Micro-imaging system using scanning DC-SQUID microscope

被引:68
|
作者
Morooka, T [1 ]
Nakayama, S [1 ]
Odawara, A [1 ]
Ikeda, M [1 ]
Tanaka, S [1 ]
Chinone, K [1 ]
机构
[1] Seiko Instruments Inc, Matsudo, Chiba 2702222, Japan
关键词
D O I
10.1109/77.783782
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A micro-imaging system in a low temperature environment has been developed for the study of superconducting films and magnetic films and for the inspection of superconducting integrated circuits. The system consists of a micro DC-SQUID, a cryostat, a precise scanning stage, and a computer. Two different types of micro DC-SQUIDs were designed. One was a magnetometer (Bz) with a one-turn pick-up coil with a diameter of 10 mu m, and the other was a gradiometer (dBz/dx) with a planar first-order derivative pick-up coil. Each micro DC-SQUID was integrated on a 3mm x 3mm Si chip using thin Nb film fabrication technology. Preliminary experiments were made using the system and several magnetic images were obtained. We present observations of a thin superconducting Nb film pattern by applying the Meissner screening and the magnetic domains of a thin garnet ((YBi)(3)(FeAl)(5)O-12) film.
引用
收藏
页码:3491 / 3494
页数:4
相关论文
共 50 条
  • [1] Scanning DC-SQUID system with high spatial resolution for NDE
    Chinone, K
    Nakayama, S
    Morooka, T
    Odawara, A
    Ikeda, M
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 1997, 7 (02) : 3271 - 3274
  • [2] MULTICHANNEL DC-SQUID SYSTEM
    FURUKAWA, H
    SHIRAE, K
    JAPANESE JOURNAL OF APPLIED PHYSICS PART 2-LETTERS, 1989, 28 (03): : L456 - L458
  • [3] IMAGING RADIOFREQUENCY FIELDS USING A SCANNING SQUID MICROSCOPE
    BLACK, RC
    WELLSTOOD, FC
    DANTSKER, E
    MIKLICH, AH
    KOELLE, D
    LUDWIG, F
    CLARKE, J
    APPLIED PHYSICS LETTERS, 1995, 66 (10) : 1267 - 1269
  • [4] DC-SQUID SENSOR SYSTEM FOR MULTICHANNEL NEUROMAGNETOMETRY
    HOUWMAN, EP
    VELDHUIS, D
    FLOKSTRA, J
    TERBRAKE, HJM
    JASZCZUK, W
    MARTINEZ, A
    ROGALLA, H
    IEEE TRANSACTIONS ON MAGNETICS, 1991, 27 (02) : 2955 - 2958
  • [5] Development of mice biomagnetic measurement system using dc-SQUID magnetometer
    Ono, Y
    Ishiyama, A
    Kasai, N
    Odawara, A
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2005, 15 (02) : 640 - 643
  • [6] Low temperature NMR measurements using a DC-SQUID
    Wurm, M
    Brison, JP
    Flouquet, J
    CZECHOSLOVAK JOURNAL OF PHYSICS, 1996, 46 : 2851 - 2852
  • [7] Prospects for imaging magnetic nanoparticles using a scanning SQUID microscope
    Kirtley, J. R.
    SUPERCONDUCTOR SCIENCE & TECHNOLOGY, 2009, 22 (06):
  • [8] A 37-CHANNEL DC-SQUID MAGNETOMETER SYSTEM
    DRUNG, D
    ZIMMERMANN, R
    CANTOR, R
    ERNE, SN
    KOCH, H
    MATTHIES, KP
    PETERS, M
    SCHEER, HJ
    STOLLFUSS, D
    CLINICAL PHYSICS AND PHYSIOLOGICAL MEASUREMENT, 1991, 12 : 21 - 29
  • [9] High temperature superconductor dc-SQUID microscope with a soft magnetic flux guide
    Poppe, U
    Faley, MI
    Zimmermann, E
    Glaas, W
    Breunig, I
    Speen, R
    Jungbluth, B
    Soltner, H
    Halling, H
    Urban, K
    SUPERCONDUCTOR SCIENCE & TECHNOLOGY, 2004, 17 (05): : S191 - S195
  • [10] MULTICHANNEL DC-SQUID SYSTEM WITH 2-STAGE CONFIGURATION
    FURUKAWA, H
    FUJITA, S
    MATANI, A
    YOSHIDA, M
    SHIRAE, K
    PHYSICA B, 1990, 165 : 99 - 100