Initial Experience with the Impella Recover LP 2.5 Micro-Axial Pump in Patients Undergoing High-Risk Coronary Angioplasty

被引:0
|
作者
Thomopoulou, Sofia [1 ]
Manginas, Athanassios [1 ]
Cokkinos, Dennis V. [1 ]
机构
[1] Onassis Cardiac Surg Ctr, Dept Cardiol A, Athens 17674, Greece
关键词
Left ventricular assist device; high-risk percutaneous coronary revascularization;
D O I
暂无
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: The Impella LP 2.5 device is a new percutaneous left ventricular assist device, an intravascular micro-axial blood pump designed for short term circulatory support in conditions characterized by profoundly reduced ventricular function. Percutaneous coronary intervention (PCI) is increasingly being used in patients who have severely compromised left ventricular contractility and complex coronary lesions, including multivessel disease. Methods: Three patients (pts) underwent elective PCI with the use of the Impella system. All were very poor candidates for surgery because of severe comorbidities, with mean logistic EuroSCORE 39.3%. Echocardiography was used to rule out anatomical contraindications before implantation of the device. All pts had previous myocardial infarctions and impaired left ventricular (LV) function with LV ejection fraction < 25-30%. We evaluated the effects of the pump on cardiac output, mean pulmonary wedge pressure and mean blood pressure, using right heart catheterization. We also assessed hemoglobin, cardiac enzymes (CK, CKMB and troponin), creatinine and NT-proBNP levels before and the day after the intervention. Results: The procedure was uncomplicated, without any acute, device-related adverse events during LV support. The Impella was used for a mean time of 9.3 hours (range 2 to 24 hours) and was removed immediately after the intervention in all but one patient. The device did not induce or increase aortic valve regurgitation. Two pts needed pharmacologic inotropic support during the procedure. Mean blood pressure increased in all 3 pts without pulmonary wedge pressure reduction, presumably due to a combination of pump function and PCI-induced ischemia. The mean drop in hemoglobin levels was 1.7 g/dl. The mean increase in creatinine levels due to contrast was 1 mg/dl. NT-proBNP showed a variable response. Two pts remained stable at follow up and one expired one month later due to worsening heart failure. Conclusions: The growing heart failure population and the benefit of revascularization in patients with severe LV dysfunction have led to an increase in the number of mechanically assisted percutaneous coronary interventions. Circulatory support with the Impella LP 2.5 pump during high risk PCI is feasible, but larger studies are required to demonstrate its efficacy in these severely compromised and vulnerable patients.
引用
收藏
页码:382 / 387
页数:6
相关论文
共 50 条
  • [1] Percutaneous Impella Recover circulatory support in high-risk coronary angioplasty
    Arieti, Mario
    Pesarini, Gabriele
    Ribichini, Flavio
    CARDIOVASCULAR REVASCULARIZATION MEDICINE, 2008, 9 (04) : 269 - 274
  • [2] Utilization of the Impella 2.5 Micro-Axial Pump in Children for Acute Circulatory Support
    Dimas, V. Vivian
    Murthy, Raghav
    Guleserian, Kristine J.
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2014, 83 (02) : 261 - 262
  • [3] Use of Impella Recover LP 2.5 in elective high risk percutaneous coronary intervention
    Luis Ferreiro, Jose
    Antoni Gomez-Hospital, Joan
    Cequier, Angel R.
    Angiolillo, Dominick J.
    Roura, Gerard
    Teruel, Luis
    Maristany, Jaume
    Gomez-Lara, Josep
    Jara, Francesc
    Bass, Theodore A.
    Esplugas, Enric
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2010, 145 (02) : 235 - 237
  • [4] High-risk left main coronary stenting supported by percutaneous Impella Recover LP 2.5 assist device
    Cohen, R.
    Foucher, R.
    Sfaxi, A.
    Hakim, M.
    Domniez, T.
    Elhadad, S.
    ANNALES DE CARDIOLOGIE ET D ANGEIOLOGIE, 2010, 59 (01): : 48 - 51
  • [5] Safety and feasibility of elective high-risk percutaneous coronary intervention procedures with left ventricular support of the Impella Recover LP 2.5
    Henriques, Jose P. S.
    Sjauw, Krischan D.
    Remmelink, Maurice
    Baon, Jan, Jr.
    van der Schaaf, Rene J.
    Vis, Marije M.
    Koch, Karel T.
    Tijssen, Jan G. P.
    Piek, Jan J.
    de Winter, Robbert J.
    AMERICAN JOURNAL OF CARDIOLOGY, 2006, 98 (8A): : 88M - 88M
  • [6] High-risk coronary angioplasty protected by an Impella pump combined with simultaneous iliac artery angioplasty
    Balak, Wojciech
    Wisniewska, Joanna
    Ziolkowski, Michal
    Walukiewicz, Marcin
    Bielawski, Gabriel
    Grzesk, Grzegorz
    KARDIOLOGIA POLSKA, 2019, 77 (7-8) : 726 - 727
  • [7] Safety and feasibility of elective high-risk percutaneous coronary intervention procedures with left ventricular support of the Impella Recover LP 2.5
    Henriques, JPS
    Remmelink, M
    Baan, J
    van der Schaaf, RJ
    Vis, MM
    Koch, KT
    Scholten, EW
    de Mol, BAJM
    Tijssen, JGP
    Piek, JJ
    de Winter, RJ
    AMERICAN JOURNAL OF CARDIOLOGY, 2006, 97 (07): : 990 - 992
  • [8] High-risk percutaneous coronary intervention using the intracardiac microaxial pump 'Impella Recover'
    Ramondo, Angelo
    Napodano, Massimo
    Tarantini, Giuseppe
    Calzolari, Diego
    Nalli, Chiara
    Cacciavillani, Luisa
    Iliceto, Sabino
    JOURNAL OF CARDIOVASCULAR MEDICINE, 2006, 7 (02) : 149 - 152
  • [9] High-risk PCI in acute coronary syndromes with Impella LP 2.5 device support
    Iliodromitis, Konstantinos E.
    Kahlert, Philipp
    Plicht, Bjorn
    Hoffmann, Andreas-Claudius
    Eggebrecht, Holger
    Erbel, Raimund
    Konorza, Thomas F.
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2011, 153 (01) : 59 - 63
  • [10] Safety and feasibility of elective high-risk percutaneous coronary intervention procedures with left ventricular support of the impella recover LP 2.5 device
    Sjauw, Krischan D.
    Remmelink, Maurice
    Vis, Marije M.
    van der Schaaf, Rene J.
    Koch, Karel T.
    Baan, Jan, Jr.
    Tijssen, Jan G.
    Piek, Jan J.
    de Winter, Robbert J.
    Henriques, Jose P.
    CIRCULATION, 2006, 114 (18) : 643 - 643