The Modern Pneumatic Tube System Transports with Reduced Speed Does Not Affect Special Coagulation Tests

被引:7
|
作者
Slavik, Ludek [1 ,2 ]
Ulehlova, Jana [1 ,2 ]
Bradacova, Pavla [1 ,2 ]
Chasakova, Katerina [1 ,2 ]
Hlusi, Antonin [1 ,2 ]
Palova, Miroslava [1 ,2 ]
Entrova, Alice [3 ]
机构
[1] Palacky Univ, Fac Med & Dent, Dept Hematooncol, Olomouc, Czech Republic
[2] Univ Hosp Olomouc, Olomouc, Czech Republic
[3] Univ Hosp Olomouc, Dept Blood Transfus, Olomouc, Czech Republic
关键词
Pneumatic tube system; Controlled acceleration; Platelet; PLATELET-FUNCTION; ROUTINE HEMATOLOGY; AGGREGOMETRY; PARAMETERS; DELIVERY;
D O I
10.1007/s10916-020-01614-6
中图分类号
R19 [保健组织与事业(卫生事业管理)];
学科分类号
摘要
Pneumatic tube transport systems (PTS) for delivery of patient samples to a hemostasis laboratory are often used to reduce turnaround time for vital analyses. PTS in our hospital has the ability to regulate the transport speed in the range of 3-6 m/s with acceleration control technology. We evaluated the effects of PTS transport for routine coagulation tests, platelet function tests and special global coagulation tests. Duplicate samples were collected from 29 patients and 40 healthy individuals. One sample was sent using PTS and the other was carried by personnel to the lab for determination of protrombin time, activated partial thromboplastin time, trombin time, fibrinogen, antitrombin and thrombin generation test. Platelet function was measured by means of a Apact 4004 (R) analyzer using the inductors (ADP, Arachidonic acid and Epinephrine). Samples transported using PTS with normal transport speed 6 m/s does not affect basic coagulation tests (PT, aPTT, FIB, TT and AT), but TGT has significantly altered. The use of PTS with controlled acceleration regulated the increase in thrombin generation from 10% to 3%, which is not statistically signifiant. The use of PTS with controlled acceleration did not show a significant difference even with the highly sensitive method of platelet aggregation. We conclude that PTS with acceleration control with transport speed from 3 to 6 m/s does not affect to platelet activity as measured by LTA and also global coagulation test - TGT. The advantage of PTS transport is very rapid assessment laboratory testing. From the above validation study, it is clear that PTS should always be validated for specialized laboratory methods and appropriately adapted to specific transport conditions.
引用
收藏
页数:5
相关论文
共 19 条
  • [1] The Modern Pneumatic Tube System Transports with Reduced Speed Does Not Affect Special Coagulation Tests
    Luděk Slavík
    Jana Úlehlová
    Pavla Bradáčová
    Kateřina Chasáková
    Antonín Hluší
    Miroslava Palová
    Alice Entrová
    Journal of Medical Systems, 2020, 44
  • [2] The effects of transport by pneumatic tube system on blood cell count, erythrocyte sedimentation and coagulation tests
    Kocak, Fatma Emel
    Yontem, Mustafa
    Yucel, Ozlem
    Cilo, Mustafa
    Genc, Ozlem
    Meral, Ayfer
    BIOCHEMIA MEDICA, 2013, 23 (02) : 206 - 210
  • [3] Comparison of pneumatic tube system with manual transport for routine chemistry, hematology, coagulation and blood gas tests
    Pupek, Alex
    Matthewson, Beverly
    Whitman, Erin
    Fullarton, Rachel
    Chen, Yu
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2017, 55 (10) : 1537 - 1544
  • [4] THE PNEUMATIC TUBE SYSTEM DOES NOT AFFECT NORMAL COMPLETE BLOOD COUNT RESULTS: A VALIDATION STUDY
    Al-Riyami, Arwa Z.
    Al-Khabori, Murtadha
    Al-Azwani, Intisar
    Al-Hadhrami, Rajaa
    Davis, Heidi
    Al-Farsi, Khalil
    Al-Hunaeini, Mohammed
    Alkindi, Salam
    Daar, Shahina
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2013, 35 : 120 - 121
  • [5] Investigation of the effects of pneumatic tube transport system on routine biochemistry, hematology, and coagulation tests in Ankara City Hospital
    Yurt, Emine F.
    Akbiyik, Filiz
    Bicer, Cemile
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2022, 60 (05) : 707 - 713
  • [6] Pre-analytical effects of pneumatic tube system transport on routine haematology and coagulation tests, global coagulation assays and platelet function assays
    Le Quellec, Sandra
    Paris, Mickael
    Nougier, Christophe
    Sobas, Frederic
    Rugeri, Lucia
    Girard, Sandrine
    Bordet, Jean-Claude
    Negrier, Claude
    Dargaud, Yesim
    THROMBOSIS RESEARCH, 2017, 153 : 7 - 13
  • [7] Pneumatic tube transport of platelet concentrates does not affect in vitro platelet function
    Henskens, Y.
    Lance, M.
    Van Oerle, R.
    Theunissen, M.
    Marcus, M.
    JOURNAL OF THROMBOSIS AND HAEMOSTASIS, 2011, 9 : 552 - 552
  • [8] The pneumatic tube system does not affect complete blood count results; a validation study at a tertiary care hospital
    Al-Riyami, A. Z.
    Al-Khabori, M.
    Al-Hadhrami, R. M.
    Al-Azwani, I. S.
    Davis, H. M.
    Al-Farsi, K. S.
    Alkindi, S. S.
    Daar, S. F.
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2014, 36 (05) : 514 - 520
  • [9] Effects of a pneumatic tube system on routine and novel hematology and coagulation parameters in healthy volunteers
    Kratz, Alexander
    Salem, Raneem O.
    Van Cott, Elizabeth M.
    ARCHIVES OF PATHOLOGY & LABORATORY MEDICINE, 2007, 131 (02) : 293 - 296
  • [10] Speed of sample transportation by a pneumatic tube system can influence the degree of hemolysis
    Tiwari, Aseem Kumar
    Pandey, Prashant
    Dixit, Surbhi
    Raina, Vimarsh
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2012, 50 (03) : 471 - 474