Experimental validation of a deforming grid 4D dose calculation for PBS proton therapy

被引:29
|
作者
Krieger, Miriam [1 ,2 ]
Klimpki, Grischa [1 ,2 ]
Fattori, Giovanni [1 ]
Hrbacek, Jan [1 ]
Oxley, David [1 ]
Safai, Sairos [1 ]
Weber, Damien C. [1 ,3 ]
Lomax, Antony J. [1 ,2 ]
Zhang, Ye [1 ]
机构
[1] Paul Scherrer Inst, Ctr Proton Therapy, CH-5232 Villigen, Switzerland
[2] Swiss Fed Inst Technol, Dept Phys, CH-8092 Zurich, Switzerland
[3] Univ Hosp Zurich, Dept Radiat Oncol, CH-8091 Zurich, Switzerland
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2018年 / 63卷 / 05期
基金
瑞士国家科学基金会;
关键词
proton therapy; 4D dose calculation; pencil beam scanning; liver tumour; SCANNED PARTICLE THERAPY; MOTION MITIGATION; BEAM THERAPY; EXPERIMENTAL-VERIFICATION; RESPIRATORY MOTION; LOG FILES; RECONSTRUCTION; DELIVERY; SIMULATION; INTERPLAY;
D O I
10.1088/1361-6560/aaad1e
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The aim of this study was to verify the temporal accuracy of the estimated dose distribution by a 4D dose calculation (4DDC) in comparison to measurements. A single-field plan (0.6 Gy), optimised for a liver patient case (CTV volume: 403cc), was delivered to a homogeneous PMMA phantom and measured by a high resolution scintillating-CCD system at two water equivalent depths. Various motion scenarios (no motion and motions with amplitude of 10 mm and two periods: 3.7 s and 4.4 s) were simulated using a 4D Quasar phantom and logged by an optical tracking system in real-time. Three motion mitigation approaches (single delivery, 6x layered and volumetric rescanning) were applied, resulting in 10 individual measurements. 4D dose distributions were retrospectively calculated in water by taking into account the delivery log files (retrospective) containing information on the actually delivered spot positions, fluences, and time stamps. Moreover, in order to evaluate the sensitivity of the 4DDC inputs, the corresponding prospective 4DDCs were performed as a comparison, using the estimated time stamps of the spot delivery and repeated periodical motion patterns. 2D gamma analyses and dose-difference-histograms were used to quantify the agreement between measurements and calculations for all pixels with >5% of the maximum calculated dose. The results show that a mean gamma score of 99.2% with standard deviation 1.0% can be achieved for 3%/3 mm criteria and all scenarios can reach a score of more than 95%. The average area with more than 5% dose difference was 6.2%. Deviations due to input uncertainties were obvious for single scan deliveries but could be smeared out once rescanning was applied. Thus, the deforming grid 4DDC has been demonstrated to be able to predict the complex patterns of 4D dose distributions for PBS proton therapy with high dosimetric and geometric accuracy, and it can be used as a valid clinical tool for 4D treatment planning, motion mitigation selection, and eventually 4D optimisation applications if the correct temporal information is available.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] Time-resolved dosimetry for validation of 4D dose calculation in PBS proton therapy
    Kostiukhina, N.
    Palmans, H.
    Stock, M.
    Knopf, A.
    Georg, D.
    Knaeusl, B.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (12):
  • [2] Experimental Validation of a 4D Dose Calculation for Scanned Proton Beam Therapy
    Krieger, M.
    Klimpki, G.
    Fattori, G.
    Hrbacek, J.
    Oxley, D.
    Safai, S.
    Weber, D.
    Lomax, A.
    Zhang, Y.
    [J]. MEDICAL PHYSICS, 2017, 44 (06) : 3180 - 3180
  • [3] 4D DOSE CALCULATION FOR SCANNED PROTON BEAMS USING A DEFORMING DOSE GRID AND SIMULATED 4DCT
    Boye, D.
    Lomax, T.
    Knopf, A.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S151 - S152
  • [4] Experimental validation of a 4D dose calculation routine for pencil beam scanning proton therapy
    Pfeiler, Tina
    Baeumer, Christian
    Engwall, Erik
    Geismar, Dirk
    Spaan, Bernhard
    Timmermann, Beate
    [J]. ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2018, 28 (02): : 121 - 133
  • [5] Time-resolved dosimetry for validation of 4D dose calculation in PBS proton beam radiotherapy
    Kostiukhina, N.
    Palmans, H.
    Stock, M.
    Georg, D.
    Knaeusl, B.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S122 - S123
  • [6] Time-resolved dosimetry for validation of 4D dose calculation in PBS proton therapy (vol 65, 125015, 2020)
    Kostiukhina, N.
    Palmans, H.
    Stock, M.
    Knopf, A.
    Georg, D.
    Knaeusl, B.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (20):
  • [7] Dosimetric uncertainties as a result of temporal resolution in 4D dose calculations for PBS proton therapy
    Zhang, Ye
    Huth, Isabel
    Weber, Damien C.
    Lomax, Antony J.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (12):
  • [8] An N-State 4D Convolution Dose Calculation Algorithm in Proton and Photon Therapy
    Li, Y.
    Li, X.
    Pan, X.
    Dong, L.
    Mohan, R.
    Zhang, X.
    [J]. MEDICAL PHYSICS, 2008, 35 (06) : 2751 - +
  • [9] Implementation of 4D proton therapy treatments with pencil beam scanning (PBS)
    Fracchiolla, F.
    Dionisi, F.
    Hild, S.
    Giacomelli, I.
    Lorentini, S.
    Engwall, E.
    Esposito, P. G.
    Amichetti, M.
    Schwarz, M.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S1110 - S1111
  • [10] Dosimetric Influence of Temporal Resolution Used by the PBS Proton 4D Dose Calculations
    Zhang, Y.
    Mueller, F.
    Weber, D.
    Lomax, A.
    [J]. MEDICAL PHYSICS, 2017, 44 (06) : 3276 - 3277