Test of the accuracy of the computerized glow curve deconvolution algorithm for the analysis of thermoluminescence glow curves

被引:17
|
作者
Sadek, A. M. [1 ]
机构
[1] Natl Inst Stand, Ionizing Radiat Metrol Dept, Giza, Egypt
关键词
CGCD; Thermoluminescence Glow Curve Analysis; Trap Parameters Evaluation Methods; GENERAL-ORDER KINETICS; QUARTZ; PEAK; TRAP; IRRADIATION; PARAMETERS; TLD-100; MODEL; TOOL; GCA;
D O I
10.1016/j.nima.2013.02.011
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The accuracy of the thermoluminescence (TL) kinetics parameters obtained using the computerized glow curve deconvolution (CGCD) algorithm was tested. The differential equation governing the electron traffic in the one trap-one recombination (OTOR) level model was solved with almost no approximation using the Fehlberg-Runge-Kutta method. A set of simulated glow peaks was generated for a wide range of kinetics parameter values. These glow peaks were then fitted using the general-order kinetics equation. Comparisons between the kinetics parameter values of the simulated glow peaks and those obtained by the CGCD method were made. The results show that the accuracy of the different kinetics parameters obtained by the CGCD method is not the same and that it varies according to the value of the kinetics order (b). The overlapping of two glow peaks with very close maximum peak positions (T-ms) results in a glow peak with unexpected values for the kinetics parameters. A set of different cases of overlapping glow peaks is also discussed. (c) 2013 Elsevier B.V. All rights reserved.