The Fluoropen: a simple low-cost device to detect intraoperative fluorescein fluorescence in stereotactic needle biopsy of brain tumors

被引:21
|
作者
Thien, Ady [1 ]
Rao, Jai Prashanth [1 ]
Ng, Wai Hoe [1 ]
King, Nicolas Kon Kam [1 ]
机构
[1] Natl Neurosci Inst, Dept Neurosurg, 11 Jalan Tan Tock Seng, Singapore 308433, Singapore
关键词
Brain tumor; Fluorescein sodium; Fluorescence; Stereotactic brain biopsy; HIGH-GRADE GLIOMAS; CONTRAST ENHANCEMENT; SURGERY; DIAGNOSIS; RESECTION;
D O I
10.1007/s00701-016-3041-8
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
The use of fluorescein fluorescence-guided stereotactic needle biopsy has been shown to improve diagnostic accuracy and to expedite operative procedure in the stereotactic needle biopsy of high-grade gliomas. We developed a device (Fluoropen) for detecting fluorescence in brain tumor tissues obtained by fluorescein fluorescence-guided stereotactic needle biopsy. The Fluoropen is a device consisting of a light source fitted with color filters to create the required emission and visualization wavelengths. The proof-of-concept study consisted of four consecutive patients who underwent fluorescein fluorescence-guided frameless stereotactic biopsy of brain tumor. Each sample was examined for the presence of fluorescence using the Fluoropen and compared with a microscope with fluorescence visualization capability. A total of six samples were obtained from four stereotactic needle biopsy procedures. Four out of five samples (80%) taken from the contrast-enhancing part of the tumors were shown to be fluorescent under the microscope fitted with fluorescence module and the Fluoropen. One non-contrast enhancing lesion was non-fluorescent using both the microscope fitted with fluorescence module and the Fluoropen. The Fluoropen was shown to have 100% concordance with the microscope fitted with fluorescence module. The Fluoropen is a low-cost and simple standalone device for the detection of fluorescein fluorescence that can expedite stereotactic needle biopsy by providing instant confirmation of the diagnostic sample and therefore avoid the need for an intraoperative frozen section. In patients with non-contrast enhancing tumors and those who were pre-treated with dexamethasone prior to surgery, fluorescein fluorescence-guided stereotactic needle biopsy will need to be used with caution.
引用
收藏
页码:371 / 375
页数:5
相关论文
共 11 条
  • [1] The Fluoropen: a simple low-cost device to detect intraoperative fluorescein fluorescence in stereotactic needle biopsy of brain tumors
    Ady Thien
    Jai Prashanth Rao
    Wai Hoe Ng
    Nicolas Kon Kam King
    [J]. Acta Neurochirurgica, 2017, 159 : 371 - 375
  • [2] Low-Cost Device for Fluorescein-Guided Surgery in Malignant Brain Tumor
    Lovato, Renan Maximilian
    Vitorino Araujo, Joao Luiz
    Esteves Veiga, Jose Carlos
    [J]. WORLD NEUROSURGERY, 2017, 104 : 61 - 67
  • [3] Fluorescein-assisted stereotactic needle biopsy of brain tumors: a single-center experience and systematic review
    Giuseppe Catapano
    Francesco Giovanni Sgulò
    Vincenzo Seneca
    Giuseppina Iorio
    Matteo de Notaris
    Giuseppe di Nuzzo
    [J]. Neurosurgical Review, 2019, 42 : 309 - 318
  • [4] Fluorescein-assisted stereotactic needle biopsy of brain tumors: a single-center experience and systematic review
    Catapano, Giuseppe
    Sgulo, Francesco Giovanni
    Seneca, Vincenzo
    Iorio, Giuseppina
    de Notaris, Matteo
    di Nuzzo, Giuseppe
    [J]. NEUROSURGICAL REVIEW, 2019, 42 (02) : 309 - 318
  • [5] FLUORESCEIN GUIDED SURGERY FOR MALIGNANT BRAIN TUMORS: A CASE SERIES WITH A LOW COST DEVICE
    Lovato, Renan Maximilian
    Vitorino Araujo, Joao Luiz
    Campos Paiva, Aline Lariessy
    Esteves Veiga, Jose Carlos
    [J]. NEURO-ONCOLOGY, 2017, 19 : 259 - 259
  • [6] Use of intraoperative fluorescein sodium fluorescence to improve the accuracy of tissue diagnosis during stereotactic needle biopsy of high-grade gliomas
    Rey-Dios, Roberto
    Hattab, Eyas M.
    Cohen-Gadol, Aaron A.
    [J]. ACTA NEUROCHIRURGICA, 2014, 156 (06) : 1071 - 1075
  • [7] Use of intraoperative fluorescein sodium fluorescence to improve the accuracy of tissue diagnosis during stereotactic needle biopsy of high-grade gliomas
    Roberto Rey-Dios
    Eyas M. Hattab
    Aaron A. Cohen-Gadol
    [J]. Acta Neurochirurgica, 2014, 156 : 1071 - 1075
  • [9] Accuracy of a low-cost localizing clip device after stereotactic 14-gauge breast core biopsy
    Curpen, NB
    Bouchard, L
    Lalonde, L
    Ouimet-Oliva, D
    [J]. RADIOLOGY, 1998, 209P : 198 - 198
  • [10] A handmade, low-cost intraoperative fluorescence navigation system with indocyanicne green for sentinel lymph node biopsy in skin cancer
    Fujisawa, Yasuhiro
    Nakamura, Yasuhiro
    Kawachi, Yasuhiro
    Otsuka, Fujio
    [J]. JOURNAL OF INVESTIGATIVE DERMATOLOGY, 2011, 131 : S36 - S36