Intrapartum signal quality with external fetal heart rate monitoring: a two way trial of external Doppler CTG ultrasound and the abdominal fetal electrocardiogram

被引:39
|
作者
Reinhard, Joscha [1 ]
Hayes-Gill, Barrie R. [2 ]
Schiermeier, Sven [3 ]
Hatzmann, Wolfgang [3 ]
Herrmann, Eva [4 ]
Heinrich, Tomas M. [1 ]
Louwen, Frank [1 ]
机构
[1] Goethe Univ Frankfurt, Fac Med, Dept Obstet & Gynaecol, D-60590 Frankfurt, Germany
[2] Univ Nottingham, Sch Elect & Elect Engn, Nottingham NG7 2RD, England
[3] Ruhr Univ Bochum, Teaching Hosp, Marien Hosp Witten, Dept Obstet & Gynaecol,Fac Med, D-58452 Witten, Germany
[4] Goethe Univ Frankfurt, Dept Biostat & Math Models, D-60590 Frankfurt, Germany
关键词
Non-invasive fetal electrocardiogram (fECG); Doppler ultrasound cardiotocogram (CTG); Fetal heart rate (FHR); First and second stage of labour; Signal quality; CARDIOTOCOGRAPHY;
D O I
10.1007/s00404-012-2413-4
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
The objective of this study was to assess the fetal heart rate (FHR) signal quality of non-invasive abdominal fetal electrocardiogram (fECG) in comparison to the Doppler ultrasound cardiotocogram (CTG) during the first and second stage of labour. This was a prospective observational study of non-invasive fECG using five abdominally sited electrodes against the traditional Doppler ultrasound CTG probe on 144 patients. Data were analysed for signal quality before and after outlier removal. Abdominal fECG signal quality was significantly better during the first stage of labour in comparison to Doppler CTG (median fECG reliability of 95.7 % vs. median 87.3 % for Doppler, p < 0.001), whereas during second stage of labour, equivalence was demonstrated (p > 0.05). For the first and second stage of labour, fECG showed 106/135 (78.5 %) and 46/98 (46.9 %) women having fetal signal loss below 20 %, respectively. Similarly, Doppler ultrasound demonstrated 104/135 (77.0 %) and 51/98 (52.0 %) women having fetal signal loss below 20 % during first and second stage of labour, respectively. The non-invasive abdominal fECG presents an improved FHR signal quality during the first stage of labour and an equivalent signal quality during the second stage.
引用
收藏
页码:1103 / 1107
页数:5
相关论文
共 50 条
  • [11] NEW TECHNIQUE FOR IMPROVING DOPPLER ULTRASOUND SIGNAL FOR FETAL HEART-RATE MONITORING
    LAUERSEN, NH
    HOCHBERG, HM
    GEORGE, MED
    TEGGE, CS
    MEIGHAN, JJ
    AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1977, 128 (03) : 300 - 302
  • [12] Monitoring the fetal heart rate and fetal electrocardiogram: Abdominal recordings are as good as direct ECG measurements
    Vullings, R
    Peters, CHL
    Andriessen, P
    Oei, SG
    Wijn, PFF
    PEDIATRIC RESEARCH, 2005, 58 (02) : 424 - 424
  • [13] RANDOMIZED CONTROLLED TRIAL OF INTRAPARTUM FETAL HEART-RATE MONITORING
    MAHOMED, K
    NYONI, R
    MULAMBO, T
    KASULE, J
    JACOBUS, E
    BRITISH MEDICAL JOURNAL, 1994, 308 (6927): : 497 - 500
  • [14] Doppler Ultrasound Technology for Fetal Heart Rate Monitoring: A Review
    Hamelmann, Paul
    Vullings, Rik
    Kolen, Alexander F.
    Bergmans, Jan W. M.
    van Laar, Judith O. E. H.
    Tortoli, Piero
    Mischi, Massimo
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2020, 67 (02) : 226 - 238
  • [15] Fetal heart rate and maternal heart rate confusion during intrapartum monitoring: comparison of transabdominal fetal ECG and Doppler telemetry
    Signaroldi, Maria
    Mastroianni, Cristina
    Rosti, Eleonora
    Loi, Giorgia
    Di Francesco, Stefania
    Ferrazzi, Enrico
    Stampalija, Tamara
    AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 2011, 204 : S261 - S262
  • [16] Noninvasive fetal heart rate monitoring: Abdominal fetal ECG v. Doppler ultrasound cardiotocogram in obese patients
    Sagaram, Deepika
    Vani, Kavita
    Kuba, Kfier
    Lambert, Calvin
    Xie, Xianhong
    Bernstein, Peter S.
    AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 2020, 222 (01) : S403 - S404
  • [17] Added value of intrapartum recording of the maternal heart rate as an adjunct to fetal monitoring using external ultrasound transducer: not only about artifacts
    Dall'Asta, Andrea
    Volpi, Lavinia
    Morganelli, Giovanni
    Ghi, Tullio
    AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 2024, 231 (01) : e47 - e50
  • [18] An Integrated Approach for Fetal Heart Rate Estimation from Abdominal Electrocardiogram Signal
    WEI Xueyun
    ZHENG Wei
    Chinese Journal of Electronics, 2019, 28 (06) : 1198 - 1203
  • [19] An Integrated Approach for Fetal Heart Rate Estimation from Abdominal Electrocardiogram Signal
    Wei, Xueyun
    Zheng, Wei
    CHINESE JOURNAL OF ELECTRONICS, 2019, 28 (06) : 1198 - 1203
  • [20] 408 Monitoring the Fetal Heart Rate and Fetal Electrocardiogram: Abdominal Recordings are as Good as Direct ECG Measurements
    R Vullings
    C H L Peters
    P Andriessen
    S G Oei
    P F F Wijn
    Pediatric Research, 2005, 58 (2) : 424 - 424