Superparamagnetic iron oxide nanoparticle enhanced percutaneous microwave ablation: Ex-vivo characterization using magnetic resonance thermometry

被引:0
|
作者
Bhagavatula, Sharath K. [1 ,4 ,5 ]
Panikkanvalappil, Sajanlal R. [2 ,3 ]
Tokuda, Junichi [1 ]
Levesque, Vincent [1 ]
Tatarova, Zuzana [1 ]
Liu, Guigen [1 ]
Markert, John E. [1 ]
Jonas, Oliver [1 ,6 ,7 ]
机构
[1] Harvard Med Sch, Brigham & Womens Hosp, Dept Radiol, Boston, MA USA
[2] Dana Farber Canc Inst, Dept Imaging, Boston, MA USA
[3] Harvard Med Sch, Boston, MA USA
[4] Harvard Med Sch, Dept Radiol, 75 Francis St, Boston, MA 02115 USA
[5] Brigham & Womens Hosp, 75 Francis St, Boston, MA 02115 USA
[6] Harvard Med Sch, Dept Radiol, 221 Longwood Ave, Boston, MA 02115 USA
[7] Brigham & Womens Hosp, 221 Longwood Ave, Boston, MA 02115 USA
基金
美国国家卫生研究院;
关键词
interventional radiology; magnetic resonance thermometry; microwave ablation; minimally invasive treatment; superparamagnetic iron oxide nanoparticles; LIVER; FERUMOXYTOL; DEVICES; TUMORS;
D O I
10.1002/mp.17040
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Percutaneous microwave ablation (pMWA) is a minimally invasive procedure that uses a microwave antenna placed at the tip of a needle to induce lethal tissue heating. It can treat cancer and other diseases with lower morbidity than conventional surgery, but one major limitation is the lack of control over the heating region around the ablation needle. Superparamagnetic iron oxide nanoparticles have the potential to enhance and control pMWA heating due to their ability to absorb microwave energy and their ease of local delivery. Purpose: The purpose of this study is to experimentally quantify the capabilities of FDA-approved superparamagnetic iron oxide Feraheme nanoparticles (FHNPs) to enhance and control pMWA heating. This study aims to determine the effectiveness of locally injected FHNPs in increasing the maximum temperature during pMWA and to investigate the ability of FHNPs to create a controlled ablation zone around the pMWA needle. Methods: PMWA was performed using a clinical ablation system at 915 MHz in ex-vivo porcine liver tissues. Prior to ablation, 50 uL 5 mg/mL FHNP injections were made on one side of the pMWA needle via a 23-gauge needle. Local temperatures at the FHNP injection site were directly compared to equidistant control sites without FHNP. First, temperatures were compared using directly inserted thermocouples. Next, temperatures were measured non-invasively using magnetic resonance thermometry (MRT), which enabled comprehensive four-dimensional (volumetric and temporal) assessment of heating effects relative to nanoparticle distribution, which was quantified using dual-echo ultrashort echo time (UTE) subtraction MR imaging. Maximum heating within FHNP-exposed tissues versus control tissues were compared at multiple pMWA energy delivery settings. The ability to generate a controlled asymmetric ablation zone using multiple FHNP injections was also tested. Finally, intra-procedural MRT-derived heat maps were correlated with gold standard gross pathology using Dice similarity analysis. Results: Maximum temperatures at the FHNP injection site were significantly higher than control (without FHNP) sites when measured using direct thermocouples (93.1 +/- 6.0 degrees C vs. 57.2 +/- 8.1 degrees C, p = 0.002) and using non-invasive MRT (115.6 +/- 13.4 degrees C vs. 49.0 +/- 10.6 degrees C, p = 0.02). Temperature difference between FHNP-exposed and control sites correlated with total energy deposition: 66.6 +/- 17.6 degrees C, 58.1 +/- 8.5 degrees C, and 20.8 +/- 9.2 degrees C at high (17.5 +/- 2.2 kJ), medium (13.6 +/- 1.8 kJ), and low (8.8 +/- 1.1 kJ) energies, respectively (all pairwise p < 0.05). Each FHNP injection resulted in a nanoparticle distribution within 0.9 +/- 0.2 cm radially of the injection site and a local lethal heating zone confined to within 1.1 +/- 0.4 cm radially of the injection epicenter. Multiple injections enabled a controllable, asymmetric ablation zone to be generated around the ablation needle, with maximal ablation radius on the FHNP injection side of 1.6 +/- 0.2 cm compared to 0.7 +/- 0.2 cm on the non-FHNP side (p = 0.02). MRT intra-procedural predicted ablation zone correlated strongly with post procedure gold-standard gross pathology assessment (Dice similarity 0.9). Conclusions: Locally injected FHNPs significantly enhanced pMWA heating in liver tissues, and were able to control the ablation zone shape around a pMWA needle. MRI and MRT allowed volumetric real-time visualization of both FHNP distribution and FHNP-enhanced pMWA heating that was useful for intra-procedural monitoring. This work strongly supports further development of a FHNP-enhanced pMWA paradigm; as all individual components of this approach are approved for patient use, there is low barrier for clinical translation.
