A comparison of HDR near source dosimetry using a treatment planning system, Monte Carlo simulation; and radiochromic film

被引:3
|
作者
Amoush, Ahmad [1 ]
Luckstead, Marcus [1 ]
Lamba, Michael [1 ]
Elson, Howard [1 ]
Kassing, William [1 ]
机构
[1] Univ Cincinnati, Coll Med, Dept Radiat Oncol, Cincinnati, OH 45267 USA
关键词
HDR; TG-43U1; Radiochromic film dosimetry; MC simulation; BRACHYTHERAPY SOURCES; IR-192; RECOMMENDATIONS; BETA;
D O I
10.1016/j.meddos.2012.10.006
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
This study aimed to investigate the high-dose rate Iridium-192 brachytherapy, including near source dosimetry, of a catheter-based applicator from 0.5 mm to 1 cm along the transverse axis. Radiochromic film and Monte Carlo (MC) simulation were used to generate absolute dose for the catheter-based applicator. Results from radiochromic film and MC simulation were compared directly to the treatment planning system (TPS) based on the American Association of Physicists in Medicine Updated Task Group 43 (TG-43U1) dose calculation formalism. The difference between dose measured using radiochromic film along the transverse plane at 0.5 mm from the surface and the predicted dose by the TPS was 24% +/- 13%. The dose difference between the MC simulation along the transverse plane at 0.5 mm from the surface and the predicted dose by the TPS was 22.1% +/- 3%. For distances from 1.5 mm to 1 cm from the surface, radiochromic film and MC simulation agreed with TPS within an uncertainty of 3%. The TPS under-predicts the dose at the surface of the applicator, i.e., 0.5 mm from the catheter surface, as compared to the measured and MC simulation predicted dose. MC simulation results demonstrated that 15% of this error is due to neglecting the beta particles and discrete electrons emanating from the sources and not considered by the TPS, and 7% of the difference was due to the photon alone, potentially due to the differences in MC dose modeling, photon spectrum, scoring techniques, and effect of the presence of the catheter and the air gap. Beyond 1 mm from the surface, the TPS dose algorithm agrees with the experimental and MC data within 3%. (C) 2013 American Association of Medical Dosimetrists.
引用
收藏
页码:160 / 164
页数:5
相关论文
共 50 条
  • [41] Monte Carlo uncertainty analysis of dose estimates in radiochromic film dosimetry with single-channel and multichannel algorithms
    Vera-Sanchez, Juan Antonio
    Ruiz-Morales, Carmen
    Gonzalez-Lopez, Antonio
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2018, 47 : 23 - 33
  • [42] Comparison of graphite calorimeter dosimetry system with Monte Carlo simulation at an industrial electron beam accelerator
    Benny, P. G.
    Selvam, T. Palani
    Sarma, K. S. S.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2013, 703 : 98 - 101
  • [43] Dosimetry in contact and magnification mammography, using Monte Carlo simulation
    Koutalonis, M.
    Delis, H.
    Spyrou, G.
    Costaridou, L.
    Tzanakos, G.
    Panayiotakis, G.
    [J]. PROCEEDINGS OF THE 15TH IASTED INTERNATIONAL CONFERENCE ON APPLIED SIMULATION AND MODELLING, 2006, : 389 - +
  • [44] Monte Carlo simulation of near-source dry deposition
    McDonaldBuller, EC
    Gilbert, RB
    Liljestrand, HM
    [J]. ENVIRONMENTAL TOXICOLOGY AND RISK ASSESSMENT: MODELING AND RISK ASSESSMENT (SIXTH VOLUME), 1997, 131 : 56 - 70
  • [45] QuAArC: A 3D VMAT QA System Based On Radiochromic Film and Monte Carlo Simulation of Log Files
    Barbeiro, A. R.
    Ureba, A.
    Baeza, J. A.
    Linares, R.
    Jimenez-Ortega, E.
    Mateos, J. C.
    Velazquez, S.
    Leal Plaza, A.
    [J]. MEDICAL PHYSICS, 2015, 42 (06) : 3484 - 3484
  • [46] Establishing an individual dosing system for patients undergoing interventional transcatheter arterial embolization: Radiochromic film and Monte Carlo simulation
    Tsai, Hui-Yu
    Lai, Pei-Ling
    Li, Yang-Ying
    Tyan, Yeu-Sheng
    [J]. RADIATION MEASUREMENTS, 2011, 46 (12) : 2107 - 2110
  • [47] Comparison of dose calculation algorithms with Monte Carlo simulation for surface dosimetry
    Chow, J.
    Jiang, R.
    [J]. MEDICAL PHYSICS, 2009, 36 (09) : 4308 - 4308
  • [48] Film Dosimetry Verification of Monte Carlo Generated Electron Treatment Plans
    Enright, S.
    Asprinio, A.
    Lu, L.
    [J]. MEDICAL PHYSICS, 2014, 41 (06) : 266 - 266
  • [49] CT-based Monte Carlo simulation tool for dosimetry planning and analysis
    DeMarco, JJ
    Solberg, TD
    Smathers, JB
    [J]. MEDICAL PHYSICS, 1998, 25 (01) : 1 - 11
  • [50] DOSIMETRY COMPARISON BETWEEN TG-43 AND MONTE CARLO CALCULATIONS USING THE FREIBURG FLAP FOR SKIN HDR BRACHYTHERAPY
    Vijande, J.
    Ballester, F.
    Domingo, G.
    Ouhib, Z.
    Pujades, M. D. C.
    Perez-Calatayud, J.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S274 - S274