Heat bump on a monochromator crystal measured with X-ray grating interferometry

被引:26
|
作者
Rutishauser, Simon [1 ]
Rack, Alexander [2 ]
Weitkamp, Timm [3 ]
Kayser, Yves [1 ]
David, Christian [1 ]
Macrander, Albert T. [4 ]
机构
[1] Paul Scherrer Inst, CH-5232 Villigen, Switzerland
[2] European Synchrotron Radiat Facil, F-38043 Grenoble, France
[3] Synchrotron Soleil, F-91192 Gif Sur Yvette, France
[4] Argonne Natl Lab, Argonne, IL 60439 USA
关键词
double-crystal monochromator; grating interferometry; wavefront characterization; heat bump; DIFFERENTIAL PHASE-CONTRAST; WAVE-FRONT CHARACTERIZATION; SHEARING INTERFEROMETER; DIFFRACTION GRATINGS; OPTICS; PERFORMANCE; FABRICATION; RESOLUTION; RADIATION; BEAMLINE;
D O I
10.1107/S0909049513001817
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Deformation of the first crystal of an X-ray monochromator under the heat load of a high-power beam, commonly referred to as 'heat bump', is a challenge frequently faced at synchrotron beamlines. Here, quantitative measurements of the deformations of an externally water-cooled silicon (111) double-crystal monochromator tuned to a photon energy of 17.6 keV are reported. These measurements were made using two-dimensional hard X-ray grating interferometry, a technique that enables in situ at-wavelength wavefront investigations with high angular sensitivity. The observed crystal deformations were of the order of 100 nm in the meridional and 5 nm in the sagittal direction, which lead to wavefront slope errors of up to 4 mu rad in the meridional and a few hundred nanoradians in the sagittal direction.
引用
收藏
页码:300 / 305
页数:6
相关论文
共 50 条
  • [1] A SIMPLE X-RAY CRYSTAL MONOCHROMATOR
    MILLAR, JJ
    BARNEA, Z
    JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1970, 3 (07): : 570 - &
  • [2] Sensitivity of X-ray grating interferometry
    Modregger, P.
    Pinzer, B. R.
    Thuering, T.
    Rutishauser, S.
    David, C.
    Stampanoni, M.
    OPTICS EXPRESS, 2011, 19 (19): : 18324 - 18338
  • [3] Envelope modulated x-ray grating interferometry
    Pandeshwar, Amogha
    Kagias, Matias
    Shi, Zhitian
    Stampanoni, Marco
    APPLIED PHYSICS LETTERS, 2022, 120 (19)
  • [4] Design Aspects of X-Ray Grating Interferometry
    Weitkamp, Timm
    Zanette, Irene
    Pfeiffer, Franz
    David, Christian
    INTERNATIONAL WORKSHOP ON X-RAY AND NEUTRON PHASE IMAGING WITH GRATINGS, 2012, 1466 : 84 - 89
  • [5] DOUBLE-CRYSTAL X-RAY MONOCHROMATOR
    METCHNIK, V
    JOURNAL OF SCIENTIFIC INSTRUMENTS, 1965, 42 (07): : 504 - &
  • [6] A SMALL ADJUSTABLE X-RAY CRYSTAL MONOCHROMATOR
    FORTEY, J
    COHEN, E
    JOURNAL OF SCIENTIFIC INSTRUMENTS, 1954, 31 (01): : 11 - 13
  • [7] Performance and optimization of X-ray grating interferometry
    Thuering, T.
    Stampanoni, M.
    PHILOSOPHICAL TRANSACTIONS OF THE ROYAL SOCIETY A-MATHEMATICAL PHYSICAL AND ENGINEERING SCIENCES, 2014, 372 (2010):
  • [8] Energy resolved X-ray grating interferometry
    Thuering, T.
    Barber, W. C.
    Seo, Y.
    Alhassen, F.
    Iwanczyk, J. S.
    Stampanoni, M.
    APPLIED PHYSICS LETTERS, 2013, 102 (19)
  • [9] Envelope modulated x-ray grating interferometry
    Swiss Light Source, Paul Scherrer Institute, Villigen
    5232, Switzerland
    不详
    8092, Switzerland
    Appl Phys Lett, 2022, 19
  • [10] From soft to hard X-ray with a single grating monochromator
    Cocco, D.
    Bianco, A.
    Kaulich, B.
    Schaefers, F.
    Mertin, M.
    Reichardt, G.
    Nelles, B.
    Heidemann, K. F.
    SYNCHROTRON RADIATION INSTRUMENTATION, PTS 1 AND 2, 2007, 879 : 497 - +