A high-resolution gas-source isotope ratio mass spectrometer

被引:76
|
作者
Eiler, John M. [1 ]
Clog, Matthieu [1 ]
Magyar, Paul [1 ]
Piasecki, Alison [1 ]
Sessions, Alex [1 ]
Stolper, Daniel [1 ]
Deerberg, Michael
Schlueter, Hans-Juergen
Schwieters, Johannes
机构
[1] CALTECH, Div Geol & Planetary Sci, Pasadena, CA 91125 USA
基金
美国国家科学基金会;
关键词
Stable isotope; Geochemistry; Hydrocarbon; STABLE-ISOTOPES; OXYGEN; CARBON; ARGON;
D O I
10.1016/j.ijms.2012.10.014
中图分类号
O64 [物理化学(理论化学)、化学物理学]; O56 [分子物理学、原子物理学];
学科分类号
070203 ; 070304 ; 081704 ; 1406 ;
摘要
We describe a new high-resolution, multi-collector gas source mass spectrometer designed for isotopic analysis of volatile and semi-volatile molecules: the Thermo Scientific MAT253-Ultra, a prototype double-focusing isotope ratio mass spectrometer installed in the Caltech laboratories for stable isotope geochemistry. This instrument achieves mass resolving power of up to similar to 27,000 (M/Delta M) and can analyze diverse gases and semi-volatile compounds using a conventional dual inlet and/or a carrier gas. It has a multi-collector array comprised of 7 detector positions with adjustable spacing, all of which can register ions through an SEM or Faraday cup and spanning up to a 10(13) range in signal strength. Abundance sensitivity in the He mass range is as good as 10(-12), and precision commonly approaches the counting statistics limit down to 0.1 parts per thousand (SEM) or 0.01 parts per thousand (Faraday) for a range of analytes. This instrument permits resolution of isobaric interferences arising from both contaminants and multiple isotopologues of an analyte that share a cardinal mass, enabling direct isotopic analysis of molecules with complex mass spectra such as hydrocarbons. This ability should enable the measurement of position-specific isotopic compositions, including multiple substitutions, by comparing isotope ratios of molecular ions with those of daughter fragment ions (assuming products of recombination and other source reactions are recognized and corrected for). The combination of high mass resolution with stable multicollection will provide a wide range of potential new tools for isotope geochemistry, including (but not limited to): singly and multiply substituted methane and larger hydrocarbons; position-specific C-13 analysis of propane and larger hydrocarbons; precise analysis of O-17/O-16 and O-18/O-16 on fragment ions from CO2 and other molecules; analysis of a variety of N2O isotopologues (including O-18,O-17, position-specific N-15, and various 'clumped' species); and high precision and abundance sensitivity noble gas analyses. These capabilities greatly extend the scope of stable isotope variations that can be utilized for problems in forensics, environmental geochemistry, biochemistry, and Earth and planetary sciences. (C) 2012 Elsevier B.V. All rights reserved.
引用
收藏
页码:45 / 56
页数:12
相关论文
共 50 条
  • [1] Preparation and Application of High-resolution Isotope Ratio Mass Spectrometer for Analysis of Hydrogen Isotope
    Shi, Lei
    Hou, Yanli
    [J]. Yuanzineng Kexue Jishu/Atomic Energy Science and Technology, 2021, 55 (03): : 544 - 549
  • [2] Determination for methane clumped isotopes by high-resolution gas stable isotope mass spectrometer
    Pang, Zhiyong
    Wang, Xinchu
    Li, Siliang
    Lang, Yunchao
    Tian, Yourong
    Lu, Li
    [J]. Chinese Journal of Analysis Laboratory, 2023, 42 (01) : 33 - 37
  • [3] Methods and limitations of 'clumped' CO2 isotope (Δ47) analysis by gas-source isotope ratio mass spectrometry
    Huntington, K. W.
    Eiler, J. M.
    Affek, H. P.
    Guo, W.
    Bonifacie, M.
    Yeung, L. Y.
    Thiagarajan, N.
    Passey, B.
    Tripati, A.
    Daeron, M.
    Came, R.
    [J]. JOURNAL OF MASS SPECTROMETRY, 2009, 44 (09): : 1318 - 1329
  • [4] HIGH-RESOLUTION MASS-SPECTROMETER
    ALEXANDRU, G
    DUTESCU, N
    MOLNAR, S
    [J]. STUDII SI CERCETARI DE FIZICA, 1972, 24 (06): : 765 - +
  • [5] Applicability of high-resolution ICP mass spectrometry for isotope ratio measurements
    Vanhaecke, F
    Moens, L
    Dams, R
    Papadakis, I
    Taylor, P
    [J]. ANALYTICAL CHEMISTRY, 1997, 69 (02) : 268 - 273
  • [6] Analysis of the site-specific carbon isotope composition of propane by gas source isotope ratio mass spectrometer
    Piasecki, Alison
    Sessions, Alex
    Lawson, Michael
    Ferreira, A. A.
    Santos Neto, E. V.
    Eiler, John M.
    [J]. GEOCHIMICA ET COSMOCHIMICA ACTA, 2016, 188 : 58 - 72
  • [7] Sensitive and selective determination of pesticides in air using a high-resolution gas chromatograph/high-resolution mass spectrometer
    Murayama, H
    Mukai, H
    Tanabe, A
    Moriyama, N
    Kitayama, Y
    [J]. BUNSEKI KAGAKU, 2001, 50 (08) : 537 - 545
  • [8] A MASS SPECTROMETER FOR HIGH PRECISION ISOTOPE RATIO DETERMINATIONS
    WANLESS, RK
    THODE, HG
    [J]. JOURNAL OF SCIENTIFIC INSTRUMENTS, 1953, 30 (11): : 395 - 398
  • [9] HIGH-RESOLUTION QUADRUPOLE MASS-SPECTROMETER
    KONENKOV, NV
    MOGILCHENKO, GA
    SILAKOV, SS
    SHAGIMURATOV, GI
    [J]. INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 1990, 33 (02) : 433 - 435
  • [10] Performance of the NGX High-Resolution Multiple Collector Noble Gas Mass Spectrometer
    Li, Yan
    Tootell, Damian
    Holland, Greg
    Zhou, Zheng
    [J]. GEOCHEMISTRY GEOPHYSICS GEOSYSTEMS, 2021, 22 (11)