Direct Determination of Cadmium in Grain by Solid Sampling Electrothermal Vaporization Atomic Fluorescence Spectrometry with a Tungsten Coil Trap

被引:16
|
作者
Zhang, Ying
Mao, Xuefei [1 ,2 ]
Wang, Min [1 ,2 ]
Gao, Chengling
Qi, Yuehan
Qian, Yongzhong
Tang, Xiaoyan
Zhou, Jian
机构
[1] Chinese Acad Agr Sci, Beijing 100193, Peoples R China
[2] Minist Agr, Inst Qual Standard & Testing Technol Agroprod, Key Lab Agrofood Safety & Qual, Beijing 100193, Peoples R China
关键词
Tungsten coil trap; Feed grain; Solid sampling; Electrothermal vaporization atomic fluorescence spectrometry; Cadmium; ABSORPTION-SPECTROMETRY; AGRICULTURAL SOILS; HEAVY-METALS; RICE; MASS; CD; PB; HOMOGENEITY; TOXICITY; ELEMENTS;
D O I
10.1080/00032719.2015.1052976
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A simple and rapid method was developed for the determination of cadmium in grain. Porous carbon was employed as an electrothermal vaporizer with a sampling boat for atomic fluorescence spectrometry with on-line ashing to eliminate matrix interferences. Grain was pulverized to a particle size less than 0.25mm to provide good homogeneity. Pauwels' equation was employed to determine that the minimum sample mass was approximately 2mg. Optimal conditions for ashing (90W and 50s), vaporization (110W and 20s), and argon-hydrogen flow (0.6L/min) were determined. Under the optimized conditions, the limit of quantification was 0.11 mu g/kg, and the relative standard deviation was less than 10%. Furthermore, the mean fortified recovery in grains was between 83.5 and 107.7% (n=6). To validate the method, cadmium was also determined in these samples by ICP-MS, and there were no significant differences between the results using the methods.
引用
收藏
页码:2908 / 2920
页数:13
相关论文
共 50 条
  • [41] Determination of cadmium in edible oils and fats by direct electrothermal atomic absorption spectrometry
    VanDalen, G
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1996, 11 (11) : 1087 - 1092
  • [42] Direct determination of major elements in solid plant materials by electrothermal vaporization inductively coupled plasma atomic emission spectrometry
    Masson, Pierre
    TALANTA, 2007, 71 (03) : 1399 - 1404
  • [43] Direct determination of traces of cadmium in cocaine and heroin by electrothermal atomic absorption spectrometry
    BermejoBarrera, P
    MoredaPineiro, A
    MoredaPineiro, J
    BermejoBarrera, A
    QUIMICA ANALITICA, 1997, 16 (01): : 35 - 40
  • [44] DIRECT DETERMINATION OF CADMIUM IN URINE BY ELECTROTHERMAL ATOMIZATION ATOMIC-ABSORPTION SPECTROMETRY
    HALLS, DJ
    BLACK, MM
    FELL, GS
    OTTAWAY, JM
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1987, 2 (03) : 305 - 309
  • [45] Electrothermal vaporization inductively coupled plasma atomic emission spectrometry for direct multielement analysis of solid food sample with slurry sampling
    Chen, SZ
    Lu, DB
    Xu, QY
    CANADIAN JOURNAL OF ANALYTICAL SCIENCES AND SPECTROSCOPY, 2004, 49 (05): : 290 - 295
  • [46] Assessment of Homogeneity and Minimum Sample Mass for Cadmium Analysis in Powdered Certified Reference Materials and Real Rice Samples by Solid Sampling Electrothermal Vaporization Atomic Fluorescence Spectrometry
    Mao, Xuefei
    Liu, Jixin
    Huang, Yatao
    Feng, Li
    Zhang, Lihua
    Tang, Xiaoyan
    Zhou, Jian
    Qian, Yongzhong
    Wang, Min
    JOURNAL OF AGRICULTURAL AND FOOD CHEMISTRY, 2013, 61 (04) : 848 - 853
  • [47] Direct phosphorus determination on solid plant samples by electrothermal vaporization-inductively coupled plasma atomic emission spectrometry
    Masson, Pierre
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2011, 26 (06) : 1290 - 1293
  • [48] Fast Detection of Cadmium in Chocolate by Solid Sampling Electrothermal Vaporization Atomic Absorption Spectrometry and Its Application on Dietary Exposure Risk Assessment
    Jia, Hongyu
    Li, Xue
    Lan, Guanyu
    Wang, Zhaohui
    Feng, Li
    Mao, Xuefei
    MOLECULES, 2022, 27 (19):
  • [49] Analysis of aluminium-based ceramic powders by electrothermal vaporization inductively coupled plasma atomic emission spectrometry using a tungsten coil and slurry sampling
    Lucic, M
    Krivan, V
    FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1999, 363 (01): : 64 - 72
  • [50] Analysis of aluminium-based ceramic powders by electrothermal vaporization inductively coupled plasma atomic emission spectrometry using a tungsten coil and slurry sampling
    M. Lucic
    V. Krivan
    Fresenius' Journal of Analytical Chemistry, 1999, 363 : 64 - 72