Objectives: non-physiological flow behaviour plays a significant role in the development of distal anastomotic intimal hyperplasia. To investigate flow patterns in four anastomotic types of femoral end-to-side distal bypass graft anastomoses, a flow visualisation study was performed. Methods: transparent 1:1 casted replicas of distal vascular graft anastomoses created by conventional technique, Miller-cuff, Taylor- and Linton-patch were fabricated. A pulsatile mock circulation with a high-speed video system was constructed. Flow pattern was determined at mean Reynolds numbers 100-500. Migrations of the stagnation points on the bottom of the anastomoses at mean Reynolds numbers 100, 230, and 350 were measured. Results: a vortex forms during early systole and increases to maximum systole in all anastomoses. During the diastolic phase the vortex moves in the Miller-cuff distally to the toe of the anastomosis and remains standing, while in the other anastomotic types the vortex moves proximally to the heal of the junction and breaks down. The shift of the stagnation point in the Miller-cuff was considerably smaller than in the other anastomoses. Conclusion: conventional, Linton and Taylor anastomoses show similar flow patterns. The Miller-cuff with its wider cavity shows lower shift of the bottom stagnation point, but a persistent washout of the anastomotic cavity, which may contribute to its reported good clinical performance.
引用
收藏
页码:191 / 200
页数:10
相关论文
共 50 条
  • [21] THE ANASTOMOSIS ANGLE DOES CHANGE THE FLOW-FIELDS AT VASCULAR END-TO-SIDE ANASTOMOSES IN-VIVO
    STAALSEN, NH
    ULRICH, M
    WINTHER, J
    PEDERSEN, EM
    HOW, T
    NYGAARD, H
    JOURNAL OF VASCULAR SURGERY, 1995, 21 (03) : 460 - 471
  • [22] Numerical investigations of the unsteady blood flow in the end-to-side arteriovenous fistula for hemodialysis
    Jodko, Daniel
    Obidowski, Damian
    Reorowicz, Piotr
    Jozwik, Krzysztof
    ACTA OF BIOENGINEERING AND BIOMECHANICS, 2016, 18 (04) : 3 - 13
  • [23] BLOOD-FLOW IN END-TO-SIDE ARTERIOVENOUS FISTULAS AND SAPHENOUS VEIN GRAFTS
    ABBOTT, JA
    CAMPBELL, M
    BUSSELL, JA
    LIM, RC
    KIDNEY INTERNATIONAL, 1973, 3 (05) : 342 - 344
  • [24] SIMULATION OF NON-NEWTONIAN BLOOD-FLOW IN AN END-TO-SIDE ANASTOMOSIS
    BALLYK, PD
    STEINMAN, DA
    ETHIER, CR
    BIORHEOLOGY, 1994, 31 (05) : 565 - 586
  • [25] The influence of out-of-plane geometry on pulsatile flow within a distal end-to-side anastomosis
    Papaharilaou, Y
    Doorly, DJ
    Sherwin, SJ
    JOURNAL OF BIOMECHANICS, 2002, 35 (09) : 1225 - 1239
  • [26] Flow and wall shear stress in end-to-side and side-to-side anastomosis of venous coronary artery bypass grafts
    Frauenfelder, Thomas
    Boutsianis, Evangelos
    Schertler, Thomas
    Husmann, Lars
    Leschka, Sebastian
    Poulikakos, Dimos
    Marincek, Borut
    BIOMEDICAL ENGINEERING ONLINE, 2007, 6 (1)
  • [27] Flow and wall shear stress in end-to-side and side-to-side anastomosis of venous coronary artery bypass grafts
    Thomas Frauenfelder
    Evangelos Boutsianis
    Thomas Schertler
    Lars Husmann
    Sebastian Leschka
    Dimos Poulikakos
    Borut Marincek
    Hatem Alkadhi
    BioMedical Engineering OnLine, 6
  • [28] DOPPLER COLOR-FLOW IMAGES OF ILIOFEMORAL GRAFT END-TO-SIDE DISTAL ANASTOMOTIC MODELS
    RITTGERS, SE
    BHAMBHANI, GH
    ULTRASOUND IN MEDICINE AND BIOLOGY, 1993, 19 (04): : 257 - 267
  • [29] Numerical Study of Blood Flow at the End-to-Side Anastomosis of a Left Ventricular Assist Device for Adult Patients
    Yang, Ning
    Deutsch, Steven
    Paterson, Eric G.
    Manning, Keefe B.
    JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2009, 131 (11):
  • [30] EFFECTS OF RETROGRADE PORTAL VENOUS FLOW FOLLOWING SIDE-TO-SIDE PORTACAVAL ANASTOMOSIS - COMPARISON WITH END-TO-SIDE SHUNTS
    MURRAY, JF
    MULDER, DG
    JOURNAL OF CLINICAL INVESTIGATION, 1961, 40 (08): : 1413 - &