10-year results of the Birmingham Hip Resurfacing: a non-designer case series

被引:15
|
作者
Hunter, Thomas J. A. [1 ]
Moores, Thomas S. [1 ]
Morley, David [1 ]
Manoharan, Gopikanthan [1 ]
Collier, Simon G. [1 ]
Shaylor, Phillip J. [1 ]
机构
[1] Univ Hosp North Midlands NHS Trust, Stoke On Trent, Staffs, England
关键词
Arthroplasty; Birmingham Hip; Hip; Hip resurfacing; Metal-on-metal; FOLLOW-UP; ARTHROPLASTY; METAL; REVISION; REPLACEMENT; YOUNGER; COHORT;
D O I
10.5301/hipint.5000518
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Introduction: Recent controversies surrounding metal-on-metal (MoM) hip resurfacing has led to a substantial decline in its use. Despite this, there is good evidence to support the use of specific implants in select patients. Patients and methods: A retrospective analysis of Birmingham Hip Resurfacing (BHR) patients with a minimum of 10 years follow-up was performed. Functional scoring was performed with the Oxford Hip Score (OHS) and failure was defined as revision for any cause. 111 patients underwent 121 BHR procedures. All patients had a minimum follow-up of 10 years. 70 patients (63%) were male. Mean patient age at surgery was 52.5 years (male 53.9 years, female 48.8 years). Results: Overall survival at 10 years was 91% (97% male, 80% female). There was a statistically significant improvement in OHS postoperatively which remains at 10-year follow-up (p = <0.05). There was no significant difference in scores between the male and female groups. Revisions were most often in patients with smaller component sizes but this was not found to be statistically significant. Conclusions: Our results reflect that of the wider literature in that good outcomes can be obtained with this implant in a select group of patients and results are comparable to that of conventional hip arthroplasty in patients of a similar age.
引用
收藏
页码:50 / 52
页数:3
相关论文
共 50 条
  • [21] Birmingham Hip Resurfacing: A Single Surgeon Series Reported at a Minimum of 10 Years Follow-Up
    Mehra, Akshay
    Berryman, Fiona
    Matharu, Gulraj S.
    Pynsent, Paul B.
    Isbister, Eric S.
    [J]. JOURNAL OF ARTHROPLASTY, 2015, 30 (07): : 1160 - 1166
  • [22] Pseudotumours associated with metal-on-metal hip resurfacing: 10-year Newcastle experience
    Malviya, Ajay
    Holland, James P.
    [J]. ACTA ORTHOPAEDICA BELGICA, 2009, 75 (04): : 477 - 483
  • [23] The Birmingham Hip Resurfacing: 5-year clinical and radiographic results from a District General Hospital
    Ollivere, B.
    Duckett, S.
    August, A.
    Porteous, M.
    [J]. INTERNATIONAL ORTHOPAEDICS, 2010, 34 (05) : 631 - 634
  • [24] The Birmingham Hip Resurfacing: 5-year clinical and radiographic results from a District General Hospital
    B. Ollivere
    S. Duckett
    A. August
    M. Porteous
    [J]. International Orthopaedics, 2010, 34 : 631 - 634
  • [25] Short-term Results of Birmingham Hip Resurfacing in the United States
    Nam, Denis
    Nunley, Ryan M.
    Ruh, Erin L.
    Engh, C. Anderson, Jr.
    Rogerson, John S.
    Brooks, Peter J.
    Raterman, Stephen J.
    Su, Edwin P.
    Barrack, Robert L.
    [J]. ORTHOPEDICS, 2015, 38 (08) : E715 - E721
  • [26] Minimum 10-year results of a tapered cementless hip replacement
    Sakalkale, DP
    Eng, K
    Hozack, WJ
    Rothman, RH
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1999, (362) : 138 - 144
  • [27] 10 YEAR SURVIVORSHIP OF THE MEDIAL OXFORD UNICOMPARTMENTAL KNEE ARTHROPLASTY. A 1000 PATIENT NON-DESIGNER SERIES - THE EFFECT OF SURGICAL GRADE AND SUPERVISON.
    Jones, L.
    Bottomley, N.
    Pandit, H.
    Beard, D.
    Jackson, W.
    Price, A.
    [J]. OSTEOARTHRITIS AND CARTILAGE, 2012, 20 : S290 - S291
  • [28] Results of 5 to 10-year follow-up after hip resurfacing. A systematic analysis of the literature on long-term results
    Klotz, M. C. M.
    Breusch, S. J.
    Hassenpflug, M.
    Bitsch, R. G.
    [J]. ORTHOPADE, 2012, 41 (06): : 442 - +
  • [29] PELVIC SURGERY IN CASE OF CONGENITAL HIP DISLOCATION .E. 10-YEAR RESULTS OF COLONNA HIP ARTHROPLASTY
    MEYSTRE, JL
    [J]. ORTHOPADE, 1979, 8 (01): : 49 - 52
  • [30] 10-YEAR TO 14-YEAR RESULTS OF A CERAMIC HIP-PROSTHESIS
    WINTER, M
    GRISS, P
    SCHELLER, G
    MOSER, T
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1992, (282) : 73 - 80