Tip shape reconstruction of the probes for shear-force and near-field microscopes

被引:1
|
作者
Chuklanov, AP [1 ]
Bukharaev, AA [1 ]
机构
[1] Russian Acad Sci, Zavoisky Phys Tech Inst, Kazan 420029, Russia
关键词
scanning near-field optical microscopy; shear-force; deconvolution;
D O I
10.1002/sia.2298
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
This work is devoted to reconstructing the tip apex of the probes for shear-force and scanning near-field optical microscopes. The lateral resolution of these microscopes strongly depends on the tip radius of the corresponding probes. The transmission electron microscope (TEM) is used for estimating the tip radius as a rule. However, it is rather difficult to use TEM in the routine experiments. We present the method for determination of the tip shape using the special test samples. Two types of tips have been used: the homemade electrochemically etched W microwires and the commercial fiber probes. The estimation of the tip radius was made from the tip apex, which was extracted by our deconvolution method from the corresponding experimental images obtained on a shear-force microscope. Additional TEM experiments allowed us to compare the results obtained by these two independent methods. Correlation was investigated between the scanning near-field optical microscopy (SNOM) lateral resolution and the reconstructed tip apex of the fiber probe used in the corresponding experiment. Copyright (C) 2006 John Wiley & Sons, Ltd.
引用
收藏
页码:587 / 589
页数:3
相关论文
共 50 条
  • [31] Assessing Probe Damage in Constant Frequency and Frequency-Modulation Shear-force Acoustic Near-field Microscopy
    Brockman, Theodore
    Bai, Jianghua
    Fernandez, Rodoflo
    La Rosa, Andres H.
    2018 IEEE 13TH NANOTECHNOLOGY MATERIALS AND DEVICES CONFERENCE (NMDC), 2018, : 365 - 368
  • [32] High-sensitivity piezoelectric tube sensor for shear-force detection in scanning near-field optical microscopy
    Lindfors, K
    Kapulainen, M
    Ryytty, P
    Kaivola, M
    OPTICS AND LASER TECHNOLOGY, 2004, 36 (08): : 651 - 656
  • [33] Tip-sample distance control for near-field scanning microwave microscopes
    Kim, MS
    Kim, S
    Kim, J
    Lee, K
    Friedman, B
    Kim, JT
    Lee, J
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2003, 74 (08): : 3675 - 3678
  • [34] PIEZOELECTRIC TIP-SAMPLE DISTANCE CONTROL FOR NEAR-FIELD OPTICAL MICROSCOPES
    KARRAI, K
    GROBER, RD
    APPLIED PHYSICS LETTERS, 1995, 66 (14) : 1842 - 1844
  • [35] KINETIC LIMITS FOR SENSING TIP MORPHOLOGY IN NEAR-FIELD SCANNING OPTICAL MICROSCOPES
    YAKOBSON, BI
    MOYER, PJ
    PAESLER, MA
    JOURNAL OF APPLIED PHYSICS, 1993, 73 (11) : 7984 - 7986
  • [36] A SIMPLE TECHNIQUE FOR THE MANUFACTURE OF OPTICAL PROBES FOR SCANNING NEAR-FIELD OPTICAL MICROSCOPES
    SCHOCH, B
    JONES, BE
    FRANKS, A
    MEASUREMENT SCIENCE AND TECHNOLOGY, 1994, 5 (06) : 663 - 666
  • [37] First overtone frequency stimulated quartz tuning fork used for shear-force scanning near-field optical microscopy
    Liu, S
    Sun, JL
    Sun, HS
    Tan, XJ
    Shi, S
    Guo, JH
    Zhao, J
    CHINESE PHYSICS LETTERS, 2003, 20 (11) : 1928 - 1931
  • [39] A shear force feedback control system for near-field scanning optical microscopes without lock-in detection
    Hsu, JWP
    McDaniel, AA
    Hallen, HD
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1997, 68 (08): : 3093 - 3095
  • [40] Bent-fiber near-field scanning optical microscopy probes for use with commercial atomic-force microscopes
    Taylor, RS
    Leopold, KE
    Wendman, M
    Gurley, G
    Elings, V
    MICROMACHINING AND IMAGING, 1997, 3009 : 119 - 129