Biological fixation of comminuted subtrochanteric fractures with proximal femur locking compression plate

被引:37
|
作者
Saini, Pramod [1 ]
Kumar, Rakesh [1 ]
Shekhawat, Vishal [2 ]
Joshi, Narendra [2 ]
Bansal, Mahesh [2 ]
Kumar, Senthil [1 ]
机构
[1] All India Inst Med Sci, Dept Orthopaed, New Delhi 110029, India
[2] Sawai Man Singh Med Coll & Hosp, Dept Orthopaed, Jaipur 302004, Rajasthan, India
关键词
Proximal femur-locking compression plate; Subtrochanteric fractures; Biological fixation; ANGLED BLADE PLATE; DYNAMIC CONDYLAR SCREW; FEMORAL NAIL PFN; LONG-GAMMA-NAIL; INTRAMEDULLARY FIXATION; INDIRECT REDUCTION; INTERNAL-FIXATION; HIP SCREW; OSTEOSYNTHESIS; EXTRAMEDULLARY;
D O I
10.1016/j.injury.2012.10.037
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Introduction: Subtrochanteric fractures are difficult fractures associated with high incidence of complications and various implants, both intramedullary and extramedullary, are available for their fixation. Traditional extramedullary implants are associated with higher rate of implant failure and varus collapse while the biomechanically better intramedullary nails are technically demanding and are associated with higher re-operation rates. This study was done to evaluate the outcome following biological (indirect) fixation of unstable comminuted subtrochanteric fractures with proximal femur-locking compression plate (PF-LCP). Methods: Thirty-five consecutive patients with comminuted subtrochanteric fractures were operated upon with PF-LCP by using an indirect reduction technique. Seinsheimer types 3-5 fractures were included in the study. Operating time, blood loss and any technical difficulty with the implant were recorded. Patients were followed clinically and radiologically for union at fracture site and implant-related complications. The Harris Hip Score was used to document hip function at final follow-up. Results: Thirty-two patients with average age of 44.7 years were available for final evaluation. The mean operating time was 79.5 min and total blood loss averaged 233.13 ml. Union was achieved in all cases with an average time of 15.62 weeks. Complications included two cases of delayed union and two cases of infection. Two cases had a shortening of 1 cm and one case had malunion with external rotation. No instances of implant failure or nonunion were recorded. Conclusion: Biological fixation of comminuted subtrochanteric fractures with PF-LCP provides stable fixation with high union rate and fewer complications. (C) 2012 Elsevier Ltd. All rights reserved.
引用
收藏
页码:226 / 231
页数:6
相关论文
共 50 条
  • [31] BENT PLATE FIXATION IN COMBINED INTERTROCHANTERIC AND SUBTROCHANTERIC FRACTURES OF THE FEMUR
    ROWE, SM
    CHUNG, JY
    MOON, ES
    SONG, EK
    ORTHOPEDICS, 1991, 14 (10) : 1123 - 1128
  • [32] SUBTROCHANTERIC FRACTURES OF FEMUR - TREATMENT WITH ASIF BLADE PLATE FIXATION
    WHATLEY, JR
    GARLAND, DE
    WHITECLOUD, T
    WICKSTROM, J
    SOUTHERN MEDICAL JOURNAL, 1978, 71 (11) : 1372 - 1375
  • [33] Locking Plate Fixation of Proximal Humerus Fractures
    Strohm, Peter C.
    Koestler, Wolfgang
    Suedkamp, Norbert P.
    TECHNIQUES IN SHOULDER AND ELBOW SURGERY, 2005, 6 (01): : 8 - 13
  • [34] Locking Plate Fixation for Proximal Humerus Fractures
    Burke, Neil G.
    Kennedy, Jim
    Green, Connor
    Dodds, Michael K.
    Mullett, Hannan
    ORTHOPEDICS, 2012, 35 (02) : E250 - E254
  • [35] Locking Compression Pilon Plate for Fixation of Comminuted Posterior Wall Acetabular Fractures: A Novel Technique
    Lee, Adam K.
    Wagner, Benjamin R.
    McPhillips, Kristin
    Horwitz, Daniel S.
    Widmaier, James C., Jr.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2017, 31 (01) : E32 - E36
  • [36] Application of proximal femur locking compression plate for unstable four part intertrochanteric fractures
    夏青
    外科研究与新技术, 2011, 20 (02) : 108 - 108
  • [37] Functional and Radiological Outcomes of Unstable Proximal Femur Fractures Fixed With Anatomical Proximal Locking Compression Plate
    Krishna, Vamsee
    Venkatesan, Aakaash
    Singh, Ayush Kumar
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2022, 14 (05)
  • [38] Indirect reduction and ''biological'' fixation of comminuted metaphyseal fractures of the femur
    Baumgaertel, F
    Hessmann, M
    LANGENBECKS ARCHIV FUR CHIRURGIE, 1996, : 1180 - 1180
  • [39] Minimally invasive plate osteosynthesis of subtrochanteric femur fractures with a locking plate: a prospective series of 20 fractures
    Oh, Chang-Wug
    Kim, Jung-Jae
    Byun, Young-Soo
    Oh, Jong-Keon
    Kim, Joon-Woo
    Kim, Shin-Yoon
    Park, Byung-Chul
    Lee, Hyun-Joo
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2009, 129 (12) : 1659 - 1665
  • [40] Minimally invasive plate osteosynthesis of subtrochanteric femur fractures with a locking plate: a prospective series of 20 fractures
    Chang-Wug Oh
    Jung-Jae Kim
    Young-Soo Byun
    Jong-Keon Oh
    Joon-Woo Kim
    Shin-Yoon Kim
    Byung-Chul Park
    Hyun-Joo Lee
    Archives of Orthopaedic and Trauma Surgery, 2009, 129 : 1659 - 1665