Long-term testing of hermetic anodically bonded glass-silicon packages

被引:21
|
作者
Harpster, TJ [1 ]
Najafi, K [1 ]
机构
[1] Univ Michigan, Ctr Wireless Integrated Microsyst, WIMS, Ann Arbor, MI 48109 USA
关键词
D O I
10.1109/MEMSYS.2002.984293
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper reviews long-term test results obtained from a series of tests on glass-Si hermetically sealed packages. Results are presented from: 1) a 6.7-year ongoing room temperature phosphate buffered saline (PBS) soak test of 4 packages. 2) a 2.3-year ongoing in-vitro 97degreesC PBS soak test of a single package; and 3) in-situ hermeticity and biocompatibility tests from 12 packages implanted in 4 guinea pigs - 3 packages implanted in each of 2 guinea pigs for I-month and another 2 guinea pigs for 19-months, and 22-months. The long-term room temperature soak test is the longest running of any micropackage reported to date.
引用
收藏
页码:423 / 426
页数:4
相关论文
共 50 条
  • [31] LONG-TERM HYDRATION EFFECT ON BONDED TITANIUM
    LOREY, R
    EDGE, M
    LANG, B
    BRANDAU, H
    [J]. JOURNAL OF DENTAL RESEARCH, 1995, 74 : 237 - 237
  • [32] ALTERATION OF INTESTINAL MICROFLORA IN LONG-TERM STAY OF MAN IN A HERMETIC CHAMBER
    SHILOV, VM
    LIZKO, NN
    BORISOVA, OK
    [J]. KOSMICHESKAYA BIOLOGIYA I MEDITSINA, 1972, 6 (01): : 78 - &
  • [33] The long-term evaluation of FRPs bonded to timber
    Toufigh, Vahab
    Yarigarravesh, Mahdireza
    Mofid, Masood
    [J]. EUROPEAN JOURNAL OF WOOD AND WOOD PRODUCTS, 2018, 76 (06) : 1623 - 1636
  • [34] The long-term evaluation of FRPs bonded to timber
    Vahab Toufigh
    Mahdireza Yarigarravesh
    Masood Mofid
    [J]. European Journal of Wood and Wood Products, 2018, 76 : 1623 - 1636
  • [35] Long-Term Performance of Adhesive Bonded Anchors
    Nilforoush, Rasoul
    Nilsson, Martin
    Soderlind, Gunnar
    Elfgren, Lennart
    [J]. ACI STRUCTURAL JOURNAL, 2016, 113 (02) : 251 - 261
  • [36] Tensile tests of micro anchors anodically bonded between Pyrex glass and aluminum thin film coated on silicon wafer
    Hu, Yu-Qun
    Zhao, Ya-Pu
    Yu, Tongxi
    [J]. MICROELECTRONICS RELIABILITY, 2008, 48 (10) : 1720 - 1723
  • [37] THE LONG-TERM CORROSION OF GLASS BY GROUNDWATER
    COX, GA
    FORD, BA
    [J]. JOURNAL OF MATERIALS SCIENCE, 1993, 28 (20) : 5637 - 5647
  • [38] Long-term evaluation of hermetically glass frit sealed silicon to Pyrex wafers with feedthroughs
    Sparks, D
    Massoud-Ansari, S
    Najafi, N
    [J]. JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 2005, 15 (08) : 1560 - 1564
  • [39] Effect of Long-Term In Vitro Testing on the Properties of Bioactive Glass-Polysulfone Composites
    Orefice, Rodrigo
    West, Jon
    LaTorre, Guy
    Hench, Larry
    Brennan, Anthony
    [J]. BIOMACROMOLECULES, 2010, 11 (03) : 657 - 665
  • [40] TESTING OF FIBROUS-GLASS PLASTICS FOR LONG-TERM STRENGTH AND CREEP IN PURE BENDING
    GLIKMAN, LA
    BERSHTEIN, VA
    [J]. INDUSTRIAL LABORATORY, 1962, 28 (04): : 494 - 500