Fused silica capillary columns of the internal diameter of 320 mu m were packed with the Nucleosil C18 stationary phase of 5 mu m using the slurry packing method. The time of the bed compaction phase, packing pressure, and the use of ultrasound varied to study their influence on the column performance. Van Deemter curves were measured and separation impedance values were calculated in order to assess both separation efficiency and kinetic performance of the columns. Selected columns were tested again after nine months to evaluate the stability of their beds. Separation efficiencies of all columns were similar, but a major difference, caused by the use of ultrasound, was observed in the bed stability. Columns sonicated for 25 minutes during the bed compaction phase exhibited unchanged performance in the course of several months, while the performance of non-sonicated columns decreased. (C) 2013 Institute of Chemistry, Slovak Academy of Sciences
引用
收藏
页码:22 / 28
页数:7
相关论文
共 50 条
  • [21] Hydrodynamic impact of particle shape in slurry packed liquid chromatography columns
    Lottes, Florian
    Arlt, Wolfgang
    Minceva, Mirjana
    Stenby, Erling H.
    JOURNAL OF CHROMATOGRAPHY A, 2009, 1216 (30) : 5687 - 5695
  • [22] QUANTITATIVE ANALYTICAL ASPECTS OF REVERSED-PHASE LIQUID-CHROMATOGRAPHY WITH SLURRY-PACKED CAPILLARY COLUMNS
    BORRA, C
    HAN, SM
    NOVOTNY, M
    JOURNAL OF CHROMATOGRAPHY, 1987, 385 : 75 - 85
  • [23] Performance of capillary columns for high-temperature gas chromatography
    Pereira, AD
    Neto, FRD
    JOURNAL OF CHROMATOGRAPHIC SCIENCE, 2000, 38 (09) : 369 - 376
  • [24] A STUDY OF PERFORMANCE OF PACKED GAS CHROMATOGRAPHY COLUMNS
    NOGARE, SD
    CHIU, J
    ANALYTICAL CHEMISTRY, 1962, 34 (08) : 890 - &
  • [25] Comparison of the chromatographic behavior of monolithic capillary columns in capillary electrochromatography and nano-high-performance liquid chromatography
    Freitag, R
    JOURNAL OF CHROMATOGRAPHY A, 2004, 1033 (02) : 267 - 273
  • [26] Monolithic columns in high-performance liquid chromatography
    Guiochon, Georges
    JOURNAL OF CHROMATOGRAPHY A, 2007, 1168 (1-2) : 101 - 168
  • [27] Application of monolithic columns in high performance liquid chromatography
    Mistry, K
    Grinberg, N
    JOURNAL OF LIQUID CHROMATOGRAPHY & RELATED TECHNOLOGIES, 2005, 28 (7-8) : 1055 - 1074
  • [28] Surfactant-bound monolithic columns for separation of proteins in capillary high performance liquid chromatography
    Gu, Congying
    He, Jun
    Jia, Jinping
    Fang, Nenghu
    Simmons, Robert
    Shamsi, Shahab A.
    JOURNAL OF CHROMATOGRAPHY A, 2010, 1217 (04) : 530 - 539
  • [29] Mutation detection by capillary denaturing high-performance liquid chromatography using monolithic columns
    Huber, CG
    Premstaller, A
    Wen, X
    Oberacher, H
    Bonn, GK
    Oefner, PJ
    JOURNAL OF BIOCHEMICAL AND BIOPHYSICAL METHODS, 2001, 47 (1-2): : 5 - 19
  • [30] SLURRY PACKING OF HIGH-PERFORMANCE COLUMNS FOR LIQUID-CHROMATOGRAPHY .2. THE EFFECT OF PRESSURE
    KARAPETYAN, SA
    YAKUSHINA, LM
    VASIJAROV, GG
    BRAZHNIKOV, VV
    JOURNAL OF HIGH RESOLUTION CHROMATOGRAPHY & CHROMATOGRAPHY COMMUNICATIONS, 1985, 8 (03): : 148 - 149