A sensitive method for determination of a broad range of perfluorinated compounds in serum suitable for large-scale human biomonitoring

被引:148
|
作者
Haug, Line Smastuen [1 ]
Thomsen, Cathrine [1 ]
Becher, Georg [1 ,2 ]
机构
[1] Norwegian Inst Publ Hlth, Div Environm Med, NO-0403 Oslo, Norway
[2] Univ Oslo, Dept Chem, NO-0315 Oslo, Norway
关键词
PFCs; LC-MS/MS; Analysis; Blood; Serum; Pseudo-MRM; Column switching; SOLID-PHASE EXTRACTION; PERFLUOROOCTANE SULFONATE PFOS; MASS-SPECTROMETRY; HUMAN BLOOD; ELECTROSPRAY-IONIZATION; PERFLUOROALKYL ACIDS; ALKYL SUBSTANCES; HUMAN PLASMA; TOXICITY; FLUOROCHEMICALS;
D O I
10.1016/j.chroma.2008.10.113
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A sensitive and reliable method based on column switching liquid chromatography (LC) coupled to a triple quadrupole mass spectrometer (MS) has been developed for quantification of 19 perfluorinated compounds (PFCs) in serum. A Volume of only 150 mu l serum is used and protein precipitation by methanol is the only sample preparation necessary prior to injection into the column switching system. Pseudo-MRM is used as a detection mode for determination of PFCs, resulting in reduced background noise and considerably increased sensitivity for the perfluorinated alkyl sulfonates and the perfluorinated alkyl sulfonamides. The estimated limits of detection for the method were as low as 0.0020-0.050 ng PFCs/ml serum. The accuracy determined from spiking experiments, reported as recovery of added amount, was between 85 and 121% in the range 0.20-50 ng PFC/ml serum, except for perfluorodecylsulfonate for which the accuracy was 146% at 0.20 ng PFC/ml serum. The low sample volume needed, the limited manual handling and the broad range of analytes which are included, make this method advantageous for large-scale epidemiological studies. This column-switching technique can easily be set up on standard LC-MS/MS instruments and is thus available to a wide range of laboratories. (C) 2008 Elsevier B.V. All rights reserved.
引用
收藏
页码:385 / 393
页数:9
相关论文
共 50 条
  • [41] Large-scale targeted metabolomics method for metabolite profiling of human samples
    Cao, Guodong
    Song, Zhengbo
    Hong, Yanjun
    Yang, Zhiyi
    Song, Yuanyuan
    Chen, Zhongjian
    Chen, Zhaobin
    Cai, Zongwei
    ANALYTICA CHIMICA ACTA, 2020, 1125 : 144 - 151
  • [42] A highly sensitive method for large-scale measurements of 1,25-dihydroxyvitamin D
    Arbour, NC
    Ross, TK
    Zierold, C
    Prahl, JM
    DeLuca, HF
    ANALYTICAL BIOCHEMISTRY, 1998, 255 (01) : 148 - 154
  • [43] A Range-Extension Method for an Indoor Standard Device for Large-Scale Length Measurement
    Hu, Chang'an
    Xue, Liang
    Li, Jiangang
    Lv, Fei
    APPLIED SCIENCES-BASEL, 2023, 13 (16):
  • [44] An innovative production method for a C/C-SiC brake disc, suitable for a large-scale production
    Nestler, D.
    Nier, N.
    Roder, K.
    Todt, A.
    Wielage, B.
    Wagner, G.
    Kroll, L.
    Paessler, E.
    Spange, S.
    Weisshuhn, J.
    H, Wuerfel
    6TH INTERNATIONAL MUNICH CHASSIS SYMPOSIUM 2015, 2015, : 605 - 627
  • [45] ENZYME REPLACEMENT THERAPY IN GAUCHERS-DISEASE - LARGE-SCALE PURIFICATION OF GLUCOCEREBROSIDASE SUITABLE FOR HUMAN ADMINISTRATION
    FURBISH, FS
    BLAIR, HE
    SHILOACH, J
    PENTCHEV, PG
    BRADY, RO
    PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 1977, 74 (08) : 3560 - 3563
  • [46] Development of an ion-pair liquid chromatography-high resolution mass spectrometry method for determination of organophosphate pesticide metabolites in large-scale biomonitoring studies
    Cequier, Enrique
    Sakhi, Amrit Kaur
    Haug, Line Smastuen
    Thomsen, Cathrine
    JOURNAL OF CHROMATOGRAPHY A, 2016, 1454 : 32 - 41
  • [47] Rapid and Sensitive HPLC Method for the Simultaneous Determination of Paraquat and Diquat in Human Serum
    Shuuji Hara
    Noriaki Sasaki
    Daizi Takase
    Shouichi Shiotsuka
    Kentaro Ogata
    Koujiro Futagami
    Kazuo Tamura
    Analytical Sciences, 2007, 23 : 523 - 526
  • [48] SENSITIVE METHOD FOR DETERMINATION OF SUBMICROGRAM QUANTITIES OF MANGANESE AND ITS APPLICATION TO HUMAN SERUM
    FERNANDEZ, AA
    JACOBS, SL
    SOBEL, C
    ANALYTICAL CHEMISTRY, 1963, 35 (11) : 1721 - &
  • [49] A SELECTIVE AND SENSITIVE METHOD FOR THE DETERMINATION OF PROPAFENON AND METABOLITES IN HUMAN-SERUM BY HPLC
    BEUMAN, G
    BOESTEN, M
    VANDERHEIJDEN, J
    PHARMACEUTISCH WEEKBLAD-SCIENTIFIC EDITION, 1986, 8 (06) : 319 - 319
  • [50] Rapid and sensitive HPLC method for the simultaneous determination of paraquat and diquat in human serum
    Hara, Shuuji
    Sasaki, Noriaki
    Takase, Daizi
    Shiotsuka, Shouichi
    Ogata, Kentaro
    Futagami, Koujiro
    Tamura, Kazuo
    ANALYTICAL SCIENCES, 2007, 23 (05) : 523 - 526