Accuracy comparison of cup positioning during total hip arthroplasty using Hip Align and AR-Hip in the supine position

被引:2
|
作者
Fujita, Masahiro [1 ]
Hayashi, Shinya [1 ]
Kuroda, Yuichi [1 ]
Nakano, Naoki [1 ]
Maeda, Toshihisa [1 ]
Matsushita, Takehiko [1 ]
Matsumoto, Tomoyuki [1 ]
Kuroda, Ryosuke [1 ]
机构
[1] Kobe Univ, Grad Sch Med, Dept Orthoped Surg, 7-5-1 Kusunoki Cho, Chuo Ku, Kobe, Hyogo 6500017, Japan
关键词
Total hip arthroplasty; Cup positioning; Portable navigation system; Accelerometer-based navigation; Augmented reality-based navigation; ACETABULAR COMPONENT; SPINAL DEFORMITY; ANTEVERSION; NAVIGATION; FRACTURES; PATIENT;
D O I
10.1007/s00402-023-04975-y
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
IntroductionThe aim of this study is to compare the accuracy of acetabular cup positioning between the accelerometer-based navigation system and the augmented reality-based navigation system during THA in the supine position.Materials and methodsThis retrospective study included 66 patients (70 hips) who underwent THA using two types of portable navigation system, Hip Align or AR-Hip, in the spine position. The absolute difference between the intraoperative navigation record and postoperative measurement using computed tomography data was evaluated. Preoperative clinical factors that decreased the accuracy of cup positioning by & GE; 3 & DEG; were analyzed via multiple logistic regression analyses.ResultsThe average absolute error of inclination was 2.8 & PLUSMN; 2.6 & DEG; in Hip Align and 2.7 & PLUSMN; 1.8 & DEG; in AR-Hip, and absolute anteversion error was 2.5 & PLUSMN; 2.0 & DEG; in Hip Align and 2.6 & PLUSMN; 2.2 & DEG; in AR-Hip, and there was no significantly different between the two navigation systems. There was a significant association between the absolute measurement error (& GE; 3 & DEG;) of cup inclination and patients' BMI in the Hip Align group [odds ratio (OR) 1.350; 95% confidence interval (CI) 1.035-1.760; p = 0.027], but not in the AR-Hip group.ConclusionsThe accuracy of the acetabular cup positioning between the Hip Align and AR-Hip showed no difference during THA in the spine position. The high BMI could have negative influence on the accuracy of cup positioning in THA using Hip Align, thus AR-Hip could be designable for obesity patients.
引用
收藏
页码:7229 / 7235
页数:7
相关论文
共 50 条
  • [21] Primary Total Hip Arthroplasty in Severe Dysplastic Hip Osteoarthritis With a Far Proximal Cup Position
    Berninger, Markus T.
    Hungerer, Sven
    Friederichs, Jan
    Stuby, Fabian M.
    Fulghum, Christian
    Schipp, Rolf
    [J]. JOURNAL OF ARTHROPLASTY, 2019, 34 (05): : 920 - 925
  • [22] Accuracy of cup orientation and learning curve of the accelerometer-based portable navigation system for total hip arthroplasty in the supine position
    Kamenaga, Tomoyuki
    Hayashi, Shinya
    Hashimoto, Shingo
    Matsumoto, Tomoyuki
    Takayama, Koji
    Fujishiro, Takaaki
    Hiranaka, Takafumi
    Niikura, Takahiro
    Kuroda, Ryosuke
    [J]. JOURNAL OF ORTHOPAEDIC SURGERY, 2019, 27 (02)
  • [23] Accuracy of a novel accelerometer-based navigation (Naviswiss) for total hip arthroplasty in the supine position
    Hasegawa, Masahiro
    Naito, Yohei
    Tone, Shine
    Sudo, Akihiro
    [J]. BMC MUSCULOSKELETAL DISORDERS, 2022, 23 (01)
  • [24] Accuracy of a novel accelerometer-based navigation (Naviswiss) for total hip arthroplasty in the supine position
    Masahiro Hasegawa
    Yohei Naito
    Shine Tone
    Akihiro Sudo
    [J]. BMC Musculoskeletal Disorders, 23
  • [25] The accuracy of digital templating: a comparison of short-stem total hip arthroplasty and conventional total hip arthroplasty
    Schmidutz, Florian
    Steinbrueck, Arnd
    Wanke-Jellinek, Lorenz
    Pietschmann, Matthias
    Jansson, Volkmar
    Fottner, Andreas
    [J]. INTERNATIONAL ORTHOPAEDICS, 2012, 36 (09) : 1767 - 1772
  • [26] Accuracy of angle and position of the cup using computed tomography-based navigation systems in total hip arthroplasty
    Iwana, Daiki
    Nakamura, Nobuo
    Miki, Hidenobu
    Kitada, Makoto
    Hananouchi, Takehito
    Sugano, Nobuhiko
    [J]. COMPUTER AIDED SURGERY, 2013, 18 (5-6) : 187 - 194
  • [27] The accuracy of digital templating: a comparison of short-stem total hip arthroplasty and conventional total hip arthroplasty
    Florian Schmidutz
    Arnd Steinbrück
    Lorenz Wanke-Jellinek
    Matthias Pietschmann
    Volkmar Jansson
    Andreas Fottner
    [J]. International Orthopaedics, 2012, 36 : 1767 - 1772
  • [28] Effect of acetabular cup position and orientation in cemented total hip arthroplasty
    Hirakawa, K
    Mitsugi, N
    Koshino, T
    Saito, T
    Hirasawa, Y
    Kubo, T
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2001, (388) : 135 - 142
  • [29] The transverse acetabular ligament may be used to align the acetabular cup in total hip arthroplasty
    Pearce, C. J.
    Sexton, S. A.
    Davies, D. C.
    Khaleel, A.
    [J]. HIP INTERNATIONAL, 2008, 18 (01) : 7 - 10
  • [30] Accuracy of computer-assisted cup placement in total hip arthroplasty
    Jolles, BM
    Genoud, P
    Hoffmeyer, P