A proposed method for correcting aperture penetration in high energy slit aperture and pinhole SPECT

被引:9
|
作者
Wrobel, MC [1 ]
Clinthorne, NH [1 ]
Fessler, JA [1 ]
Zhang, Y [1 ]
Wilderman, SJ [1 ]
Rogers, WL [1 ]
机构
[1] UNIV MICHIGAN, ANN ARBOR, MI 48109 USA
关键词
D O I
10.1109/23.632733
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Aperture penetration in high energy pin-hole and slit aperture SPECT results in a decrease of both resolution and contrast. At 511 keV, approximately 50% of projection counts can be attributed to penetration, Correction for this effect by measuring the penetration from a ''blocked'' slit has been investigated using SPRINT-II. Slit apertures were blocked with small bars of depleted uranium (DU) and projections acquired to serve as an estimate of the penetration background. The block's finite size and DU's limited attenuation of 511 keV photons required the blocked projection data to be corrected by a spatially variant, Monte Carlo based window function. F-18 line sources imaged without correction for penetration had an 8.8 mm FWHM and 18 mm FWTM due to large penetration tails. Lines reconstructed with penetration correction achieved 5.0 mm FWHM and 8.6 mm FWTM resolution. An F-18 filled mu-Jaszczak phantom (hot lines from 1.5 mm to 4 mm) imaged without correction produced a barely discernible 4 mm line set and uniform background. Images produced with correction produced clear images of 3 mm holes with negligible background The findings should be generally applicable to clinical gamma camera systems.
引用
收藏
页码:1564 / 1570
页数:7
相关论文
共 50 条
  • [1] High energy slit aperture and pinhole SPECT, a proposed method for correcting aperture penetration to enhance resolution and minimize penetration background.
    Wrobel, MC
    Clinthorne, NH
    Fessler, JA
    Zhang, Y
    Rogers, WL
    [J]. JOURNAL OF NUCLEAR MEDICINE, 1996, 37 (05): : 18 - 18
  • [2] An analytic model of pinhole aperture penetration for 3D pinhole SPECT image reconstruction
    Smith, MF
    Jaszczak, RJ
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (04): : 761 - 775
  • [3] Use of an analytic model of pinhole aperture penetration for resolution recovery in I-131 pinhole SPECT.
    Smith, MF
    Jaszczak, RJ
    Coleman, RE
    [J]. JOURNAL OF NUCLEAR MEDICINE, 1997, 38 (05): : 210 - 210
  • [4] High energy slit aperture SPECT and simplified in vitro methods for the dosimetry of positron emitting radiotracers
    Wrobel, MC
    [J]. MEDICAL PHYSICS, 1997, 24 (10) : 1648 - 1648
  • [5] A HIGH-RESOLUTION SLIT APERTURE FOR IMAGING SMALL ANIMALS WITH SPECT
    ROGERS, WL
    SLOSAR, J
    HUA, L
    CHIAO, P
    ZHANG, Y
    CLINTHORNE, NH
    [J]. JOURNAL OF NUCLEAR MEDICINE, 1993, 34 (05): : P9 - P9
  • [6] Finite aperture modeling in small animal pinhole SPECT Imaging
    Huang, PC
    Hsu, CH
    [J]. 2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 3454 - 3457
  • [7] Determination of the analytical point spread function of a pinhole aperture SPECT system
    Bekaert, V
    Brasse, D
    Gizard, B
    Guyonnet, JL
    [J]. 2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 2565 - 2569
  • [8] Geometrical calibration and aperture configuration design in multi-pinhole spect
    Vunckx, K.
    Defrise, M.
    Beque, D.
    Vanhove, C.
    Andreyev, A.
    Nuyts, J.
    [J]. 2008 IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING: FROM NANO TO MACRO, VOLS 1-4, 2008, : 1403 - +
  • [9] A geometric system model of finite aperture in small animal pinhole SPECT imaging
    Hsu, Ching-Han
    Huang, Po-Chia
    [J]. COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2006, 30 (03) : 181 - 185
  • [10] Calibration Method of Pinhole Aperture Attenuation in Reflectance Transfer Spectrometer
    Jiang Feng
    Niu Xinhua
    Yin Kai
    [J]. LASER & OPTOELECTRONICS PROGRESS, 2022, 59 (01)