1] Univ Munich, Dept Clin Radiol, D-80336 Munich, Germany
关键词
IMAGE-QUALITY; DOSE REDUCTION; ALGORITHM; OPTIMIZATION; ANGIOGRAPHY; PERFORMANCE; RISKS; MDCT;
D O I
10.1148/radiol.12112707
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To compare objective and subjective image quality parameters of three image reconstruction algorithms of different generations at routine multidetector computed tomographic (CT) examinations of the abdomen. Materials and Methods: This institutional review board-approved study included 22 consecutive patients (mean age, 56.1 years +/- 15.8 [standard deviation]; mean weight, 79.1 kg +/- 14.8) who underwent routine CT examinations of the abdomen. A low-contrast phantom was used for objective quality control. Raw data sets were reconstructed by using filtered back projection (FPB), adaptive statistical iterative reconstruction (ASIR), and a model-based iterative reconstruction (MBIR). Radiologists used a semiquantitative scale (-3 to +3) to rate subjective image quality and artifacts, comparing both FBP and MBIR images with ASIR images. The Wilcoxon test and the intraclass correlation coefficient were used to evaluate the data. Measurements of objective noise and CT numbers of soft tissue structures were compared with analysis of variance. Results: The phantom study revealed an improved detectability of low-contrast targets for MBIR compared with ASIR or FBP. Subjective ratings showed higher image quality for MBIR, with better resolution (median value, 2; range, 1 to 3), lower noise (2; range, 1 to 3), and finer contours (2; range, 1 to 2) compared with ASIR (all P,.001). FBP performed inferiorly (0, range, 22 to 0]; -1 [range, -3 to 0]; 0 [range, 21 to 0], respectively; all, P,.001). Mean interobserver correlation was 0.9 for image perception and 0.7 for artifacts. Objective noise for FBP was 14%-68% higher and for MBIR was 18%-47% lower than that for ASIR (P < .001). Conclusion: The MBIR algorithm considerably improved objective and subjective image quality parameters of routine abdominal multidetector CT images compared with those of ASIR and FBP. (C) RSNA, 2012
引用
收藏
页码:197 / 206
页数:10
相关论文
共 50 条
  • [1] Model-based Iterative Reconstruction Compared to Adaptive Statistical Iterative Reconstruction and Filtered Back-projection in CT of the Kidneys and the Adjacent Retroperitoneum
    Olcott, Eric W.
    Shin, Lewis K.
    Sommer, Graham
    Chan, Ian
    Rosenberg, Jarrett
    Molvin, F. Lior
    Boas, F. Edward
    Fleischmann, Dominik
    [J]. ACADEMIC RADIOLOGY, 2014, 21 (06) : 774 - 784
  • [2] Abdominal CT: Comparison of Adaptive Statistical Iterative and Filtered Back Projection Reconstruction Techniques
    Singh, Sarabjeet
    Kalra, Mannudeep K.
    Hsieh, Jiang
    Licato, Paul E.
    Do, Synho
    Pien, Homer H.
    Blake, Michael A.
    [J]. RADIOLOGY, 2010, 257 (02) : 373 - 383
  • [3] Comparison of the image qualities of filtered back-projection, adaptive statistical iterative reconstruction, and model-based iterative reconstruction for CT venography at 80 kVp
    Kim, Jin Hyeok
    Choo, Ki Seok
    Moon, Tae Yong
    Lee, Jun Woo
    Jeon, Ung Bae
    Kim, Tae Un
    Hwang, Jae Yeon
    Yun, Myeong-Ja
    Jeong, Dong Wook
    Lim, Soo Jin
    [J]. EUROPEAN RADIOLOGY, 2016, 26 (07) : 2055 - 2063
  • [4] Comparison of the image qualities of filtered back-projection, adaptive statistical iterative reconstruction, and model-based iterative reconstruction for CT venography at 80 kVp
    Jin Hyeok Kim
    Ki Seok Choo
    Tae Yong Moon
    Jun Woo Lee
    Ung Bae Jeon
    Tae Un Kim
    Jae Yeon Hwang
    Myeong-Ja Yun
    Dong Wook Jeong
    Soo Jin Lim
    [J]. European Radiology, 2016, 26 : 2055 - 2063
  • [5] Assessment of Filtered Back Projection, Adaptive Statistical, and Model-Based Iterative Reconstruction for Reduced Dose Abdominal Computed Tomography
    Padole, Atul
    Singh, Sarabjeet
    Lira, Diego
    Blake, Michael A.
    Pourjabbar, Sarvenaz
    Khawaja, Ranish Deedar Ali
    Choy, Garry
    Saini, Sanjay
    Do, Synho
    Kalra, Mannudeep K.
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2015, 39 (04) : 462 - 467
  • [6] Quantitative analysis of emphysema and airway measurements according to iterative reconstruction algorithms: comparison of filtered back projection, adaptive statistical iterative reconstruction and model-based iterative reconstruction
    Ji Yung Choo
    Jin Mo Goo
    Chang Hyun Lee
    Chang Min Park
    Sang Joon Park
    Mi-Suk Shim
    [J]. European Radiology, 2014, 24 : 799 - 806
  • [7] Quantitative analysis of emphysema and airway measurements according to iterative reconstruction algorithms: comparison of filtered back projection, adaptive statistical iterative reconstruction and model-based iterative reconstruction
    Choo, Ji Yung
    Goo, Jin Mo
    Lee, Chang Hyun
    Park, Chang Min
    Park, Sang Joon
    Shim, Mi-Suk
    [J]. EUROPEAN RADIOLOGY, 2014, 24 (04) : 799 - 806
  • [8] Comparison of image quality from filtered back projection, statistical iterative reconstruction, and model-based iterative reconstruction algorithms in abdominal computed tomography
    Kuo, Yu
    Lin, Yi-Yang
    Lee, Rheun-Chuan
    Lin, Chung-Jung
    Chiou, Yi-You
    Guo, Wan-Yuo
    [J]. MEDICINE, 2016, 95 (31)
  • [9] Computed Tomography Image Quality Evaluation of a New Iterative Reconstruction Algorithm in the Abdomen (Adaptive Statistical Iterative Reconstruction-V) a Comparison With Model-Based Iterative Reconstruction, Adaptive Statistical Iterative Reconstruction, and Filtered Back Projection Reconstructions
    Goodenberger, Martin H.
    Wagner-Bartak, Nicolaus A.
    Gupta, Shiva
    Liu, Xinming
    Yap, Ramon Q.
    Sun, Jia
    Tamm, Eric P.
    Jensen, Corey T.
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2018, 42 (02) : 184 - 190
  • [10] Image quality of CT angiography with model-based iterative reconstruction in young children with congenital heart disease: comparison with filtered back projection and adaptive statistical iterative reconstruction
    Son, Sung Sil
    Choo, Ki Seok
    Jeon, Ung Bae
    Jeon, Gye Rok
    Nam, Kyung Jin
    Kim, Tae Un
    Yeom, Jeong A.
    Hwang, Jae Yeon
    Jeong, Dong Wook
    Lim, Soo Jin
    [J]. INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2015, 31 : 31 - 38