Validation of the Iowa Head-Mounted Open-Source Perimeter

被引:2
|
作者
Heinzman, Zachary [1 ]
Linton, Edward [1 ]
Marin-Franch, Ivan [2 ,3 ]
Turpin, Andrew [4 ]
Alawa, Karam [5 ]
Wijayagunaratne, Anushi [1 ]
Wall, Michael [1 ]
机构
[1] Univ Iowa, Carver Coll Med, Dept Ophthalmol & Visual Sci, Iowa City, IA 52242 USA
[2] Computat Optometry, Atarfe, Spain
[3] Southwest Eye Inst, Tavistock, England
[4] Curtin Univ, Curtin Sch Populat Hlth, Bentley, WA, Australia
[5] Harvard Med Sch, Dept Ophthalmol, Massachusetts Eye & Ear, Boston, MA 02115 USA
来源
关键词
virtual reality; glaucoma; perimetry; visual field; STANDARD AUTOMATED PERIMETRY; VISUAL-FIELD PROGRESSION; EFFECTIVE DYNAMIC RANGES; STIMULUS SIZES III; MATRIX; MOTION;
D O I
10.1167/tvst.12.9.19
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To assess the validity of visual field (VF) results from the Iowa Head-Mounted Display (HMD) Open-Source Perimeter and to test the hypothesis that VF defects and test-retest repeatability are similar between the HMD and Octopus 900 perimeters. Methods: We tested 20 healthy and nine glaucoma patients on the HMD and Octopus 900 perimeters using the Open Perimetry Interface platform with size V stimuli, a custom grid spanning the central 26 degrees of the VF, and a ZEST thresholding algorithm. Historical data from the Humphrey Field Analyzer (HFA) were also analyzed. Repeatability was analyzed with the repeatability coefficient (RC), and VF defect detection was determined through side-by-side comparisons. Results: The pointwise RCs were 2.6 dB and 3.4 dB for the HMD and Octopus 900 perimeters in ocular healthy subjects, respectively. Likewise, the RCs were 4.2 dB and 3.5 dB, respectively, in glaucomatous patients. Limits of agreement between the HMD and Octopus 900 perimeters were +/- 4.6 dB (mean difference, 0.4 dB) for healthy patients and +/- 8.9 dB (mean difference, 0.1 dB) for glaucomatous patients. Retrospective analysis showed that pointwise RCs on the HFA2 perimeter were between 3.4 and 3.7 dB for healthy patients and between 3.9 and 4.7 dB for glaucoma patients. VF defects were similar between the HMD and Octopus 900 for glaucoma subjects. Conclusions: The Iowa Virtual Reality HMD Open-Source Perimeter is as repeatable as the Octopus 900 perimeter and is a more portable and less expensive alternative than traditional perimeters. Translational Relevance: This study demonstrates the validity of the visual field results from the Iowa HMD Open-Source Perimeter which may help expand perimetry access.
引用
收藏
页数:10
相关论文
共 50 条
  • [41] Distortion matching with head-mounted displays
    Vizgaitis, JN
    [J]. HELMET- AND HEAD-MOUNTED DISPLAYS VII, 2002, 4711 : 162 - 173
  • [42] Holographic head-mounted display with RGB light emitting diode light source
    Moon, Eunkyong
    Kim, Myeongjae
    Roh, Jinyoung
    Kim, Hwi
    Hahn, Joonku
    [J]. OPTICS EXPRESS, 2014, 22 (06): : 6526 - 6534
  • [43] Contrast budget of head-mounted displays
    de Wit, GC
    [J]. OPTICAL ENGINEERING, 2002, 41 (10) : 2419 - 2426
  • [44] Reading with a head-mounted video magnifier
    Ortiz, A
    Chung, STL
    Legge, GE
    Jobling, JT
    [J]. OPTOMETRY AND VISION SCIENCE, 1999, 76 (11) : 755 - 763
  • [45] Head-mounted display system for microneurosurgery
    Chen, JCT
    Moffitt, K
    Levy, ML
    [J]. STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 1997, 68 (1-4) : 25 - 32
  • [46] Optical schemes of the head-mounted displays
    Romanova, G. E.
    Bakholdin, A. V.
    Vasilyev, V. N.
    [J]. OPTICAL MODELING AND PERFORMANCE PREDICTIONS IX, 2017, 10374
  • [47] Mixed reality head-mounted display
    Choy, CST
    Lu, KC
    Chow, WM
    Sit, KM
    Lay, KH
    Ho, W
    [J]. ICCE: 2005 INTERNATIONAL CONFERENCE ON CONSUMER ELECTRONICS, DIGEST OF TECHNICAL PAPERS, 2005, : 427 - 428
  • [48] Storage and Memory in Head-Mounted Displays
    Coughlin, Tom
    [J]. IEEE CONSUMER ELECTRONICS MAGAZINE, 2021, 10 (05) : 51 - 53
  • [49] International Standardization of Head-mounted Displays
    Hyodo, Kei
    [J]. Kyokai Joho Imeji Zasshi/Journal of the Institute of Image Information and Television Engineers, 2019, 73 (06): : 1090 - 1093
  • [50] A MINIATURE HEAD-MOUNTED SURGICAL MICROSCOPE
    BECKER, SC
    [J]. ARCHIVES OF OPHTHALMOLOGY, 1969, 82 (02) : 216 - &