Numerical jitter estimation for swept source optical coherence tomography

被引:1
|
作者
Zhu, Lida [1 ]
Miyazawa, Arata [1 ]
Mukherjee, Pradipta [1 ]
Abd El-Sadek, Ibrahim [1 ]
Oikawa, Kensuke [1 ]
Oida, Daisuke [1 ]
Yasuno, Yoshiaki [1 ]
机构
[1] Univ Tsukuba, Computat Opt Grp, 1-1-1 Tennodai, Tsukuba, Ibaraki 3058573, Japan
关键词
Optical coherence tomography; swept source; jitter; numerical; phase stabilization; HUMAN SKIN;
D O I
10.1117/12.2573281
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Due to asynchronization between the acquisition trigger and K-clock trigger in a swept source optical coherence tomography (SS-OCT) system, trigger jitter causes the spectrum a temporal shift in the spectral domain and thus corrupts the measurement. We study ternary distribution of the jitter signal by measuring TiO2 phantom using a SS-OCT system, and it shows one-pixel spectral shift in the spectral domain.
引用
收藏
页数:3
相关论文
共 50 条
  • [11] Retinal applications of swept source optical coherence tomography (OCT) and optical coherence tomography angiography (OCTA)
    Lains, Ines
    Wang, Jay C.
    Cui, Ying
    Katz, Raviv
    Vingopoulos, Filippos
    Staurenghi, Giovanni
    Vavvas, Demetrios G.
    Miller, Joan W.
    Miller, John B.
    [J]. PROGRESS IN RETINAL AND EYE RESEARCH, 2021, 84
  • [12] Novel perspectives on swept-source optical coherence tomography
    Lavinsky F.
    Lavinsky D.
    [J]. International Journal of Retina and Vitreous, 2 (1)
  • [13] Swept source optical coherence tomography in eyes with macular hemorrhage
    Oyamada, Hiroshi
    Maruko, Ichiro
    Sugano, Yukinori
    Sekiryu, Tetsuju
    Iida, Tomohiro
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2013, 54 (15)
  • [14] Gastric Myxoma Imaging with Swept Source Optical Coherence Tomography
    Luo, Site
    Huo, Li
    Liu, Hao
    Li, Xinguo
    An, Xin
    Zhao, Hui
    Xie, Huikai
    [J]. 2017 CONFERENCE ON LASERS AND ELECTRO-OPTICS EUROPE & EUROPEAN QUANTUM ELECTRONICS CONFERENCE (CLEO/EUROPE-EQEC), 2017,
  • [15] Ocular biometry with swept-source optical coherence tomography
    Montes-Mico, Robert
    Pastor-Pascual, Francisco
    Ruiz-Mesa, Ramon
    Tana-Rivero, Pedro
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2021, 47 (06): : 802 - 814
  • [16] Optimizing numerical k-sampling for swept-source optical coherence tomography angiography
    Huang, Naixing
    Hormel, Tristan t.
    Liang, Guangru b.
    Wei, Xiang
    Guo, Yukun
    Chen, Siyu
    Jia, Yali
    [J]. OPTICS LETTERS, 2024, 49 (05) : 1201 - 1204
  • [17] Swept source optical coherence tomography artifacts in glaucoma patients
    Lee, S. Y.
    Seo, S. J.
    Lee, Y. H.
    Bae, H. Y.
    Kim, C. Y.
    [J]. ACTA OPHTHALMOLOGICA, 2014, 92
  • [18] High axial resolution swept source for optical coherence tomography
    Cordes, A. H.
    Xavier, G. B.
    de Faria, G. Vilela
    von der Weid, J. P.
    [J]. ELECTRONICS LETTERS, 2010, 46 (01) : 27 - 28
  • [19] Reproducibility of the measurements taken with swept source optical coherence tomography
    Orduna Hospital, E.
    Vilades Palomar, E.
    Cipres, M.
    Obis, J.
    Rodrigo, M. J.
    Satue, M.
    Garcia-Martin, E.
    [J]. ACTA OPHTHALMOLOGICA, 2017, 95
  • [20] SPECTRAL SIGNAL PROCESSING IN SWEPT SOURCE OPTICAL COHERENCE TOMOGRAPHY
    Chang, Shoude
    Mao, Youxin
    Flueraru, Costel
    [J]. PHOTONICS NORTH 2010, 2010, 7750