Intraoperative Cone Beam-Computed Tomography With Navigation (O-ARM) Versus Conventional Fluoroscopy (C-ARM)

被引:106
|
作者
Tabaraee, Ehsan [1 ]
Gibson, Anthony G. [1 ]
Karahalios, Dean G. [2 ]
Potts, Eric A. [3 ]
Mobasser, Jean-Pierre [3 ]
Burch, Shane [1 ]
机构
[1] Univ Calif San Francisco, San Francisco, CA 94143 USA
[2] Rush Med Coll, NorthShore Med, Chicago, IL 60612 USA
[3] Goodman Campbell Brain & Spine, Indianapolis, IN USA
关键词
radiation; O-ARM; image-guided surgery; radiation exposure; intraoperative imaging; PEDICLE SCREW PLACEMENT; RADIATION-EXPOSURE; IMAGE-GUIDANCE; THORACIC SPINE; ACCURACY; LUMBAR; THORACOLUMBAR; SURGEON; SYSTEM; COMPLICATIONS;
D O I
10.1097/BRS.0b013e3182a51d1e
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. Cadaveric laboratory study. Objective. To compare the accuracy, efficiency, and safety of intraoperative cone beam-computed tomography with navigation (O-ARM) with traditional intraoperative fluoroscopy (C-ARM) for the placement of pedicle screws. Summary of Background Data. Radiation exposure remains a concern with traditional methods of intraoperative imaging in spine surgery. The use of O-ARM has been proposed for more accurate and efficient spinal instrumentation. Understanding radiation imparted to patients and surgeons by O-ARM is important for assessing risks and benefits of this technology, especially in light of evolving indications. Methods. Four surgeons placed 160 pedicle screws on 8 cadavers without deformity. Eighty pedicle screws were placed using O-ARM and C-ARM each. Instrumentation was placed bilaterally in the thoracic (T1-T6) spine and lumbosacral junction (L5-S1) using a standard open technique, whereas minimally invasive surgery technique was used at the lumbar 3 to 4 (L3-L4) level. A "postoperative" computed tomography (CT) scan was performed on cadavers where instrumentation was done using the C-ARM. An independent musculoskeletal radiologist assessed final images for screw position. Time required to set up and instrumentation was recorded. Dosimeters were placed on multiple aspects of cadavers and surgeons to record radiation exposure. Results. There were no differences in breach rate between the O-ARM and C-ARM groups (5 vs. 7,. 2 = 0.63, P = 0.4). The setup time for the O-ARM group was longer than that for the C-ARM group (592 vs. 297 s, P < 0.05). However, the average total time was statistically the same (1629 vs. 1639 s, P = 0.96). Radiation exposure was higher for surgeons in the C-ARM group and cadavers in the O-ARM group. When a "postoperative" CT scan was included in the estimation of the total radiation exposure, there was less of difference between the groups, but still more for the O-ARM group. Conclusion. In cadavers without deformity, O-ARM use results in similar breach rates as C-ARM for the placement of pedicle screws. Time for instrumentation is shorter with the O-ARM, but requires a longer setup time. The O-ARM exposes less radiation to the surgeon, but higher doses to the cadaver.
引用
收藏
页码:1953 / 1958
页数:6
相关论文
共 50 条
  • [41] A Novel Field-of-View Augmentation Wand for C-arm Computed Tomography-Like Fluoroscopy-Based Intraoperative Navigation New Technology
    Peleg, Eran
    Liebergall, Meir
    Weil, Yoram
    Khoury, Amal
    Joskowicz, Leo
    Mosheiff, Rami
    [J]. JOURNAL OF ORTHOPAEDIC TRAUMA, 2010, 24 (07) : 452 - 456
  • [42] Towards cardiac C-arm computed tomography
    Lauritsch, Guenter
    Boese, Jan
    Wigstroem, Lars
    Kemeth, Herbert
    Fahrig, Rebecca
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2006, 25 (07) : 922 - 934
  • [43] Application of C-arm computed tomography in cardiology
    Rieber, J.
    Rohkohl, C.
    Lauritsch, G.
    Rittger, H.
    Meissner, O.
    [J]. ZEITSCHRIFT FUR HERZ THORAX UND GEFASSCHIRURGIE, 2010, 24 (03): : 191 - 196
  • [44] Intraoperative Computed Tomography Versus 3D C-Arm Imaging for Navigated Spinal Instrumentation
    Hecht, Nils
    Yassin, Hadya
    Czabanka, Marcus
    Foehre, Bettina
    Arden, Klaus
    Liebig, Thomas
    Vajkoczy, Peter
    [J]. SPINE, 2018, 43 (05) : 370 - 377
  • [45] Intraoperative 2D C-arm and 3D O-arm in children: a comparative phantom study
    Prod'homme, M.
    Sans-Merce, M.
    Pitteloud, N.
    Damet, J.
    Lascombes, P.
    [J]. JOURNAL OF CHILDRENS ORTHOPAEDICS, 2018, 12 (05) : 550 - 557
  • [46] Percutaneous transthoracic needle biopsy for pulmonary nodules: a retrospective study of a comparison between C-arm cone-beam computed tomography and conventional computed tomography guidance
    Akkakrisee, Surasit
    Hongsakul, Keerati
    [J]. POLISH JOURNAL OF RADIOLOGY, 2020, 85 : E309 - E315
  • [47] C-arm cone-beam computed tomography-guided minimally invasive open excision of an osteoid osteoma undetectable on fluoroscopy: A case report
    Higuchi, Takashi
    Yamamoto, Norio
    Hayashi, Katsuhiro
    Takeuchi, Akihiko
    Abe, Kensaku
    Taniguchi, Yuta
    Araki, Yoshihiro
    Tsuchiya, Hiroyuki
    [J]. INTERNATIONAL JOURNAL OF SURGERY CASE REPORTS, 2019, 61 : 14 - 19
  • [48] Dosimetric Characterization of a Cone Beam O-Arm Imaging System
    Weir, V.
    Zhang, J.
    Fajardo, L.
    Hsiung, H.
    Ritenour, E.
    [J]. MEDICAL PHYSICS, 2008, 35 (06)
  • [49] Survival Impact of C-Arm Cone-Beam Computed Tomography on Hepatocellular Carcinoma Patients Undergoing Chemoembolization
    Wallace, Michael J.
    Abdelsalam, Mohamed E.
    Ensor, Joseph E.
    Kaseb, Ahmed O.
    Tam, Alda
    Avritscher, Rony
    Gupta, Sanjay
    [J]. HEPATOLOGY, 2012, 56 : 466A - 467A
  • [50] Learning Curve of C-Arm Cone-beam Computed Tomography Virtual. Navigation-Guided Percutaneous Transthoracic Needle Biopsy
    Ahn, Su Yeon
    Park, Chang Min
    Yoon, Soon Ho
    Kim, Hyungjin
    Goo, Jin Mo
    [J]. KOREAN JOURNAL OF RADIOLOGY, 2019, 20 (05) : 844 - 853