Validation of an LC-MS/MS method for the quantitative determination of mavoglurant (AFQ056) in human plasma

被引:4
|
作者
Jakab, Annamaria [1 ]
Winter, Serge [1 ]
Raccuglia, Marc [1 ]
Picard, Frank [2 ]
Dumitras, Swati [1 ]
Woessner, Ralph [1 ]
Mistry, Sanket [3 ]
Chudasama, Jayraj [3 ]
Guttikar, Swati [3 ]
Kretz, Olivier [1 ]
机构
[1] Novartis Pharma AG, Drug Metab & Pharmacokinet, Novartis Inst BioMed Res, CH-4056 Basel, Switzerland
[2] Novartis Inst BioMed Res, Bioanalyt Div, CH-4056 Basel, Switzerland
[3] Veeda Clin Res Pvt Ltd, Ahmadabad 380015, Gujarat, India
关键词
AFQ056; Mavoglurant; mGluR5; LC-MS/MS; Human plasma; Recovery; Validation; FRAGILE-X-SYNDROME; GLUTAMATE; PHYSIOLOGY;
D O I
10.1007/s00216-012-6456-y
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A simple, sensitive, and selective liquid chromatography/tandem mass spectrometry method was validated for the identification and quantification of mavoglurant (AFQ056) in human plasma. The chromatographic separation was performed using a Cosmosil 5 C18 (150 x 4.6 mm, 5 mu m) column at 40 +/- 0.5 A degrees C with a mobile phase consisting of acetic acid in water (0.1 %, v/v)/methanol (10:90, v/v) with a flow rate of 1.0 mL/min followed by quantification with tandem mass spectrometry, operating with electrospray ionization in positive ion mode and applying multiple reaction monitoring. The validated method described in this paper presents high absolute recovery with precision and accuracy meeting the acceptance criteria. The method was precise and accurate for 2- and 10-fold dilution of samples. The method was validated using sodium heparin as specific anticoagulant, and the anticoagulant effect was tested by lithium heparin and K(3)EDTA. The method was successfully cross-validated between two bioanalytical sites. The method was specific for mavoglurant within the given criteria for acceptance (apparent peak area at the retention time of mavoglurant in zero samples was less than 20 % compared with the mean peak area at LLOQ) in human plasma. The method was fully validated for the quantitative determination of mavoglurant in human plasma between the range of 2.00 and 2,500 ng/mL.
引用
收藏
页码:215 / 223
页数:9
相关论文
共 50 条
  • [1] Validation of an LC-MS/MS method for the quantitative determination of mavoglurant (AFQ056) in human plasma
    Annamaria Jakab
    Serge Winter
    Marc Raccuglia
    Frank Picard
    Swati Dumitras
    Ralph Woessner
    Sanket Mistry
    Jayraj Chudasama
    Swati Guttikar
    Olivier Kretz
    [J]. Analytical and Bioanalytical Chemistry, 2013, 405 : 215 - 223
  • [2] LC-MS/MS Method Development and Validation for the Determination of Nifedipine in Human Plasma
    Logoyda, Liliya S.
    [J]. BIOINTERFACE RESEARCH IN APPLIED CHEMISTRY, 2020, 10 (05): : 6220 - 6227
  • [3] Development and validation of a LC-MS/MS method for the determination of viaminate in human plasma
    Yin, Jing
    Ni, Kun-Yi
    Shen, Yue
    Ma, Peng-Cheng
    Cao, Ling
    Wang, Wei-Peng
    Wang, Yu
    [J]. JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2007, 856 (1-2): : 376 - 380
  • [4] Validation of a selective method for determination of paroxetine in human plasma by LC-MS/MS
    Massaroti, P
    Cassiano, NM
    Duarte, LF
    Campos, DR
    Marchioretto, MAM
    Bernasconi, G
    Calafatti, S
    Barros, FAP
    Meurer, EC
    Pedrazzoli, J
    [J]. JOURNAL OF PHARMACY AND PHARMACEUTICAL SCIENCES, 2005, 8 (02): : 340 - 347
  • [5] Validation of LC-MS/MS method for determination of ertapenem in human plasma and urine
    Kim, Yun-Jeong
    Han, Song-Hee
    Jeon, Ji-Young
    Hwang, Min-Ho
    Im, Yong-Jin
    Chae, Soo-Wan
    Kim, Min-Gul
    [J]. ANALYTICAL SCIENCE AND TECHNOLOGY, 2012, 25 (01): : 19 - 24
  • [6] Development and Validation of an LC-MS/MS Method for the Quantitative Determination of Contezolid in Human Plasma and Cerebrospinal Fluid
    Zhang, Guanxuanzi
    Zhang, Na
    Dong, Liuhan
    Bai, Nan
    Cai, Yun
    [J]. PHARMACEUTICALS, 2023, 16 (01)
  • [7] Development and validation of LC-MS/MS method for quantitative determination of (-)-securinine in mouse plasma
    Wabuyele, Simuli L.
    Wald, David
    Xu, Yan
    [J]. JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2014, 960 : 19 - 26
  • [8] Development and validation of a quantitative LC-MS/MS method for the simultaneous determination of ceftolozane and tazobactam in human plasma and urine
    Putnam, William C.
    Kallem, Raja Reddy
    Edpuganti, Vindhya
    Subramaniyan, Indhu
    Hall, Ronald G., II
    [J]. JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2020, 1159
  • [9] Development and Validation of New LC-MS/MS Method for the Determination of armodafinil in Human Plasma
    Ramesh, Devi
    Ramakrishna, Singirikonda
    Habibuddin, Mohammad
    [J]. CURRENT PHARMACEUTICAL ANALYSIS, 2012, 8 (03) : 295 - 305
  • [10] LC-MS/MS method for the determination of clodronate in human plasma
    Hasan, Mahmoud
    Schumacher, Gitta
    Seekamp, Anne
    Taedken, Tobias
    Siegmund, Werner
    Oswald, Stefan
    [J]. JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2014, 100 : 341 - 347