The impact of software updates on accuracy of intraoral scanners

被引:20
|
作者
Schmalzl, Judit [1 ]
Roth, Ivett [1 ]
Borbely, Judit [1 ]
Hermann, Peter [1 ]
Vecsei, Balint [1 ]
机构
[1] Semmelweis Univ, Dept Prosthodont, Szentkiralyi St 47, H-1088 Budapest, Hungary
关键词
Accuracy; Trueness; Precision; Software update; Intraoral scanner; DIGITAL IMPRESSIONS; CONVENTIONAL IMPRESSIONS; FIXED PROSTHODONTICS; PRECISION; TRUENESS; SYSTEMS;
D O I
10.1186/s12903-023-02926-y
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
BackgroundDigital workflow is showing an increasing tendency in everyday dentistry. Accuracy is essential during digital dental workflows for all indication areas. The present study aimed to evaluate the effect of software updates on the accuracy of intraoral scanner (IOS) devices.Methods3Shape Trios 3 Pod with software versions 18.1.2. (TRI3_1) and 20.1.2. (TRI3_2); 3Shape Trios 4 Move, version 19.2.2. (TRI4_1); and 3Shape Trios 4 Pod, version 20.1.1. (TRI4_2) were used to take direct optical impressions from a polymethyl methacrylate (PMMA) full arch reference model with prepared teeth (FDI 11,14,17 for crowns and FDI 26 for onlay) and an edentulous region (between FDI 14 and 17). The scanners were used eight times; STL files were imported into Geomagic Control X for accuracy assessment by comparing them to a reference data set created by an industrial high-precision scanner (AICON SmartScan-3D C5). The average deviation of the surface points was calculated in three locations: across a full arch (Parameter 1), the region of a four-unit bridge (Parameter 2), and a single prepared abutment (Parameter 3).ResultsIn parameter 1 and 2, the newest model with the latest software (TRI4_2) reached the highest accuracy (31.06 +/- 5.24 mu m and 21.69 +/- 7.50 mu m). In parameter 3, an older generation scanner running legacy software produced the highest accuracy: TRI4_1, 11.75 +/- 0.35 mu m.ConclusionAppropriate software updates can significantly increase the trueness and precision of intraoral scanner devices. With updated software, the older generation can match the accuracy level of latest equipment.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] The impact of software updates on accuracy of intraoral scanners
    Judit Schmalzl
    Ivett Róth
    Judit Borbély
    Péter Hermann
    Bálint Vecsei
    [J]. BMC Oral Health, 23
  • [2] The effect of software updates on the trueness and precision of intraoral scanners
    Vag, Janos
    Renne, Walter
    Revell, Griffin
    Ludlow, Mark
    Mennito, Anthony
    Teich, Sorin T.
    Gutmacher, Zvi
    [J]. QUINTESSENCE INTERNATIONAL, 2021, 52 (07): : 636 - 644
  • [3] Accuracy of commercial intraoral scanners
    Sacher, Mattia
    Schulz, Georg
    Deyhle, Hans
    Jaeger, Kurt
    Mueller, Bert
    [J]. JOURNAL OF MEDICAL IMAGING, 2021, 8 (03)
  • [4] A comparison of the accuracy of intraoral scanners using an intraoral environment simulator
    Park, Hye-Nan
    Lim, Young-Jun
    Yi, Won-Jin
    Han, Jung-Suk
    Lee, Seung-Pyo
    [J]. JOURNAL OF ADVANCED PROSTHODONTICS, 2018, 10 (01): : 58 - 64
  • [5] Comparison of Accuracy of Current Ten Intraoral Scanners
    Amornvit, Pokpong
    Rokaya, Dinesh
    Sanohkan, Sasiwimol
    [J]. BIOMED RESEARCH INTERNATIONAL, 2021, 2021
  • [6] Accuracy and Reliability of Intraoral Scanners: Are They the Better Option?
    Ebeid K.
    Salah T.
    Nossair S.
    [J]. Current Oral Health Reports, 2017, 4 (3) : 209 - 214
  • [7] Exploring the impact of the extent of the partially edentulous area on the accuracy of two intraoral scanners
    Alfaraj, Amal
    Khanlar, Leila Nasiry
    Lin, Wei-Shao
    Zandinejad, Amirali
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 2024, 131 (01): : 163.e1 - 163.e8
  • [8] Complete-arch accuracy of intraoral scanners
    Treesh, Joshua C.
    Liacouras, Peter C.
    Taft, Robert M.
    Brooks, Daniel, I
    Raiciulescu, Sorana
    Ellert, Daniel O.
    Grant, Gerald T.
    Ye, Ling
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 2018, 120 (03): : 382 - 388
  • [9] Comparative analysis of intraoral scanners accuracy using 3D software: an in vivo study
    Federica Pellitteri
    Paolo Albertini
    Angelica Vogrig
    Giorgio Alfredo Spedicato
    Giuseppe Siciliani
    Luca Lombardo
    [J]. Progress in Orthodontics, 23
  • [10] Comparative analysis of intraoral scanners accuracy using 3D software: an in vivo study
    Pellitteri, Federica
    Albertini, Paolo
    Vogrig, Angelica
    Spedicato, Giorgio Alfredo
    Siciliani, Giuseppe
    Lombardo, Luca
    [J]. PROGRESS IN ORTHODONTICS, 2022, 23 (01)