Miniarc® single-incision sling for treatment of stress urinary incontinence: 2-year clinical outcomes

被引:28
|
作者
Kennelly, Michael J. [1 ,2 ]
Moore, Robert [3 ]
Nguyen, John N. [4 ]
Lukban, James [5 ]
Siegel, Steven [6 ]
机构
[1] McKay Dept Urol, Charlotte, NC 28207 USA
[2] Carolinas Med Ctr, Dept Urol, Charlotte, NC 28203 USA
[3] Atlanta Urogynecol Associates, Alpharetta, GA USA
[4] So Calif Permanente Med Grp, Downey, CA USA
[5] Eastern Virginia Med Sch, Norfolk, VA 23501 USA
[6] Metro Urol, St Paul, MN USA
关键词
Single-incision sling; MiniArc; Stress incontinence; FREE VAGINAL TAPE; FOLLOW-UP;
D O I
10.1007/s00192-012-1734-y
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
We report 2-year data on the effectiveness and safety of the MiniArc single-incision sling in women with stress urinary incontinence. This multi-center, prospective, single-arm, industry-sponsored study measured the effectiveness of the MiniArc sling via quantitative (cough stress test and 1-h pad weight test) and qualitative (Urogenital Distress Inventory-Short Form and Incontinence Impact Questionnaire-Short Form) measurements. The objective efficacy rate was defined as the number of patients with a negative cough stress test or 1-h pad weight test a parts per thousand currency signaEuro parts per thousand 1 g at 2 years. The subjective efficacy rate was determined by patient responses to the UDI-6 question # 3, "Do you experience, and if so, how much are you bothered by urine leakage related to physical activity, coughing, or sneezing?" Secondary objectives were to evaluate procedural variables of implantation and long-term safety. One hundred and eighty women with a mean age of 51.1 years were implanted in the study. Mean procedure time, blood loss, and length of stay were 11.0 min, 41.7 mL and 9.5 h respectively. At 2 years, 142 patients were available for analysis. The objective efficacy rates for the cough stress test (CST) and pad weight test (PWT) were 84.5 % and 80.1 % respectively and the subjective efficacy rate was 92.9 %. Median Urogenital Distress Inventory-Short Form and Incontinence Impact Questionnaire-Short Form scores showed statistically significant improvement (p < .001). The most common adverse events included UTI (4.8 %), constipation (3.7 %), and temporary urinary retention (3.2 %). MiniArc is a safe and effective surgical procedure for the treatment of SUI in women with follow-up through 2 years.
引用
收藏
页码:1285 / 1291
页数:7
相关论文
共 50 条
  • [31] Comparison of the clinical outcomes of transobturator and single-incision slings for stress urinary incontinence
    Wu, Ling-Ying
    Yang, Tsai-Hwa
    Kung, Fu-Tsai
    Chuang, Fei-Chi
    Huang, Kuan-Hui
    [J]. KAOHSIUNG JOURNAL OF MEDICAL SCIENCES, 2016, 32 (07): : 367 - 372
  • [32] Risk factors for stress urinary incontinence recurrence after single-incision sling
    Palmieri, Stefania
    Frigerio, Matteo
    Spelzini, Federico
    Manodoro, Stefano
    Milani, Rodolfo
    [J]. NEUROUROLOGY AND URODYNAMICS, 2018, 37 (05) : 1711 - 1716
  • [33] LONG TERM FOLLOW-UP DATA ON THE MINIARC™ SINGLE INCISION SLING SYSTEM FOR THE TREATMENT OF STRESS URINARY INCONTINENCE
    Pickens, Ryan
    Hardin, Brent
    Beddies, John
    White, Wesley
    Waters, W. Bedford
    Klein, Frederick
    [J]. JOURNAL OF UROLOGY, 2011, 185 (04): : E534 - E534
  • [34] The MiniArc® Sling System in the Treatment of Female Stress Urinary Incontinence
    Gauruder-Burmester, Annett
    Popken, Gralf
    [J]. INTERNATIONAL BRAZ J UROL, 2009, 35 (03): : 334 - 341
  • [35] LONG TERM FOLLOW-UP DATA ON THE MINIARC™ SINGLE INCISION SLING SYSTEM FOR THE TREATMENT OF STRESS URINARY INCONTINENCE
    Pickens, Ryan
    Stewart, Adam
    White, Wesley
    Mobley, Joe
    Klein, Frederick
    [J]. NEUROUROLOGY AND URODYNAMICS, 2011, 30 (02) : 261 - 261
  • [36] Two-year outcomes for the Altis® adjustable single incision sling system for treatment of stress urinary incontinence
    Kocjancic, Ervin
    Erickson, Ty
    Tu, Le-Mai
    Gheiler, Edward
    Van Drie, Douglas
    [J]. NEUROUROLOGY AND URODYNAMICS, 2017, 36 (06) : 1582 - 1587
  • [37] Transobturator Sling Compared With Single-Incision Mini-Sling for the Treatment of Stress Urinary Incontinence A Randomized Controlled Trial
    Djehdian, Lucyana M.
    Araujo, Maita P.
    Takano, Claudia C.
    Del-Roy, Carlos A.
    Sartori, Marair G. F.
    Girao, Manoel J. B. C.
    Castro, Rodrigo A.
    [J]. OBSTETRICS AND GYNECOLOGY, 2014, 123 (03): : 553 - 561
  • [38] Single-incision mini-sling and trans-obturator sling for stress urinary incontinence: A 5-year comparison
    Ruffolo, Alessandro F.
    Bonavina, Giulia
    Serati, Maurizio
    Casiraghi, Arianna
    Degliuomini, Rebecca
    Parma, Marta
    Porcari, Irene
    Candiani, Massimo
    Salvatore, Stefano
    [J]. EUROPEAN JOURNAL OF OBSTETRICS & GYNECOLOGY AND REPRODUCTIVE BIOLOGY, 2022, 270 : 90 - 94
  • [39] The Comparison of Adjustable Single-incision Mini-Sling and Transobturator Tape for the Treatment of Stress Urinary Incontinence
    Karatas, Suat
    Tekelioglu, Meltem
    Cift, Tayfur
    Temur, Muzaffer
    Yumru, Ayse Ender
    [J]. JOURNAL OF ACADEMIC RESEARCH IN MEDICINE-JAREM, 2020, 10 (01): : 36 - 40
  • [40] Prospective study of a single-incision sling versus a transobturator sling in women with stress urinary incontinence: 3-year results
    White, Amanda B.
    Kahn, Bruce S.
    Gonzalez, Ricardo R.
    Rosamilia, Anna
    Anger, Jennifer T.
    Eilber, Karyn S.
    Schaffer, Joseph, I
    [J]. AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 2020, 223 (04)