Quantifying impacts on the measurement uncertainty in flow calibration arising from dynamic flow effects

被引:14
|
作者
Engel, Rainer [1 ]
Baade, Hans-Joachim [2 ]
机构
[1] Phys Tech Bundesanstalt, Dept Liquid Flow, Braunschweig, Germany
[2] Droge Baade Drescher Consulting Engineers, Salzgitter, Germany
关键词
Liquid flow calibration; Fluctuation of process parameters; Measurement uncertainty analysis; Traceability in fluid flow; Element-by-element chain of traceability; Dynamic system behavior;
D O I
10.1016/j.flowmeasinst.2014.11.003
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
Traceability of the measurement units in fluid flow metering, as the state-of-the-art approach, is practiced as a so-called element-by-element method which relies upon the idealistic assumption that the measurement process in a flow calibration facility can be run under exact steady-state conditions, that is, absolutely no fluctuations of the flow quantities are assumed to occur during the flow measurement process. Practical experiences, combined with a model-based analytical view of the measurement process in a flow standard, have revealed that dynamic impacts on the measurement uncertainty due to flowrate fluctuations have to be taken into account, in addition to the steady-state traceability chain. (C) 2014 Elsevier Ltd. All rights reserved.
引用
收藏
页码:51 / 60
页数:10
相关论文
共 50 条
  • [31] Nonparametric estimation of measurement uncertainty arising from sampling
    Lourenco, Felipe Rebello
    de Oliveira, Elcio Cruz
    [J]. ANALYTICA CHIMICA ACTA, 2024, 1305
  • [32] Measurement uncertainty arising from sampling of environmental samples
    Glavic-Cindro, Denis
    Hazou, Eyakifama
    Korun, Matjaz
    Kristof, Romana
    Osterman, Petra
    Petrovic, Toni
    Vodenik, Branko
    Zorko, Benjamin
    [J]. APPLIED RADIATION AND ISOTOPES, 2020, 156
  • [33] Dynamic calibration of laminar flow sensor for gases
    Funaki, T
    Sengoku, K
    Kawashima, K
    Kagawa, T
    [J]. SICE 2004 ANNUAL CONFERENCE, VOLS 1-3, 2004, : 2437 - 2442
  • [34] Uncertainty analysis and evaluation of gas flow measurement
    Lu, Ping
    [J]. PROGRESS IN MINE SAFETY SCIENCE AND ENGINEERING II, 2014, : 1095 - 1098
  • [35] UNDERSTANDING MEASUREMENT UNCERTAINTY IN DP FLOW DEVICES
    Hering, Michael S.
    Mesnard, David R.
    [J]. PROCEEDINGS OF THE ASME POWER CONFERENCE, 2014, VOL 2, 2014,
  • [36] Quantifying the water balance in a planar hillslope plot: Effects of measurement errors on flow prediction
    Kampf, Stephanie K.
    Burges, Stephen J.
    [J]. JOURNAL OF HYDROLOGY, 2010, 380 (1-2) : 191 - 202
  • [37] MEASUREMENT OF GAS FLOW-RATES AND CALIBRATION OF FLOW METERS FOR FLAME SPECTROSCOPY
    KIRKBRIGHT, GF
    SARGENT, M
    [J]. ANALYST, 1968, 93 (1109) : 552 - +
  • [38] Quantifying Uncertainty in Flow Functions Derived from SCAL DataUSS Relative Permeability and Capillary Pressure
    S. Subbey
    H. Monfared
    M. Christie
    M. Sambridge
    [J]. Transport in Porous Media, 2006, 65 : 265 - 286
  • [39] Helical flow arising from the yielded annular flow of a Bingham fluid
    Shepherd, J. J.
    Stacey, A. J.
    Khan, A. A.
    [J]. APPLIED MATHEMATICAL MODELLING, 2014, 38 (23) : 5382 - 5391
  • [40] Quantifying parameter uncertainty in reservoir operation associated with environmental flow management
    He, Shan
    Yin, Xin'an
    Yu, Chunxue
    Xu, Zhihao
    Yang, Zhifeng
    [J]. JOURNAL OF CLEANER PRODUCTION, 2018, 176 : 1271 - 1282