引用
收藏
页码:3195 / 3206
页数:12
相关论文
共 50 条
  • [1] Magnetic resonance-based thermometry during laser ablation on ex-vivo swine pancreas and liver
    Allegretti, G.
    Saccomandi, P.
    Giurazza, F.
    Caponero, M. A.
    Frauenfelder, G.
    Di Matteo, F. M.
    Zobel, B. Beomonte
    Silvestri, S.
    Schena, E.
    MEDICAL ENGINEERING & PHYSICS, 2015, 37 (07) : 631 - 641
  • [2] Detection of Macrophage Localization in the Abdominal Aortic Aneurysm Wall Using Ex Vivo Superparamagnetic Iron Oxide-Enhanced Magnetic Resonance Imaging
    Umetsu, Michihisa
    Goto, Hitoshi
    Nakamura, Yasuhiro
    Ota, Hideki
    Shimizu, Takuya
    Hashimoto, Munetaka
    Akamatsu, Daijirou
    Kamei, Takashi
    ANNALS OF VASCULAR SURGERY, 2020, 68 : 344 - 350
  • [3] Dual energy computed tomography thermometry during hepatic microwave ablation in an ex-vivo porcine model
    Paul, Jijo
    Vogl, Thomas J.
    Chacko, Annamma
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2015, 31 (07): : 683 - 691
  • [4] Ultrasmall Superparamagnetic Iron Oxide Nanoparticles as Nanocarriers for Magnetic Resonance Imaging: Development and In Vivo Characterization
    Si, Ge
    Hapuarachchige, Sudath
    Artemov, Dmitri
    ACS APPLIED NANO MATERIALS, 2022, 5 (07) : 9625 - 9632
  • [5] Temperature map of kidneys undergoing microwave ablation using computed tomography-thermometry: ex-vivo experiments and numerical simulations
    Saccomandi, Paola
    De Landro, Martina
    Massaroni, Carlo
    Fong, Yuman
    Park, Jinha
    Park, John
    Schena, Emiliano
    2019 IEEE INTERNATIONAL SYMPOSIUM ON MEDICAL MEASUREMENTS AND APPLICATIONS (MEMEA), 2019,
  • [6] The assessment of ex-vivo human pancreata using magnetic resonance imaging
    Khambalia, Hussein
    Zhao, Sha
    van Dellen, David
    Summers, Angela
    Moinuddin, Zia
    Giuffrida, Giuseppe
    Yiannoullou, Petros
    Parker, Geoffrey
    Augustine, Titus
    XENOTRANSPLANTATION, 2015, 22 : S2 - S2
  • [7] THE ASSESSMENT OF EX-VIVO HUMAN PANCREATA USING MAGNETIC RESONANCE IMAGING
    Khambalia, Hussein
    Zhao, Sha
    van Dellen, David
    Summers, Angela
    Moinuddin, Zia
    Giuffrida, Giuseppe
    Yiannoullou, Petros
    Parker, Geoffrey
    Augustine, Titus
    TRANSPLANTATION, 2015, 99 (11) : S2 - S2
  • [8] Improving nuclear magnetic resonance and electron spin resonance thermometry with size reduction of superparamagnetic iron oxide nanoparticles
    Lin, Pei-Yun
    Chalise, Darshan
    Cahill, David G.
    PHYSICAL REVIEW APPLIED, 2024, 22 (04):
  • [9] Structural and magnetic characterization of superparamagnetic iron platinum nanoparticle contrast agents for magnetic resonance imaging
    Taylor, Robert M.
    Huber, Dale L.
    Monson, Todd C.
    Esch, Victor
    Sillerud, Laurel O.
    JOURNAL OF VACUUM SCIENCE & TECHNOLOGY B, 2012, 30 (02):
  • [10] Magnetic resonance angiography of the portal venous system enhanced by superparamagnetic iron oxide
    Schmitz, SA
    Jensen, K
    Wolf, KJ
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2000, 172 (01): : 51 - 54