Criterion for determining the analytical emission lines in glow discharge optical emission spectrometry

被引:0
|
作者
Wagatsuma, K [1 ]
机构
[1] Tohoku Univ, Mat Res Inst, Aoba Ku, Sendai, Miyagi 9808577, Japan
关键词
optical emission spectrometry; glow discharge plasma; analytical line; emission; intensity; self-absorption; vanadium; copper;
D O I
暂无
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Two major factors for selecting analytical lines, the sensitivity and the self-absorption effect, were compared between atomic lines and singly-ionized lines of copper as well as vanadium in glow discharge optical emission spectrometry. They have intense ionic lines which are excited through resonance charge-transfer collisions. It was found that the ionic lines are more sensitive compared to the atomic resonance lines; in addition, the ionic lines are essentially free from self-absorption, while the atomic lines suffer from self-absorption to a certain extent. The Cu II 224.70 nm, V II 309.31 nm, and V II 266.32 nm lines in the argon plasma, and the Cu II 270.32 nm line in the neon plasma are recommended as analytical lines of these elements.
引用
收藏
页码:457 / 461
页数:5
相关论文
共 50 条