Characterizing the HeartMate II Left Ventricular Assist Device Outflow Using Particle Image Velocimetry

被引:7
|
作者
Rowlands, Grant W. [1 ]
Good, Bryan C. [1 ]
Deutsch, Steven [1 ]
Manning, Keefe B. [1 ,2 ]
机构
[1] Penn State Univ, Dept Biomed Engn, 205 Hallowell Bldg, University Pk, PA 16802 USA
[2] Penn State Hershey Med Ctr, Dept Surg, Hershey, PA 17033 USA
关键词
HeartMate II; Reynolds stresses; ventricular assist device; outlet cannula; particle image velocimetry; von Willebrand factor; fluid dynamics; VON-WILLEBRAND-FACTOR; SHEAR-STRESS; FLOW-FIELD; PUMP;
D O I
10.1115/1.4039822
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Ventricular assist devices (VADs) are implanted in patients with a diseased ventricle to maintain peripheral perfusion as a bridge-to-transplant or as destination therapy. However, some patients with continuous flow VADs (e.g., HeartMate II (HMII)) have experienced gastrointestinal (GI) bleeding, in part caused by the proteolytic cleavage or mechanical destruction of von Willebrand factor (vWF), a clotting glycoprotein. in vitro studies were performed to measure the flow located within the HMII outlet cannula under both steady and physiological conditions using particle image velocimetry (PIV). Under steady flow, a mock flow loop was used with the HMII producing a flow rate of 3.2 L/min. The physiological experiment included a pulsatile pump operated at 105 BPM with a ventricle filling volume of 50mL and in conjunction with the HMII producing a total flow rate of 5.0 L/min. Velocity fields, Reynolds normal stresses (RNSs), and Reynolds shear stresses (RSSs) were analyzed to quantify the outlet flow's potential contribution to vWF degradation. Under both flow conditions, the HMII generated principal Reynolds stresses that are, at times, orders of magnitude higher than those needed to unfurl vWF, potentially impacting its physiological function. Under steady flow, principal RNSs were calculated to be approximately 500 Pa in the outlet cannula. Elevated Reynolds stresses were observed throughout every phase of the cardiac cycle under physiological flow with principal RNSs approaching 1500 Pa during peak systole. Prolonged exposure to these conditions may lead to acquired von Willebrand syndrome (AvWS), which is accompanied by uncontrollable bleeding episodes.
引用
收藏
页数:13
相关论文
共 50 条
  • [1] Evaluation of patients with a HeartMate 3 left ventricular assist device using echocardiographic particle image velocimetry
    Arend F. L. Schinkel
    Sakir Akin
    Mihai Strachinaru
    Rahatullah Muslem
    Dan Bowen
    Yunus C. Yalcin
    Jasper J. Brugts
    Alina A. Constantinescu
    Olivier C. Manintveld
    Kadir Caliskan
    [J]. Journal of Ultrasound, 2021, 24 : 499 - 503
  • [2] Evaluation of patients with a HeartMate 3 left ventricular assist device using echocardiographic particle image velocimetry
    Schinkel, Arend F. L.
    Akin, Sakir
    Strachinaru, Mihai
    Muslem, Rahatullah
    Bowen, Dan
    Yalcin, Yunus C.
    Brugts, Jasper J.
    Constantinescu, Alina A.
    Manintveld, Olivier C.
    Caliskan, Kadir
    [J]. JOURNAL OF ULTRASOUND, 2021, 24 (04) : 499 - 503
  • [3] Outflow graft thrombectomy during HeartMate II left ventricular assist device exchange
    Berg, Kent B.
    Klodell, Charles T.
    [J]. INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY, 2014, 19 (04) : 714 - 715
  • [4] Migration of a HeartMate II Left Ventricular Assist Device
    Jabbar, Ali Abdul
    Etheridge, Whitson B.
    Frazier, O. H.
    [J]. CIRCULATION-HEART FAILURE, 2013, 6 (03) : E30 - E30
  • [5] Outflow Graft Occlusion of the HeartMate 3 Left Ventricular Assist Device
    Posada, Juan G. Duero
    Moayedi, Yasbanoo
    Alhussein, Mosaad
    Rodger, Marnie
    Alvarez, Juglans
    Wintersperger, Bernd J.
    Ross, Heather J.
    Butany, Jagdish
    Billia, Filio
    Rao, Vivek
    [J]. CIRCULATION-HEART FAILURE, 2017, 10 (09)
  • [6] Outflow Graft Narrowing of the HeartMate 3 Left Ventricular Assist Device
    Jain, Sneha S.
    Clerkin, Kevin J.
    Anstey, Edmund
    Liu, Qi
    Fried, Justin A.
    Raikhelkar, Jayant
    Griffin, Jan M.
    Marshall, Dylan
    Colombo, Paolo
    Yuzefpolskaya, Melana
    Topkara, Veli
    Naka, Yoshifumi
    Takeda, Koji
    Sayer, Gabriel
    Uriel, Nir
    Leb, Jay
    [J]. ANNALS OF THORACIC SURGERY, 2023, 115 (05): : 1282 - 1288
  • [7] Use of Sealed Grafts in the HeartMate II Left Ventricular Assist Device Inflow and Outflow Conduits
    Gregoric, Igor D.
    Radovancevic, Rajko
    Patel, Manish
    Fenik, Yelena
    Cohn, William E.
    Frazier, O. H.
    [J]. ASAIO JOURNAL, 2014, 60 (03) : 300 - 303
  • [8] Device exchange from HeartMate II to HeartMate 3 left ventricular assist device
    Takeda, Koji
    Takayama, Hiroo
    Sanchez, Joseph
    Cevasco, Marisa
    Yuzefpolskaya, Melana
    Colombo, Paolo C.
    Naka, Yoshifumi
    [J]. INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY, 2019, 29 (03) : 430 - 433
  • [9] Exchange of a HeartMate II left ventricular assist device with a HeartMate 3 pump
    Hanke, Jasmin S.
    Haverich, Axel
    Schmitto, Jan D.
    [J]. JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2016, 35 (07): : 944 - 946
  • [10] Predictors of Stroke on Heartmate II Left Ventricular Assist Device
    Kar, B.
    Basra, S.
    Loyalka, P.
    Gregoric, I.
    Delgado, R.
    Civitello, A.
    Bogaev, R.
    Cohn, W.
    Attallah, A.
    Wagle, R.
    Frazier, O. H.
    [J]. JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2009, 28 (02): : S244 - S245