Tissue composition effect on dose distribution in radiotherapy with a 6 MV photon beam of a medical linac

被引:3
|
作者
Ghorbani, Mahdi [1 ]
Noghreiyan, Atefeh Vejdani [2 ]
Tabatabaei, Zahra Sadat [2 ]
Pakravan, Delaram [3 ]
Davenport, David [4 ]
机构
[1] Shahid Beheshti Univ Med Sci, Fac Med, Dept Biomed Engn & Med Phys, Tehran, Iran
[2] Mashhad Univ Med Sci, Phys Res Ctr, Mashhad, Iran
[3] Islamic Azad Univ, Fac Basic Sci, Dept Phys, Ahvaz, Iran
[4] Comprehens Canc Ctr Nevada, Las Vegas, NV USA
关键词
Dose distribution; photon beam; radiotherapy; tissue composition; MONTE-CARLO; ELECTRON-BEAMS; BRACHYTHERAPY; ENHANCEMENT;
D O I
10.4103/jcrt.JCRT_706_16
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Aim: The aim of this study is to evaluate soft-tissue composition effect on dose distribution for various soft tissues in radiotherapy with a 6 MV photon beam of a medical linac. Background: The compositions of various soft tissues are different which could affect dose calculations. Materials and Methods: A phantom and Siemens Primus linear accelerator were simulated using MCNPX Monte Carlo code. In a homogeneous cubic phantom, six types of soft tissue and three types of tissue-equivalent materials were defined separately. The soft tissues were muscle (skeletal), adipose tissue, blood (whole), breast tissue, soft tissue (9-component), and soft tissue (4-component). The tissue-equivalent materials included water, A-150 tissue equivalent plastic and perspex. Photon dose relative to dose in 9-component soft tissue at various depths on the beam's central axis was determined for the 6 MV photon beam. The relative dose was also calculated and compared for various MCNPX tallies including *F8, F6, and *F4. Results: The results of the relative photon dose in various materials relative to dose in 9-component soft tissue using different tallies are reported in the form of tabulated data. Minor differences between dose distributions in various soft tissues and tissue-equivalent materials were observed. The results from F6 and F4 were practically the same but differ with the *F8 tally. Conclusions: Based on the calculations performed, the differences in dose distributions in various soft tissues and tissue-equivalent materials are minor but they could be corrected in radiotherapy calculations to upgrade the accuracy of the dosimetric calculations.
引用
收藏
页码:237 / 244
页数:8
相关论文
共 50 条
  • [41] Integral dose in external beam photon radiotherapy
    Mackie, T
    Kissick, M
    Jeraj, R
    Tome, W
    Fenwick, J
    Olivera, G
    Ritter, M
    MEDICAL PHYSICS, 2004, 31 (06) : 1721 - 1721
  • [42] Monte Carlo Simulation of 6 MV Flattening Filter Free Photon Beam of TrueBeam STx LINAC at Songklanagarind Hospital
    Efendi, M. Arif
    Funsian, Amporn
    Chittrakarn, Thawat
    Bhongsuwan, Tripob
    SAINS MALAYSIANA, 2017, 46 (09): : 1407 - 1411
  • [44] Determination of initial electron parameters by means of Monte Carlo simulations for the Siemens Artiste Linac 6 MV photon beam
    Tugrul, Taylan
    Erogul, Osman
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2019, 24 (04) : 331 - 337
  • [45] A theoretical multileaf collimator model for fast Monte Carlo dose calculation of linac 6/10 MV photon beams
    Aboulbanine, Zakaria
    El Khayati, N.
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2019, 5 (05):
  • [46] Dose-Free Monitoring of Radiotherapy Treatments with Scattered Photons: First Experimental Results at a 6-MV Linac
    Simoes, Hugo
    Pinto, Marco
    Cunha, Micaela
    Goncalves, Joana
    Sampaio, Liliana
    Ferreira, Ricardo J.
    Saraiva, Henrique M.
    Rita Barbeiro, Ana
    Capela, Miguel
    Ferreira, Brigida
    Fonte, Paulo
    Ghithan, Sharif
    Leal Plaza, Antonio
    Lopes, Maria do Carmo
    Martins, Paulo
    Crespo, Paulo
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 4296 - 4302
  • [47] Dose-Free Monitoring of Radiotherapy Treatments With Scattered Photons: First Experimental Results at a 6-MV Linac
    Simoes, Hugo
    Cunha, Micaela
    Pinto, Marco
    Goncalves, Joana
    Sampaio, Liliana
    Ferreira, Ricardo J.
    Saraiva, Henrique M.
    Rita Barbeiro, Ana
    Capela, Miguel
    Ferreira, Brigida
    Fonte, Paulo
    Ghithan, Sharif
    Leal Plaza, Antonio
    Lopes, Maria do Carmo
    Martins, Paulo
    Crespo, Paulo
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2013, 60 (04) : 3110 - 3118
  • [48] CALCULATION AND MEASUREMENT OF THE DOSE AT POINTS OUTSIDE THE PRIMARY BEAM FOR PHOTON ENERGIES OF 6-MV, 10-MV, AND 23-MV
    VANDERGIESSEN, PH
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1994, 30 (05): : 1239 - 1246
  • [49] CHARACTERISTICS OF AN MM-22 MEDICAL MICROTRON 6-MV PHOTON-BEAM
    GEORGE, RE
    MEDICAL PHYSICS, 1984, 11 (06) : 862 - 865
  • [50] Commissioning a superposition dose calculation model for 6 MV photon beams on a radiotherapy treatment planning system
    Ambolt, P.
    Bowden, J.
    Cooper, M.
    Henry, A.
    McCormack, S.
    Morgan, A.
    Patrick, S.
    Thompson, R.
    Thwaites, D.
    CLINICAL ONCOLOGY, 2007, 19 (03) : S40 - S40