Dental Implants with Conical Implant-Abutment Interface: Influence of the Conical Angle Difference on the Mechanical Behavior of the Implant

被引:22
|
作者
Aguirrebeitia, Josu [1 ]
Abasolo, Mikel [1 ]
Vallejo, Javier [1 ]
Ansola, Ruben [1 ]
机构
[1] Univ Basque Country, ETSI BILBAO, Dept Mech Engn, Bilbao 48013, Spain
关键词
conical angle difference; conical implant; finite element model; sensitivity study; FINITE-ELEMENT-ANALYSIS; IN-VITRO EVALUATION; BACTERIAL LEAKAGE; PROSTHESIS; COMPONENTS; SYSTEM; DOGS; FIT;
D O I
10.11607/jomi.2775
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: Misfit in the conical implant-abutment interface plays an important role on the mechanical behavior of the implant when masticatory forces are applied. The origin of the misfit adopted in this work is a conical angle difference between implant and abutment, which can be due to a combination of design decisions and manufacturing tolerances. The goal of this work was to investigate the effects of the implant-abutment conical angle difference in the following mechanical features: interfacial microgap, preload loss on the bolt, stress level in the bone, and abutment removal force and/or torque. Materials and Methods: A simplified three-dimensional nonlinear monoparametric finite element model of an OsseoSpeed TX 4.5 S 9-mm implant (Astra Tech) with a tapered implant-abutment interface was built to evaluate the variability of the mechanical features cited above with the conical angle difference, keeping constant the overall geometry, load and boundary conditions, material properties, frictional behavior, and mesh structure. Results: As the conical angle difference increased, the following effects were observed: the microgap decreased and remained almost constant for values over a given positive angle difference, the stress level in the bone increased sensitively, the removal force and/or torque needed to separate the abutment from the implant varied slightly, and the bolt preload loss increased. Conclusions: In light of the results provided, the conical angle difference in the implant-abutment interface had a significant influence on the overall mechanical behavior of the implant. Among the four mechanical features considered, the interfacial microgap and the bone stress were demonstrated to be the most sensitive to the conical angle difference, and therefore the most relevant when selecting an optimum value in the design process of a conical interface.
引用
收藏
页码:E72 / E82
页数:11
相关论文
共 50 条
  • [1] INVESTIGATION OF AXIAL DISPLACEMENT OF SUPRASTRUCTURES AT THE CONICAL IMPLANT-ABUTMENT INTERFACE
    Mishin, Dmitry
    Kalinovsky, Sergei
    Ilyasov, Vyacheslav
    Dedova, Veronika
    STRUCTURAL INTEGRITY AND LIFE-INTEGRITET I VEK KONSTRUKCIJA, 2023, 23 (01): : 65 - 69
  • [2] Analytical and experimental position stability of the abutment in different dental implant systems with a conical implant-abutment connection
    Semper-Hogg, Wiebke
    Kraft, Silvan
    Stiller, Sebastian
    Mehrhof, Juergen
    Nelson, Katja
    CLINICAL ORAL INVESTIGATIONS, 2013, 17 (03) : 1017 - 1023
  • [3] The Impact of Conical and Nonconical Abutments on Bacterial Infiltration at the Implant-Abutment Interface
    Guerra, Elen
    Pereira, Cristiane
    Faria, Renata
    Cardoso Jorge, Antonio Olavo
    Bottino, Marco Antonio
    de Melo, Renata Marques
    INTERNATIONAL JOURNAL OF PERIODONTICS & RESTORATIVE DENTISTRY, 2016, 36 (06) : 825 - 831
  • [4] The Influence of Torque Tightening on the Position Stability of the Abutment in Conical Implant-Abutment Connections
    Hogg, Wiebke Semper
    Zulauf, Kris
    Mehrhof, Juergen
    Nelson, Katja
    INTERNATIONAL JOURNAL OF PROSTHODONTICS, 2015, 28 (05) : 538 - 541
  • [5] Influence of abutment materials on the implant-abutment joint stability in internal conical connection type implant systems
    Jo, Jae-Young
    Yang, Dong-Seok
    Huh, Jung-Bo
    Heo, Jae-Chan
    Yun, Mi-Jung
    Jeong, Chang-Mo
    JOURNAL OF ADVANCED PROSTHODONTICS, 2014, 6 (06): : 491 - 497
  • [6] Assessment of lipopolysaccharide microleakage at conical implant-abutment connections
    Sönke Harder
    Elgar Susanne Quabius
    Lars Ossenkop
    Matthias Kern
    Clinical Oral Investigations, 2012, 16 : 1377 - 1384
  • [7] Fatigue induced changes in conical implant-abutment connections
    Blum, Kai
    Wiest, Wolfram
    Fella, Christian
    Balles, Andreas
    Dittmann, Jonas
    Rack, Alexander
    Maier, Dominik
    Thomann, Ralf
    Spies, Benedikt Christopher
    Kohal, Ralf Joachim
    Zabler, Simon
    Nelson, Katja
    DENTAL MATERIALS, 2015, 31 (11) : 1415 - 1426
  • [8] Assessment of lipopolysaccharide microleakage at conical implant-abutment connections
    Harder, Soenke
    Quabius, Elgar Susanne
    Ossenkop, Lars
    Kern, Matthias
    CLINICAL ORAL INVESTIGATIONS, 2012, 16 (05) : 1377 - 1384
  • [9] Load Fatigue Performance of Conical Implant-Abutment Connections
    Seetoh, Y. L.
    Tan, Keson B.
    Chua, E. K.
    Quek, H. C.
    Nicholls, Jack I.
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2011, 26 (04) : 797 - 806
  • [10] Effects of different torque levels on the implant-abutment interface in a conical internal connection
    Alexandre Gehrke, Sergio
    Shibli, Jamil Awad
    Aramburu Junior, Jaime Sarda
    Mate Sanchez de Val, Jose Eduardo
    Luis Calvo-Girardo, Jose
    Dedavid, Berenice Anina
    BRAZILIAN ORAL RESEARCH, 2016, 30 (01):