Calibration and characterization of a digital camera for bidirectional reflectance distribution function retrieval of vicarious calibration sites

被引:5
|
作者
Czapla-Myers, Jeffrey S. [1 ]
Thome, Kurtis J. [1 ]
Biggar, Stuart F. [1 ]
机构
[1] Univ Arizona, Remote Sensing Grp, Coll Opt Sci, Tucson, AZ 85721 USA
来源
关键词
Digital camera; calibration; radiometry; bidirectional reflectance; ABSOLUTE RADIOMETRIC CALIBRATION; LANDSAT-7 ETM+; TERRA ASTER; BRDF DATA; MODIS; INSTRUMENT; SENSORS; SYSTEM; MISR;
D O I
10.1117/1.3116662
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
This work describes the calibration and characterization of a commercial digital camera that is being evaluated as a potential inexpensive imaging system to measure the bidirectional reflectance factor of the test site surfaces that are used by the Remote Sensing Group for the ground-based vicarious calibration of airborne and spaceborne systems. The main advantage to using a lightweight digital camera in combination with a fisheye lens is that a test site can be simultaneously spatially and spectrally sampled with greater ease. The motivation for using an inexpensive, lightweight digital camera to measure surface bidirectional distribution factor is the ability to spatially sample many points on the ground. An additional benefit to such cameras is their ability to operate without the need for cooling, which was required on a previous camera system developed at the Remote Sensing Group. Laboratory measurements include the gain and offset, spectral responsivity, linearity, cosine response, and point spread function. The results show that the uniformity of the Nikon detector array ranges from 0.7%-1.2%. The spectral bandwidth is approximately 90 nm for the blue channel, and 110 nm for the green, and 75 nm for the red channel. Linearity measurements show that the camera responsivity is nonlinear due to the internal image-processing algorithms, and a logarithmic fitting function is suggested. Measurements of the camera response versus the angle of incidence reveal that the camera does not exhibit a cos(4) falloff on detector irradiance as one would expect, which is another consequence of the internal algorithms.
引用
收藏
页数:15
相关论文
共 50 条
  • [1] Sensitivity analysis of a CCD-based camera system for the retrieval of bidirectional reflectance distribution function for vicarious calibration
    Nandy, P
    Thome, K
    Biggar, S
    [J]. IMAGING SPECTROMETRY VI, 2000, 4132 : 279 - 289
  • [2] Characterization and field use of a CCD camera system for retrieval of bidirectional reflectance distribution function
    Nandy, P
    Thome, K
    Biggar, S
    [J]. JOURNAL OF GEOPHYSICAL RESEARCH-ATMOSPHERES, 2001, 106 (D11) : 11957 - 11966
  • [3] The measured polarized bidirectional reflectance distribution function of a spectralon calibration target
    Betty, CL
    Fung, AK
    Irons, J
    [J]. IGARSS '96 - 1996 INTERNATIONAL GEOSCIENCE AND REMOTE SENSING SYMPOSIUM: REMOTE SENSING FOR A SUSTAINABLE FUTURE, VOLS I - IV, 1996, : 2183 - 2185
  • [4] BIDIRECTIONAL SPECTRAL REFLECTANCE FACTOR OF BAOTOU SANDY CALIBRATION SITE AND ITS APPLICATION IN VICARIOUS RADIOMETRIC CALIBRATION
    Zhao, Yongguang
    Ma, Lingling
    Liu, Yaokai
    Qian, Yonggang
    Li, Kun
    Wang, Ning
    Gao, Caixia
    Zhu, Xiaohua
    Li, Wan
    [J]. IGARSS 2020 - 2020 IEEE INTERNATIONAL GEOSCIENCE AND REMOTE SENSING SYMPOSIUM, 2020, : 6250 - 6253
  • [5] Sources of uncertainty in vicarious calibration: Understanding calibration target reflectance
    Anderson, K
    Milton, EJ
    Rollin, EM
    [J]. IGARSS 2003: IEEE INTERNATIONAL GEOSCIENCE AND REMOTE SENSING SYMPOSIUM, VOLS I - VII, PROCEEDINGS: LEARNING FROM EARTH'S SHAPES AND SIZES, 2003, : 2072 - 2074
  • [6] Temporal Variation in Surface Bidirectional Reflectance of the Railroad Valley Vicarious Calibration Test Site in Nevada
    Byford, Nicole
    Coburn, Craig A.
    [J]. CANADIAN JOURNAL OF REMOTE SENSING, 2022, 48 (06) : 722 - 736
  • [7] Instrument for retrieval of BRDF data for vicarious calibration
    Nandy, P
    Thome, K
    Biggar, S
    [J]. IGARSS '98 - 1998 INTERNATIONAL GEOSCIENCE AND REMOTE SENSING SYMPOSIUM, PROCEEDINGS VOLS 1-5: SENSING AND MANAGING THE ENVIRONMENT, 1998, : 562 - 564
  • [8] Vicarious calibration - A reflectance-based experiment with AirMISR
    Abdou, WA
    Conel, JE
    Pilorz, SH
    Helmlinger, MC
    Bruegge, CJ
    Gaitley, BJ
    Ledeboer, WC
    Martonchik, JV
    [J]. REMOTE SENSING OF ENVIRONMENT, 2001, 77 (03) : 338 - 353
  • [9] Ocean Sites Calibration and Water Color Retrieval for Directional Polarimetric Camera
    Pan, Yuwei
    Chen, Jingjing
    Sun, Liang
    Zhang, Pengfei
    Chen, Feinan
    Meng, Binghuan
    Xiang, Guangfeng
    Hong, Jin
    [J]. Guangxue Xuebao/Acta Optica Sinica, 2024, 44 (18):
  • [10] Vicarious calibration of GOCI for the SeaDAS ocean color retrieval
    Concha, Javier
    Mannino, Antonio
    Franz, Bryan
    Bailey, Sean
    Kim, Wonkook
    [J]. INTERNATIONAL JOURNAL OF REMOTE SENSING, 2019, 40 (10) : 3984 - 4001