Quantification of Artifacts in MR Images Caused by Commonly Used Dental Materials

被引:2
|
作者
Valizadeh, Solmaz [1 ]
Pouraliakbar, Hamidreza [2 ]
Arzani, Vida [1 ]
Nahardani, Ali [3 ]
机构
[1] Shahid Beheshti Univ Med Sci, Sch Dent, Dept Radiol, Tehran, Iran
[2] Iran Univ Med Sci, Rajaie Cardiovasc Med & Res Ctr, Tehran, Iran
[3] Shahid Beheshti Univ Med Sci, Sch Allied Med Sci, Tehran, Iran
关键词
Magnetic Resonance Imaging; Dental Materials; Artifact; MAGNETIC-RESONANCE IMAGES; ORTHODONTIC APPLIANCES; IMAGING ARTIFACTS; SUSCEPTIBILITY; OBJECTS; JOINT;
D O I
10.5812/iranjradiol.55566
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Although new approaches for application of magnetic resonance imaging (MRI) in dentistry have been promising, the presence of dental materials in the oral cavity represents a major obstacle to generating quality MRI data. The resulting artifacts can adversely affect diagnostic accuracy of oral and maxillofacial conditions. Objectives: This study aimed to detect and assess artifacts caused by metallic and non-metallic dental materials in MR images. Materials and Methods: Nineteen commonly used dental materials were used in an in vitro assay. An empty cube-shaped phantom with internal dimensions of 15 x 15 x 15 cm(3) was fabricated of polyester and filled with paraffin or distilled water to create optimal contrast for T1-weighted (T1W; for anatomical structures), T2-weighted (T2W, for pathological conditions) and proton density-weighted (PDW) MR images. Dental material samples were manufactured and placed at the center of the mold using polyester tubes. All samples were then subjected to MRI using two-dimensional (2D) and three-dimensional (3D) T1W, T2W and PDW imaging sequences. The data were subjected to one-way analysis of variance (ANOVA) followed by mean comparison using the Tukey's test. Results: Comparison of mean artifact size in 2D and 3D images showed that ceramic caused the smallest (189.48 mm(3)) and orthodontic metal brackets caused the largest (4307.5 mm(3)) artifact areas. Assessment of the amount of artifacts created in different imaging sequences revealed minimal artifact in inversion recovery sequence and maximum artifact in gradient echo sequence of the amount of artifacts at different weighing times in 2D and 3D images showed minimal artifacts in PDW and maximum artifacts in T2W. Conclusion: Bracket, titanium implants, and the nickel-chromium (Ni-Cr) materials are the three most susceptible materials to external B0 magnetic field; thus creating huge metal-induced artifacts. The Siemens default two dimensional turbo spin echo (2D-TSE), 2D-BLADE, three dimensional volumetric interpolated breath-hold examination (3D-VIBE) and three dimensional sampling perfection with application optimized contrasts using different flip angle evolution (3D-SPACE) are the most appropriate pulse sequences for the 2D-T1W, 2D-T2W, 3D-T1W, and 3D-T2W evaluations of the bracket/titanium/Ni-Cr materials, respectively. The rest of the materials' magnetic susceptibility effects are negligible, and artifacts are small.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] ARTIFACTS CAUSED BY DENTAL FILLING MATERIALS IN MR-IMAGING
    VIKHOFF, B
    RIBBELIN, S
    KOHLER, B
    EKHOLM, S
    BORRMAN, H
    ACTA RADIOLOGICA, 1995, 36 (03) : 323 - 325
  • [2] DENTAL MATERIAL ARTIFACTS ON MR IMAGES
    HINSHAW, DB
    HOLSHOUSER, BA
    ENGSTROM, HIM
    TJAN, AHL
    CHRISTIANSEN, EL
    CATELLI, WF
    RADIOLOGY, 1988, 166 (03) : 777 - 779
  • [3] Artifacts on temporomandibular joint MR images caused by mascara used as hair dye
    Wakae-Morita, Satsuki
    Igarashi, Chinami
    Kobayashi, Kaoru
    Sano, Tsukasa
    ORAL RADIOLOGY, 2017, 33 (03) : 231 - 236
  • [4] Artifacts on temporomandibular joint MR images caused by mascara used as hair dye
    Satsuki Wakae-Morita
    Chinami Igarashi
    Kaoru Kobayashi
    Tsukasa Sano
    Oral Radiology, 2017, 33 : 231 - 236
  • [5] MR IMAGING ARTIFACTS PRODUCED BY DENTAL MATERIALS
    FACHE, JS
    PRICE, C
    HAWBOLT, EB
    LI, DKB
    AMERICAN JOURNAL OF NEURORADIOLOGY, 1987, 8 (05) : 837 - 840
  • [6] Dental bur fragments causing metal artifacts on MR images
    Kaneda, T
    Minami, M
    Curtin, HD
    Utsunomiya, T
    Shirouzu, I
    Yamashiro, M
    Kiba, H
    Yamamoto, H
    Ohba, S
    AMERICAN JOURNAL OF NEURORADIOLOGY, 1998, 19 (02) : 317 - 319
  • [7] INVITRO PLAQUE-FORMATION ON COMMONLY USED DENTAL MATERIALS
    DUMMER, PMH
    HARRISON, KA
    JOURNAL OF ORAL REHABILITATION, 1982, 9 (05) : 413 - 417
  • [8] Evaluation of MRI artifacts caused by metallic dental implants and classification of the dental materials in use
    Starcuk, Z
    Bartusek, K
    Hubálková, H
    Bachorec, T
    Starcuková, J
    Krupa, P
    Measurement 2005, Proceedings, 2005, : 246 - 249
  • [9] ARTIFACTS IN MAGNETIC RESONANCE IMAGING CAUSED BY DENTAL MATERIALS: A SYSTEMATIC REVIEW
    Hegde, Shwetha
    JOURNAL OF EVIDENCE-BASED DENTAL PRACTICE, 2022, 22 (03)
  • [10] Artifacts In Magnetic Resonance Imaging and Computed Tomography Caused By Dental Materials
    Klinke, Thomas
    Daboul, Amro
    Maron, Juliane
    Gredes, Tomasz
    Puls, Ralf
    Jaghsi, Ahmad
    Biffar, Reiner