Usefulness of Ag Additional Filter on Image Quality and Radiation Dose for Low-Dose Chest Computed Tomography

被引:2
|
作者
Kawamoto, Keishin [1 ,2 ]
Sato, Hideyuki [1 ]
Kogure, Yosuke [1 ]
机构
[1] Juntendo Univ Hosp, Dept Radiol Technol, Tokyo, Japan
[2] Juntendo Univ Hosp, Dept Radiol Technol, 3-1-3 Hongo,Bunkyo-ku, Tokyo 1138421, Japan
关键词
Ag additional filter; low-dose chest CT; radiation dose; image quality; SPECTRAL FILTRATION; 100; KVP; CT; REDUCTION; ARTIFACTS;
D O I
10.1097/RCT.0000000000001538
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
ObjectiveThis study aimed to evaluate the effect of a silver (Ag) additional filter on dose characteristics and image quality in low-dose chest computed tomography (CT).MethodsA dose evaluation phantom, physical evaluation phantom, and chest phantom were scanned with and without an Ag additional filter. The doses were adjusted so that the displayed the volume CT dose indexes (CTDIvol) were from 0.3 to 1.6 mGy. For dose characteristics, the spectrum of photon energies and the measured CTDIvol were calculated for each scanning condition. For task-based image quality analysis, task transfer function, noise power spectrum, and system performance were evaluated. Streak artifacts, image noise, and contrast-to-noise ratio were quantified using a chest phantom.ResultsWith the Ag additional filter, mean energy was 22% higher and the CTDIvol was approximately 30% lower than those without the Ag additional filter. The task transfer function and noise power spectrum with the Ag additional filter were lower than those without the Ag additional filter. The system performance with the Ag additional filter was similar to that without the Ag additional filter. The Ag additional filter reduced streak artifact near the lung apex and image noise in the lung fields. The contrast-to-noise ratio was slightly higher with the Ag additional filter than that without the Ag additional filter.ConclusionsThe output dose and spatial resolution with the Ag additional filter were lower than those without the Ag additional filter. However, this filter helped reduce the radiation dose, image noise, and streak artifacts, particularly when scanning at ultralow doses.
引用
收藏
页码:236 / 243
页数:8
相关论文
共 50 条
  • [1] ULTRA-LOW-DOSE COMPUTED TOMOGRAPHY IN UROLITHIASIS-EFFECT OF AN ADDITIONAL TIN FILTER ON IMAGE QUALITY AND RADIATION DOSE
    Haubold, Johannes
    Zensen, Sebastian
    Erfanian, Youssef
    Guberina, Nika
    Opitz, Marcel
    Sawicki, Lino Morris
    Forsting, Michael
    Umutlu, Lale
    Theysohn, Jens M.
    [J]. RADIATION PROTECTION DOSIMETRY, 2021, 197 (3-4) : 146 - 153
  • [2] A multi-institutional assessment of low-dose protocols in chest computed tomography: Dose and image quality
    Tonkopi, Elena
    Tetteh, Mercy Afadzi
    Gunn, Catherine
    Ashraf, Haseem
    Rusten, Sigrid Lia
    Safi, Perkhah
    Tinsoe, Nora Suu
    Colford, Kylie
    Ouellet, Olivia
    Naimi, Salma
    Johansen, Safora
    [J]. ACTA RADIOLOGICA OPEN, 2024, 13 (01)
  • [3] Dosimetry for low-dose computed tomography of the chest
    Lenzen, H
    Roos, N
    Diederich, S
    Meier, N
    [J]. RADIOLOGE, 1996, 36 (06): : 483 - 488
  • [4] Low-dose abdominopelvic computed tomography in patients with lymphoma: An image quality and radiation dose reduction study
    Yoon, Sungjin
    Yoo, Kwai Han
    Park, So Hyun
    Kim, Hawk
    Lee, Jae Hoon
    Park, Jinny
    Park, Seong Ho
    Kim, Hwa Jung
    [J]. PLOS ONE, 2022, 17 (08):
  • [5] Low-dose computed tomography perceptual image quality assessment
    Lee, Wonkyeong
    Wagner, Fabian
    Galdran, Adrian
    Shi, Yongyi
    Xia, Wenjun
    Wang, Ge
    Mou, Xuanqin
    Ahamed, Md. Atik
    Imran, Abdullah Al Zubaer
    Oh, Ji Eun
    Kim, Kyungsang
    Baek, Jong Tak
    Lee, Dongheon
    Hong, Boohwi
    Tempelman, Philip
    Lyu, Donghang
    Kuiper, Adrian
    van Blokland, Lars
    Calisto, Maria Baldeon
    Hsieh, Scott
    Han, Minah
    Baek, Jongduk
    Maier, Andreas
    Wang, Adam
    Gold, Garry Evan
    Choi, Jang-Hwan
    [J]. Medical Image Analysis, 2025, 99
  • [6] Radiation Dose Reduction for Computed Tomography Localizer Radiography Using an Ag Additional Filter
    Nomura, Keiichi
    Fujii, Keisuke
    Goto, Takahiro
    Tsukagoshi, Shinsuke
    Ota, Hiroyuki
    Iwabuchi, Yuto
    Suzuki, Hidenobu
    Muramatsu, Yoshihisa
    Kobayashi, Tatsushi
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2021, 45 (01) : 84 - 92
  • [7] Radiation dose and image quality in computed tomography
    Prokop, M
    [J]. ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2002, 174 (05): : 631 - 636
  • [8] Low-dose, volumetric helical CT - Image quality, radiation dose, and usefulness for evaluation of bronchiectasis
    Jung, KJ
    Lee, KS
    Kim, SY
    Kim, TS
    Pyeun, YS
    Lee, JY
    [J]. INVESTIGATIVE RADIOLOGY, 2000, 35 (09) : 557 - 563
  • [9] Application of Aortic Computed Tomography Angiography with Low Radiation Dose: Assessment of Image Quality and Radiation Dose
    Liu, Jianhua
    Li, Long
    Zhang, Yan
    Yuan, Qinghai
    Sun, Yong
    [J]. 2012 IEEE FIFTH INTERNATIONAL CONFERENCE ON ADVANCED COMPUTATIONAL INTELLIGENCE (ICACI), 2012, : 148 - 151
  • [10] Low-dose dental computed tomography:: Significant dose reduction without loss of image quality
    Rustemeyer, P
    Streubühr, U
    Suttmoeller, J
    [J]. ACTA RADIOLOGICA, 2004, 45 (08) : 847 - 853