Shaping ability of ProFile rotary nickel-titanium instruments with ISO sized tips in simulated root canals: Part 2

被引:1
|
作者
Bryant, ST [1 ]
Thompson, SA [1 ]
Al-Omari, MAO [1 ]
Dummer, PMH [1 ]
机构
[1] Univ Wales Coll Med, Sch Dent, Dept Restorat Dent, Cardiff CF4 4XY, S Glam, Wales
关键词
canal shaping; endodontics; nickel-titanium; root canals; rotary files;
D O I
暂无
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
The aim of this study was to determine the shaping ability of ProFile 0.04 taper rotary nickel-titanium instruments with ISO sized tips ill simulated canals, A total of 40 simulated root canals made up of four different shapes in terms of angle and position of curvature were prepared by ProFile instruments using the 'crown down' approach recommended by the manufacturer, Part 2 of this two-part report describes the efficacy of the instruments in terms of prevalence of canal aberrations, the amount and direction of canal transportation and the overall post-operative shape. Out of 37 completed specimens 9 zips (24%) and one ledge (3%) were created, but no perforations or danger zones were found. There were significant differences(P < 0.01) between canal shapes for the incidence of zips and elbows but not for their distance from the end-point of preparation. At specific positions along the canal length there were significant differences between the canal types in terms of their mean total width; overall, at the end-point of preparation and along the curved portion of the canals those specimens with 40 degrees curves were widest. This trend continued for the width of material removed from the outer aspect of the canal curves, whereas along the inner aspect of curves more material was removed in the 20 degrees canals, Overall, transportation was towards the outer aspect of the curve at the end-point of preparation and along the curved portion of canals but more balanced along the straight coronal section. Absolute transportation was small and below 0.1 mm at every position including the zips, Under the conditions of this study, Profile nickel-titanium rotary instruments with ISO sized tips produced a larger number of zips than expected; however, the degree of zipping was limited and relatively minor.
引用
收藏
页码:282 / 289
页数:8
相关论文
共 50 条
  • [31] Comparison of the shaping ability of GT® Series X, Twisted Files and AlphaKite rotary nickel-titanium systems in simulated canals
    Raidan Ba-Hattab
    Anne-Kathrin Pröhl
    Hermann Lang
    Dieter Pahncke
    BMC Oral Health, 13
  • [32] Evaluation of the root canal shaping ability of two rotary nickel-titanium systems
    Al-Manei, K. K.
    Al-Hadlaq, S. M. S.
    INTERNATIONAL ENDODONTIC JOURNAL, 2014, 47 (10) : 974 - 979
  • [33] Comparison of the shaping ability of GT® Series X, Twisted Files and AlphaKite rotary nickel-titanium systems in simulated canals
    Ba-Hattab, Raidan
    Proehl, Anne-Kathrin
    Lang, Hermann
    Pahncke, Dieter
    BMC ORAL HEALTH, 2013, 13
  • [34] The influence of various automated devices on the shaping ability of Mtwo rotary nickel-titanium instruments
    Buerklein, S.
    Schaefer, E.
    INTERNATIONAL ENDODONTIC JOURNAL, 2006, 39 (12) : 945 - 951
  • [35] Shaping ability of three profile rotary instrumentation techniques in simulated resin root canals
    Kum, KY
    Spängberg, L
    Cha, BY
    Il-Young, J
    Seung-Jong, L
    Chan-Young, L
    JOURNAL OF ENDODONTICS, 2000, 26 (12) : 719 - 723
  • [36] Comparative investigation of two rotary nickel-titanium instruments:: ProTaper versus RaCe.: Part 2.: Cleaning effectiveness and shaping ability in severely curved root canals of extracted teeth
    Schäfer, E
    Vlassis, M
    INTERNATIONAL ENDODONTIC JOURNAL, 2004, 37 (04) : 239 - 248
  • [37] Efficiency of rotary nickel-titanium K3 instruments compared with stainless steel hand K-Flexofile.: Part 1.: Shaping ability in simulated curved canals
    Schäfer, E
    Florek, H
    INTERNATIONAL ENDODONTIC JOURNAL, 2003, 36 (03) : 199 - 207
  • [38] Cyclic fatigue behavior of nickel-titanium dental rotary files in clinical simulated root canals
    Chi, Chih-Wen
    Li, Chun-Chieh
    Lin, Chun-Pin
    Shin, Chow-Shing
    JOURNAL OF THE FORMOSAN MEDICAL ASSOCIATION, 2017, 116 (04) : 306 - 312
  • [39] Shaping ability of five different nickel-titanium systems in simulated S-shaped canals
    Koring, Stella
    Schwahn, Christian
    Kocher, Thomas
    Steffen, Heike
    ENDO-ENDODONTIC PRACTICE TODAY, 2020, 14 (02): : 135 - 143
  • [40] The study of the shaping ability of three nickel-titanium rotary instruments to two apical dimensions.
    Wei, X
    Ling, JQ
    Cheung, GSP
    JOURNAL OF DENTAL RESEARCH, 2003, 82 : 334 - 334